h1. MSI set-up in more details
2
3
4 10 Amber Herold
This section shows some of the more important and complicated steps in the set-up checklist. It assumes that the user knows the nomenclatures. See the full documentation if you want the detailed explanation. It also assumes that general set-up and calibrations exist from previous sessions so that most calibrations can remain unchanged. The steps that are NRAMM specific are marked. MSI now features stage Eucentric height correction at each square.
5 1 Amber Herold
6 10 Amber Herold
NOTE: The Tietz imaging software (EMMenu) must be turned OFF before Leginon is started. In contrast to this, the Gatan imaging software (Digital Micrograph) must be ON before Leginon is started.
7 1 Amber Herold
8
9
10 10 Amber Herold
h2. Starting Up
11 1 Amber Herold
12
13
14 10 Amber Herold
Set up is most stable with a good microscope alignment.
15 1 Amber Herold
16 11 Amber Herold
1.  Initial alignment at the microscope See Chapter on [[Microscope Set-up|Microscope setup]] for details.
17 1 Amber Herold
18 10 Amber Herold
  *  scope> Insert specimen holder
19
  *  scope> Put stage to eucenter and reset defocus in both HM mode AND LM mode
20
  *  scope> Check gun tilt and shift
21
  *  scope> Check rotation center and beam tilt pivot points in the HM mode
22
  *  scope> Save the goniomenter position for the square with intact supporting film where the alignment was done and do the same for a square where the supporting film is broken (where bright/dark images can be acquired).
23 1 Amber Herold
24
25 11 Amber Herold
2. Start python-based leginon main program amd leginon
26 1 Amber Herold
client. See Chapter on [[Microscope Set-up|Start Leginon]].
27
28 10 Amber Herold
  *  scope> Set up [[MSI set-up in more details#Starting-Up|CCD camera]]
29 11 Amber Herold
  *  [[Run Leginon Client on the Microscope Computer|scope> Start Leginon Client]]
30 10 Amber Herold
  *  scope> If the following error message appears, DO NOT touch it, or Leginon Client will need restarting
31 3 Alex Kadokura
32 10 Amber Herold
bq. _"Exception EAccessViolation in module adaExp.exe at...._
33
_Access violation at address xxxx in module 'adaExp.exe'. Read of address xxxx"_
34 1 Amber Herold
35 10 Amber Herold
  * [[Start Leginon at your main working station remotely|main> start-leginon.py]]
36
  *  Leginon Setup Wizard> Create new or Select old session, connect to the instrument host, and start.
37 1 Amber Herold
38
39 11 Amber Herold
3.  Start your preferred "MSI" application.
40 1 Amber Herold
41 13 Amber Herold
 *  Leginon/Application/Run> Select the "MSI" application you would like to use (Edge,T,Raster, or Tomography), the scope, and the main launcher. Click Run.
42
 *  Leginon/Node Selector> Select "Presets Manager" node.
43
 * Leginon/Presets Manager/Preset Selector> Import presets from a previous working session !http://emg.nysbc.org/software/leginon/images/icons/import.png!
44
 ** Select the TEM and camera to "Find" sessions closest to the condition of your data collection.
45
 ** Use shift-key to select a series of presets if needed. Hold-down to ctrl-key allows individual selection/deselection.
46
 ** Click "Import" button to import, the presets should appear in the Presets Manager/Preset Selector window.
47
 ** Click "OK" to exit "Import Presets" window.
48
 Note: Creating new preset and adjusting the preset are covered in Calibration Chapter under [[Creating Presets for the first time|Creating presets the first time]].
49
 *  The cycle order should match the sequence of data acquisition to minimize hysteresis. If the presets are imported from previous session, the cycle order is imported automatically. As an example, the cycle for naming scheme used here at NRAMM is gr -> sq -> hl -> fa -> fc -> en -> ef.
50
 * Leginon/Presets Manager> Select a preset in the list box and then click "To Scope" !http://emg.nysbc.org/software/leginon/images/icons/instrumentset.png! to apply the settings to the scope and synchronize Leginon and the scope settings.
