MSI set-up in more details

Starting Up

Set up is most stable with a good microscope alignment.

  • Initial alignment at the microscope See Chapter on Microscope setup for details.
  • scope> Insert specimen holder
  • scope> Put stage to eucenter and reset defocus in both HM mode AND LM
    mode
  • scope> Check gun tilt and shift
  • scope> Check rotation center and beam tilt pivot points in the HM
    mode
  • scope> Save the goniomenter position for the square with intact
    supporting film where the alignment was done and do the same for a square where
    the supporting film is broken (where bright/dark images can be acquired).
  • Start python-based leginon main program amd leginon
    client. See Chapter on Start Leginon.
  • scope> If the error message

"Exception EAccessViolation in module adaExp.exe at
....

Access violation at address xxxx in module 'adaExp.exe'. Read of
address xxxx"

appears, DO NOT touch it, or Leginon Client will need restarting

  • Leginon Setup Wizard> Create new or Select old session, connect to the
    instrument host, and start.
  • Start your preferred "MSI" application.
  • Leginon/Application/Run> Select the "MSI" application you would like to
    use (Edge,T,Raster, or Tomography), the scope, and the main launcher. Click
    Run.
  • Leginon/Node Selector> Select "Presets Manager" node.
  • Leginon/Presets Manager/Preset Selector> Import
    presets from a previous working session <inlinegraphic
    fileref="http://emg.nysbc.org/software/leginon/images/icons/import.png"
    format="PNG"/>
  • Select the TEM and camera to "Find" sessions closest to the condition of
    your data collection.
  • Use shift-key to select a series of presets if needed. Hold-down to
    ctrl-key allows individual selection/deselection.
  • Click "Import" button to import, the presets should appear in the
    Presets Manager/Preset Selector window.
  • Click "OK" to exit "Import Presets" window.
  • The cycle order should match the sequence of data acquisition to minimize
    hysterisis. If the presets are imported from previous session, the cycle order
    is imported automatically. As an example, the cycle for naming scheme used here
    at NRAMM is gr -> sq -> hl -> fa -> fc -> en -> ef.
  • Leginon/Presets Manager> Select a preset in
    the list box and then click "To Scope" <inlinegraphic
    fileref="http://emg.nysbc.org/software/leginon/images/icons/instrumentset.png"
    format="PNG"/> to apply the settings to the scope and synchronize Leginon and
    the scope settings.
  • Save the alignment and empty position in Leginon
  • scope> Recall the stored intact or empty position or go to such
    location.
  • Leginon/Navigation/Toolbar/Stage Locations>
    <inlinegraphic
    fileref="http://emg.nysbc.org/software/leginon/images/icons/stagelocations.png"
    format="PNG"/> Enter a New name to store Stage Location such as, for example,
    "empty" and "align", respectively. Save the xy coordinate only. To recall the
    position, select the named location under the same tool and click on "send to
    scope".

Grid preset image shift refinement

  • Leginon/Presets Manager> Send a preset at an intermediate mag at HM mode as
    the reference (e.g. hl) "To Scope"<inlinegraphic
    fileref="http://emg.nysbc.org/software/leginon/images/icons/instrumentset.png"
    format="PNG"/>

(Note: sq preset in LM mode can be used if the image alignment between LM and HM
is known to be good).