51 1 Amber Herold
52
53 11 Amber Herold
4. Save the alignment and empty position in Leginon
54 1 Amber Herold
55 14 Amber Herold
*  scope> Recall the stored intact or empty position or go to such
56 1 Amber Herold
location.
57
58 14 Amber Herold
*  Leginon/Navigation/Toolbar/Stage Locations> !http://emg.nysbc.org/software/leginon/images/icons/stagelocations.png! Enter a New name to store Stage Location such as, for example, "empty" and "align", respectively. Save the xy coordinate only. To recall the position, select the named location under the same tool and click on "send to scope".
59 1 Amber Herold
60
61
h2. Grid preset image shift refinement
62
63 11 Amber Herold
64 16 Amber Herold
1.  Leginon/Presets Manager> Send a preset at an intermediate mag at HM mode as the reference (e.g. hl) "To Scope" !http://emg.nysbc.org/software/leginon/images/icons/instrumentset.png!
65 1 Amber Herold
66 15 Amber Herold
Note: sq preset in LM mode can be used if the image alignment between LM and HM
67 14 Amber Herold
is known to be good.
68 1 Amber Herold
69
70 14 Amber Herold
2.  scope> Move the grid to a position on the "align" square, either on a recognizable object or burn a good hole through ice at the center of the CCD as preparation for lower mag preset image shift correction.
71 1 Amber Herold
72 14 Amber Herold
73 16 Amber Herold
3.  Leginon/Presets Manager> Click on the tool "Align presets to each other" !http://emg.nysbc.org/software/leginon/images/icons/alignpresets.png!.
74 1 Amber Herold
75
76 14 Amber Herold
4.  Check, adjust, and save the preset(s) to be aligned.
77 1 Amber Herold
78 18 Amber Herold
The left panel displays and navigate by "Stage Position" of the reference preset while the right panel displays and navigate by "Image Shift" of the preset to be aligned
79 14 Amber Herold
 *  Leginon/Presets Manager/Align Presets> Choose the reference (hl) preset on the left panel.
80 1 Amber Herold
81 14 Amber Herold
 *  Leginon/Presets Manager/Align Presets> Choose the aligning (gr) preset on the right panel.
82 1 Amber Herold
83 14 Amber Herold
 *  Leginon/Presets Manager/Align Presets> Select "Similar look across mags" mode at the top left corner of the window. This will use a smaller part of the CCD to acquire the image for the lower mag preset so that it appears to be magnified in the 512x512 image panel.
84 1 Amber Herold
85 14 Amber Herold
 *  Leginon/Presets Manager/Align Presets> "Start" image acquisition.
86 1 Amber Herold
87 14 Amber Herold
 *  Leginon/Presets Manager/Align Presets> If the recognizable object can not be found in both panels, go back to step one to choose a lower mag preset or a better object.
88 1 Amber Herold
89 14 Amber Herold
 *  Leginon/Presets Manager/Align Presets> If the recognizable object can be found in both panels but not at the center on the reference left panel, navigate* it to the center using its "Click Tool". Wait for the move to complete and both images refreshed.
90 1 Amber Herold
91 14 Amber Herold
 *  Leginon/Presets Manager/Align Presets> Navigate* the object on the right panel to the center using its "Click Tool".The new image shift is automatically saved at each move.
92 1 Amber Herold
93 14 Amber Herold
 *  Repeat 7 if necessary.
94 1 Amber Herold
95
96 14 Amber Herold
Note: "Navigate" means selecting the navigate tool !http://emg.nysbc.org/software/leginon/images/icons/arrow.png! on the top right of image display and then double clicking the location of the new center that is translated by the given TEM Parameter.