  • scope> Move the grid to a position on the "align" square, either on a
    recognizable object or burn a good hole through ice at the center of the CCD as
    preparation for lower mag preset image shift correction.
  • Leginon/Presets Manager> Click on the tool "Align presets to each other"
    <inlinegraphic
    fileref="http://emg.nysbc.org/software/leginon/images/icons/alignpresets.png"
    format="PNG"/> .
  • Check, adjust, and save the preset(s) to be aligned.
The left panel displays and navigate by "Stage Position" of the reference preset
while the right panel displays and navigate by "Image Shift" of the preset to be
aligned
  • Leginon/Presets Manager/Align Presets> Choose the reference (hl) preset
    on the left panel.
  • Leginon/Presets Manager/Align Presets> Choose the aligning (gr) preset
    on the right panel.
  • Leginon/Presets Manager/Align Presets> Select "Similar look across mags"
    mode at the top left corner of the window. This will use a smaller part of the
    CCD to acquire the image for the lower mag preset so that it appears to be
    magnified in the 512x512 image panel.
  • Leginon/Presets Manager/Align Presets> "Start" image acquisition.
  • Leginon/Presets Manager/Align Presets> If the recongnizable object can
    not be found in both panels, go back to step one to choose a lower mag preset or
    a better object.
  • Leginon/Presets Manager/Align Presets> If the recongnizable object can
    be found in both panels but not at the center on the reference left panel,
    navigate* it to the center using its "Click Tool". Wait for the move to complete
    and both images refreshed.
  • Leginon/Presets Manager/Align Presets> Navigate* the object on the right
    panel to the center using its "Click Tool".The new image shift is automatically
    saved at each move.
  • Repeat 7 if necessary.
  • "Navigate" means selecting the navigate tool<inlinegraphic
    fileref="http://emg.nysbc.org/software/leginon/images/icons/arrow.png" format="PNG"/> on
    the top right of image display and then double clicking the location of the new center that
    is translated by the given TEM Parameter.

Grid preset stage postion refinement

Stage position model Scale and Rotation is stable as long as the LM eucentric height
defocus is set to zero before each session, although variation from day to day is normal.
Check the stage movement calibration of gr preset by navigating it in Navigation node.
Within 10% error is satisfactory since reaching a square does not require high accuracy. If
not, perform stage model position scale and rotation
calibration
> at gr preset
using "Calibration" application as
following:

Why goniometer model is
better
.

  • scope> Remove the objective apperture.
  • Leginon> Run "Calibrations" Application
  • Leginon/Presets Manager> Send "gr" preset to scope.
  • Leginon/GonioModeling/Settings/Measurement> Set axis=x,
    Points=5,Tolerance=25%, and Interval=1e-05 m. Default model label is the session
    name.
  • Leginon/GonioModeling/Settings/Modeling> Set "Scale and Rotation Adjustment
    Only"=Yes. Other settings will be set automatically once measurements are
    finished.
  • Leginon/GonioModeling> Click on <inlinegraphic
    fileref="http://emg.nysbc.org/software/leginon/images/icons/cam_ruler.png"
    format="PNG"/> will start "Measure" the shifts of between points that are moved on x
    axis of the goniometer.
  • Leginon/GonioModeling> Click on <inlinegraphic
    fileref="http://emg.nysbc.org/software/leginon/images/icons/play.png" format="PNG"
    /> will "Calibrate" the scale and rotation of the model on x axis of the
    goniometer.
  • Leginon/GonioModeling> repeat 4-7 with the axis set to y.

Acquire mosaic atlas of the grid

  • scope> Remove the objective apperture.
  • Leginon/Grid Targeting> Click "Settings" to check the preset used, the size
    and name of the atlas.
  • Leginon/Grid Targeting> Click "Calculate Atlas" <inlinegraphic
    fileref="http://emg.nysbc.org/software/leginon/images/icons/calculate.png"
    format="PNG"/>
  • Leginon/Grid Targeting> Click "Submit Targets" <inlinegraphic
    fileref="http://emg.nysbc.org/software/leginon/images/icons/play.png" format="PNG"
    />to begin acquiring the atlas.
  • Leginon/Square Targeting> Follow the atlas progress in the display. (***There
    is no sign at the end currently other than that nothing happens for a long time and
    that the atlas dimension is the same in other direction when it is completed.)
  • scope> Insert the objective aperture.
  • Leginon/Presets Manager/Send an HM preset "To Scope" several times to cycle all
    magnifications and warm up the objective lens to minimize the image shift hysteresis
    in switching between LM and HM modes. If the scope has been in LM mode for a long
    time, wait for at least 30 min before final image shift refinement of needed LM preset
    during acquisition (i.e. sq)

Determine the proper exposure condition for exposure
presets

Exposure presets are the base presets for data collection. To keep it to low dose
condition such as 10 electron/Angstrum, measurement of dose is performed and the exposure
time is adjusted. If scope or camera parameters are changed during the session, the dose
value is no longer valid and should be remeasured.