97 1 Amber Herold
98
99
100 14 Amber Herold
h2. Grid preset stage position refinement
101 1 Amber Herold
102 3 Alex Kadokura
103 1 Amber Herold
104 14 Amber Herold
Stage position model Scale and Rotation is stable as long as the LM eucentric height defocus is set to zero before each session, although variation from day to day is normal. Check the stage movement calibration of gr preset by navigating it in Navigation node. Within 10% error is satisfactory since reaching a square does not require high accuracy. If not, perform stage model position _[[Modeled Stage Position calibration#Magnification-and-image-rotation-adjustment-of-the-model|scale and rotation calibration]]> at gr preset_ using "Calibration" application as following:
105 1 Amber Herold
106 14 Amber Herold
_[[Modeled Stage Position calibration|Why goniometer model is better]]._
107 1 Amber Herold
108
109
110 14 Amber Herold
1.  scope> Remove the objective aperture.
111 1 Amber Herold
112 14 Amber Herold
2.  Leginon> Run "Calibrations" Application
113 1 Amber Herold
114 14 Amber Herold
3.  Leginon/Presets Manager> Send "gr" preset to scope.
115 1 Amber Herold
116 14 Amber Herold
4.  Leginon/GonioModeling/Settings/Measurement> Set axis=x, Points=5,Tolerance=25%, and Interval=1e-05 m. Default model label is the session name.
117 1 Amber Herold
118 14 Amber Herold
5.  Leginon/GonioModeling/Settings/Modeling> Set "Scale and Rotation Adjustment Only"=Yes. Other settings will be set automatically once measurements are finished.
119 1 Amber Herold
120 14 Amber Herold
6.  Leginon/GonioModeling> Click on !http://emg.nysbc.org/software/leginon/images/icons/cam_ruler.png! will start "Measure" the shifts of between points that are moved on x axis of the goniometer.
121 1 Amber Herold
122 14 Amber Herold
7.  Leginon/GonioModeling> Click on !http://emg.nysbc.org/software/leginon/images/icons/play.png! will "Calibrate" the scale and rotation of the model on x axis of the goniometer.
123 1 Amber Herold
124 14 Amber Herold
8.  Leginon/GonioModeling> repeat 4-7 with the axis set to y.
125 1 Amber Herold
126
127
128
129
h2. Acquire mosaic atlas of the grid
130
131
132 14 Amber Herold
1.  scope> Remove the objective apperture.
133 1 Amber Herold
134 14 Amber Herold
2.  Leginon/Grid Targeting> Click "Settings" to check the preset used, the size and name of the atlas.
135 1 Amber Herold
136 14 Amber Herold
3. Leginon/Grid Targeting> Click "Calculate Atlas" !http://emg.nysbc.org/software/leginon/images/icons/calculate.png!
137 1 Amber Herold
138 14 Amber Herold
4.  Leginon/Grid Targeting> Click "Submit Targets" !http://emg.nysbc.org/software/leginon/images/icons/play.png! to begin acquiring the atlas.
139 1 Amber Herold
140 14 Amber Herold
5.  Leginon/Square Targeting> Follow the atlas progress in the display. (***There is no sign at the end currently other than that nothing happens for a long time and that the atlas dimension is the same in other direction when it is completed.)
141 1 Amber Herold
142 14 Amber Herold
6.  scope> Insert the objective aperture.
143 1 Amber Herold
144 14 Amber Herold
7.  Leginon/Presets Manager/Send an HM preset "To Scope" several times to cycle all magnifications and warm up the objective lens to minimize the image shift hysteresis in switching between LM and HM modes. If the scope has been in LM mode for a long time, wait for at least 30 min before final image shift refinement of needed LM preset during acquisition (i.e. sq)
145 1 Amber Herold
146
147
148
149 14 Amber Herold
h2. Determine the proper exposure condition for exposure presets
150 1 Amber Herold
151
152
153 17 Amber Herold
Exposure presets are the base presets for data collection. To keep it to low dose condition such as 10 electron/Angstrum, measurement of dose is performed and the exposure time is adjusted. If scope or camera parameters are changed during the session, the dose value is no longer valid and should be remeasured.
154 1 Amber Herold
155
156 14 Amber Herold
1.  Leginon/Navigation/Toolbar/Stage Locations !http://emg.nysbc.org/software/leginon/images/icons/stagelocations.png! Send the "empty" position "To scope".
157 1 Amber Herold
158 14 Amber Herold
2.  Leginon/Presets Manager/Preset Selector> Select a high mag exposure (i.e. "en" or "ef"), confirm that the parameters are what you desire for the experiment, and then send "To Scope" !http://emg.nysbc.org/software/leginon/images/icons/instrumentset.png!.
159 3 Alex Kadokura
160 14 Amber Herold
3.  scope> Expand the beam to at least cover the CCD (> medium circle mark on the viewing screen) and to the size you desire.
161 1 Amber Herold
162 14 Amber Herold
4.  Leginon/Presets Manager/Preset Selector> If you have changed the size or location of the beam or other preset scope parameters at the scope, you must retrieve
163
the preset parameter "From scope" !http://emg.nysbc.org/software/leginon/images/icons/instrumentget.png! first before checking and saving the dose value.
164 1 Amber Herold
165 14 Amber Herold
5.  Leginon/Presets Manager/Preset Selector> "Acquire dose image" !http://emg.nysbc.org/software/leginon/images/icons/acquire.png! for the selected preset.
166 1 Amber Herold
167 14 Amber Herold
6.  Leginon/Presets Manager/Dose Image> Check the "Dose" value.
168 1 Amber Herold
169 14 Amber Herold
7.  Leginon/Presets Manager/Dose Image> If desired dose is known, fill in the value and then click "Match". This will change exposure time to give a close match of
170 1 Amber Herold
the value entered.
171 3 Alex Kadokura
172 14 Amber Herold
8.  Leginon/Presets Manager/Dose Image> If not satisfied with the value, click "No", change the exposure time by editing the preset, and try again (No need to send
173 1 Amber Herold
the preset "To scope" again for the repeat)
174
175 14 Amber Herold
9.  If the scope can not give such illumination to satisfy dose, exposure time and area requirement, alignment is required or change of gun lens or spot size is needed.
176 1 Amber Herold
177 14 Amber Herold
10.  Leginon/Presets Manager> Copy the preset parameter to the other exposure preset.
178 1 Amber Herold
179 14 Amber Herold
 * Preset Selector> send the preset that has the dose measurement "To Scope" !http://emg.nysbc.org/software/leginon/images/icons/instrumentset.png!
180
 * Preset Selector> select the preset to receive the copy
181
 * Preset Selector> retrieve the parameters "From Scope" !http://emg.nysbc.org/software/leginon/images/icons/instrumentget.png! for the selected preset.
182
 * Preset Selector/Edit> !http://emg.nysbc.org/software/leginon/images/icons/edit.png! Modify defocus for the preset after the retrieval.
183 1 Amber Herold
184
185
186 14 Amber Herold
h2. Update bright/dark correction images for exposure presets
187 1 Amber Herold
188
189
190 14 Amber Herold
Exposure presets are the base presets for data collection. To make sure the best correction is applied to them, the correction images should be acquired prior to the experiment at session. Since no correlation is normally done on these images, only one [[Bright and Dark reference images#Correction-Channels|channel]] is needed. Normalization image might need to be [[Bright and Dark reference images#Correction-Planmodified]] if a large chunk of region is blocked by a fallen debris.
191 1 Amber Herold
192
193 14 Amber Herold
1.  Leginon/Presets Manager> Select a dose measured high mag exposure (i.e. "en" or "ef"), and then send "To Scope".
194 1 Amber Herold
195 14 Amber Herold
2.  Leginon/Correction/Settings>
196 1 Amber Herold
197 14 Amber Herold
 *  Select TEM and camera to match that will be used.
198
 
199
 *  Camera Configuration: The exposure preset camera configuration
200 1 Amber Herold
201 14 Amber Herold
3.  Leginon/Correction/Settings> Click OK to exit settings.
202 1 Amber Herold
203 14 Amber Herold
4.  Leginon/Correction/Toolbar> Select "Raw image " from the pull down list of acquisition modes and then click on "Acquire" button next to the selector to view an image that is not corrected.
204 1 Amber Herold
205 14 Amber Herold
5.  Leginon/Correction/Toolbar> Select "Dark reference" in the acquisition mode selector and then click "Acquire" to acquire the Dark reference image for this particular camera configuration.
206 1 Amber Herold
207 14 Amber Herold
6.  Leginon/Correction/Toolbar> Select "Bright reference" and repeat the acquisition to obtain the Bright reference.
208 1 Amber Herold
209 14 Amber Herold
7.  Leginon/Correction/Toolbar> Select "Corrected image" and then "Acquire" to view the corrected image. A corrected image should be free of artifacts and have
210 1 Amber Herold
smaller standard deviation than the raw image, in general.
211
212 14 Amber Herold
8.  Leginon/Correction/Toolbar> If there are regions consistently over normalizes the acquired image, either select the region and use the "+" tool to "Add Region to Bad Pixel List", or click on !http://emg.nysbc.org/software/leginon/images/icons/stagelocations.png! to "Add Extreme Points to Bad Pixel List" one of more times to find the outliers. Right click on a shown bad pixel removes it from the list.
213 1 Amber Herold
214
215
216
h2. Medium mag preset image and beam shift refinement
217
218
219 14 Amber Herold
Preset image shift alignment can be done follow the same procedure as the section "[[MSI set-up in more details#Grid-preset-image-shift-refinement|Grid preset image shift refinement]]" but is more efficient using the automated mode as follows.
220 1 Amber Herold
221
222 22 Amber Herold
1.  Leginon/Presets Manager> Send one of the exposure preset as the reference (e.g. en) "To Scope"  !http://emg.nysbc.org/software/leginon/images/icons/instrumentset.png!
223 1 Amber Herold
224 14 Amber Herold
2.  scope> Move the grid to a position on the "align" square, either on a recognizable object or burn a good hole through ice at the center of the CCD as preparation for lower mag preset image shift correction.
225 1 Amber Herold
226 14 Amber Herold
3.  Leginon/Presets Manager> Click on the tool "Align presets to each other" !http://emg.nysbc.org/software/leginon/images/icons/alignpresets.png!. The tool automatically select the reference preset that was selected in the Presets Manager mainpanel and choose a preset at closest magnification as the starting aligning preset.
227 1 Amber Herold
228 14 Amber Herold
4.  Leginon/Presets Manager/Align Presets> Click "Start" to acquire the first image pair.
229 1 Amber Herold
230 14 Amber Herold
5.  Check, adjust, and save the preset(s) to be aligned.
231 1 Amber Herold
232 19 Amber Herold
The left panel displays and navigate by "Stage Position" of the reference preset while the right panel displays and navigate by "Image Shift" of the preset to be aligned
233 1 Amber Herold
234 14 Amber Herold
 *  Leginon/Presets Manager/Align Presets> Changing Selection of "Similar look across mags" or "Full CCD" mode at the top left corner of the window will cause reacquisition of the image pair in the automated mode.
235 1 Amber Herold
236 14 Amber Herold
 *  Leginon/Presets Manager/Align Presets> If the recognizable object can be found in both panels but not at the center on the reference left panel, navigate* it to the center using its "Click Tool". Wait for the move to complete and both images refreshed.
237 1 Amber Herold
238 14 Amber Herold
 *  Leginon/Presets Manager/Align Presets> Navigate* the object on the right panel to the center using its "Click Tool". The new image shift is automatically saved after each move.
239 1 Amber Herold
240 14 Amber Herold
 *  Repeat 3 if necessary.
241 1 Amber Herold
242 14 Amber Herold
 *  Leginon/Presets Manager/Align Presets> Click "Continue" to save the alignment to all presets of the same magnification and continue to the next preset pair.
243 1 Amber Herold
244 14 Amber Herold
6.  Repeat the process if desired.
245 1 Amber Herold
246
247
248 8 Anchi Cheng
h2. Go to MSI Operation section
249
250 1 Amber Herold
251
252 21 Amber Herold
The setup that requires you to be next to the microscope is completed. Go to the section on [[MSI Operation|MSI Operation]] You can come back and continue on this section after you have learned how to go through the operation without automated hole finding.
253 1 Amber Herold
254
255
256
h2. Hole Targeting Set-up
257
258
259 21 Amber Herold
We can use any image that is already loaded into Hole Targeting to adjust these parameters.  The values in this node vary so much from grid to grid that the best way to find these parameters is by going through a step-by-step trial-and-error process. To see the final acquisition and focus targets, enable only the Original, acquisition, and focus display selections. Refer to Introduction Chapter on [[MSI set-up in more details|how the various tools behave in the control panel]].
260 1 Amber Herold
261 4 Alex Kadokura
262 21 Amber Herold
The loaded image normally comes from the image acquired through "Square" node when you submit targets on the grid atlas in "Square Targeting" node.  You can also do a practice run using the default image comes with Leginon. Leave the filename entry blank when you click "Load" in the hole finding settings window for the step "Original". The default values in the node gives an example of reasonable target finding with such default image.
263 1 Amber Herold
264
265
266
Setting up Hole Targeting Depends on which MSI flavor you are running:
267
268
269 21 Amber Herold
1. [[Hole Targeting Set-up for MSI-Edge|MSI-Edge]] uses template matching of the edge image
270 1 Amber Herold
271 21 Amber Herold
2. [[JAHC finder nodes set-up|MSI-T]] uses template matching of the original image
272 1 Amber Herold
273 21 Amber Herold
3. [[raster finder nodes set-up|MSI-Raster]] creates raster points from a given origin
274 1 Amber Herold
275 21 Amber Herold
4. [[Initial MSI application preference setup#Settings-that-are-the-same-as-in-other-MSI-applications-in-queuing-mode|MSI-Tomography]] does not use automatic target finding. Therefore, it does not requires set up.
276 1 Amber Herold
277
278
279
h2.  Exposure Targeting Set-up
280
281
282
283 21 Amber Herold
Like hole targeting, we can use any image that is already loaded into Exposure Targeting to adjust these parameters.
284 1 Amber Herold
285
Setting up Exposure Targeting Depends on which MSI flavor you are running.
286
287
288 21 Amber Herold
1. [[Hole Targeting Set-up for MSI-Edge|MSI-Edge]] uses template matching of the edge image
289 1 Amber Herold
290
291 21 Amber Herold
2. [[JAHC finder nodes set-up|MSI-T]] uses template matching of the original image
292 1 Amber Herold
293 4 Alex Kadokura
294 21 Amber Herold
3. [[raster finder nodes set-up|MSI-Raster]] creates raster points from a given origin
295 1 Amber Herold
296
297 21 Amber Herold
4. [[Initial MSI application preference setup#Settings-that-are-the-same-as-in-other-MSI-applications-in-queuing-mode|MSI-Tomography]] does not use automatic target finding. Therefore, it does not requires set up.
298 1 Amber Herold
299
300
301
h2. Autofocus Test
302
303
304 4 Alex Kadokura
305 21 Amber Herold
There are two focusers in MSI. "Z Focus" node usually moves stage in Z direction to Eucentric height at lower mags. "Focus" node usually just correct the defocus in the way equivalent to turning the focus adjustment knob on the scope at a higher mag. The defocus measurements in both nodes are based on the same principle of beam-tilt induced image shift. A correlation (normally phase) is used to obtain the image shift. Therefore, it is important to test and preform autofocus measurement at area with recognizable feature that does not move significantly during the defocus measurement.
306 1 Amber Herold
307
308 21 Amber Herold
The test use "Manual_after" focus step in the focus sequence for checking. Two procedures can be used to reach a area for good correlation measurements:
309 1 Amber Herold
310
311 21 Amber Herold
1.  Using Simulate Target tool
312 1 Amber Herold
313 21 Amber Herold
 *  Leginon/Presets Manager> Send the autofocus preset "To Scope"
314 1 Amber Herold
315 21 Amber Herold
 *  scope> Move to an area that has recognizable feature on the carbon support at the center.
316 1 Amber Herold
317 21 Amber Herold
 *  Leginon/(Z) Focus/Focus Sequence !http://emg.nysbc.org/software/leginon/images/icons/focus_sequence.png! > Enable desired autofocus and "Manual_after" focus steps.
318 1 Amber Herold
319 21 Amber Herold
 *  Leginon/(Z) Focus/Toolbar> Click Simulate Target !http://emg.nysbc.org/software/leginon/images/icons/simulatetarget.png! and follow the autofocus
320 1 Amber Herold
321
322 21 Amber Herold
2.  Submitting focus target from Hole Targeting
323 1 Amber Herold
324 21 Amber Herold
Note: To use this procedure, there has to be a grid mosaic, and the Hole Targeting should allow user verification of targets
325 1 Amber Herold
326
327 23 Amber Herold
 *  Leginon/(Z) Focus/Focus Sequence !http://emg.nysbc.org/software/leginon/images/icons/focus_sequence.png! > Enable desired autofocus and "Manual_after" focus steps.
328 1 Amber Herold
329 21 Amber Herold
 *  Leginon/Square Targeting> Select a test square and submit it !http://emg.nysbc.org/software/leginon/images/icons/play.png!
330 1 Amber Herold
331 21 Amber Herold
 *  Leginon/Hole Targeting> Submit a focus target to test autofocus in Z Focus node and an acquisition target for the next stage. Both targets should be in an area that has recognizable feature with carbon support.
332 1 Amber Herold
333 21 Amber Herold
 *  Leginon/Z focus> follow the autofocus test in this node.
334 1 Amber Herold
335 21 Amber Herold
 *  Leginon/Exposure Targeting> Submit a focus target at a location that has recognizable feature with carbon support to test autofocus in Focus node.
336 1 Amber Herold
337 21 Amber Herold
 *  Leginon/Focus> follow autofocus test in this node.
338 1 Amber Herold
339
340 21 Amber Herold
Check the performance of autofocus by the following criteria:
341 1 Amber Herold
342
343 21 Amber Herold
1.  During autofocusing:
344 1 Amber Herold
345 21 Amber Herold
 *  Display the image and correlation peak: The observed movement of images upon beam tilt should be consistent with the correlation peak and the edge of the beam should not be visible in any of the beam-tilt images.
346 1 Amber Herold
347 21 Amber Herold
 *  Display the correlation image: The correlation peak at its correct position should be significantly higher than background.
348 1 Amber Herold
349
350 21 Amber Herold
2.  During manual focusing
351 1 Amber Herold
352 21 Amber Herold
Manual Focusing can be activated either by enabling "Manual_before/Manual_after" steps or directly when leginon is idle using (Z) Focus/Toolbar/MF. To activate manual focus directly, it is necessary to send to scope the preset of the choice (normally fc) first, though.
353 1 Amber Herold
354 21 Amber Herold
 *  MF> The FFT should show visible and isotropic Thon ring according to the fc preset defocus.
355 1 Amber Herold
356 21 Amber Herold
 *  MF/Toolbar> Make sure the move type is "Defocus" and the set value in the box is 0 m then click "Set instrument" botton !http://emg.nysbc.org/software/leginon/images/icons/instrumentset.png!. This will set the defocus to the current 0.
357 1 Amber Herold
358 21 Amber Herold
 *  MF> The perfect autofocus result is that the FFT indicates that the scope is at Gaussian focus.
359 1 Amber Herold
360 21 Amber Herold
 *  For relative movement, press "+" or "-". The step size is determined in MF/Toolbar/Settings>
361 2 Amber Herold
362 24 Amber Herold
 *  To reset defocus to zero at the new focus, press the so-named button !http://emg.nysbc.org/software/leginon/images/icon/instrumentsetnew.png!.
363 2 Amber Herold
364 21 Amber Herold
 *  Clicking the Stop button !http://emg.nysbc.org/software/leginon/images/icons/stop.png! exits the manual check.
365 2 Amber Herold
366
367
368 1 Amber Herold
______
369
370
[[MSI Quick-start|< MSI Quick-start]] | [[Special Operation setup|Special Operation setup >]]
371