  • Leginon/Navigation/Toolbar/Stage Locations<inlinegraphic
    fileref="http://emg.nysbc.org/software/leginon/images/icons/stagelocations.png"
    format="PNG"/>> Send the "empty" position "To scope".
  • Leginon/Presets Manager/Preset Selector> Select a high mag exposure (i.e. "en"
    or "ef"), confirm that the parameters are what you desire for the experiment, and then
    send "To Scope"<inlinegraphic
    fileref="http://emg.nysbc.org/software/leginon/images/icons/instrumentset.png"
    format="PNG"/>.
  • scope> Expand the beam to at least cover the CCD (> medium circle mark on
    the viewing screen) and to the size you desire.
  • Leginon/Presets Manager/Preset Selector> If you have changed the size or
    location of the beam or other preset scope parameters at the scope, you must retrieve
    the preset parameter "From scope"<inlinegraphic
    fileref="http://emg.nysbc.org/software/leginon/images/icons/instrumentget.png"
    format="PNG"/> first before checking and saving the dose value.
  • Leginon/Presets Manager/Preset Selector> "Acquire dose image"<inlinegraphic
    fileref="http://emg.nysbc.org/software/leginon/images/icons/acquire.png"
    format="PNG"/> for the selected preset.
  • Leginon/Presets Manager/Dose Image> Check the "Dose" value.
  • Leginon/Presets Manager/Dose Image> If desired dose is known, fill in the
    value and then click "Match". This will change exposure time to give a close match of
    the value entered.
  • Leginon/Presets Manager/Dose Image> If not satisfied with the value, click
    "No", change the exposure time by editing the preset, and try again (No need to send
    the preset "To scope" again for the repeat)
  • If the scope can not give such illumination to satisfy dose, exposure time and
    area requirement, alignment is required or change of gun lens or spot size is
    needed.
  • Leginon/Presets Manager> Copy the preset parameter to the other exposure
    preset.
  • Preset Selector> send the preset that has the dose measurement "To
    Scope"<inlinegraphic
    fileref="http://emg.nysbc.org/software/leginon/images/icons/instrumentset.png"
    format="PNG"/>
  • Preset Selector> select the preset to recieve the copy
  • Preset Selector> retrieve the parameters "From Scope"<inlinegraphic
    fileref="http://emg.nysbc.org/software/leginon/images/icons/instrumentget.png"
    format="PNG"/> for the selected preset.
  • Preset Selector/Edit><inlinegraphic
    fileref="http://emg.nysbc.org/software/leginon/images/icons/edit.png"
    format="PNG"/> Modify defocus for the preset after the retrieval.

Update bright/dark correction images for exposure
presets

Exposure presets are the base presets for data collection. To make sure the best
correction is applied to them, the correction images should be acquired prior to the
experiment at session. Since no correlation is normally done on these images, only one channel is needed. Normalization image might need to be
Bright and Dark reference images if a large chunk of region is blocked by a
fallen debris.

  • Leginon/Presets Manager> Select a dose measured high mag exposure (i.e. "en"
    or "ef"), and then send "To Scope".
  • Leginon/Correction/Settings>
  • Select TEM and camera to match that will be used.
  • Camera Configuration: The exposure preset camera configuration
  • Leginon/Correction/Settings> Click OK to exit settings.
  • Leginon/Correction/Toolbar> Select "Raw image " from the pull down list of
    acquisition modes and then click on "Acquire" button next to the selector to view an
    image that is not corrected.
  • Leginon/Correction/Toolbar> Select "Dark reference" in the acquisition mode
    selector and then click "Acquire" to acquire the Dark reference image for this
    particular camera configuration.
  • Leginon/Correction/Toolbar> Select "Bright reference" and repeat the
    acquisition to obtain the Bright reference.
  • Leginon/Correction/Toolbar> Select "Corrected image" and then "Acquire" to
    view the corrected image. A corrected image should be free of artifacts and have
    smaller standard deviation than the raw image, in general.
  • Leginon/Correction/Toolbar> If there are regions consistently over normalizes
    the acquired image, either select the region and use the "+" tool to "Add Region to
    Bad Pixel List", or click on<inlinegraphic
    fileref="http://emg.nysbc.org/software/leginon/images/icons/stagelocations.png"
    format="PNG"/>to "Add Extreme Points to Bad Pixel List" one of more times to find
    the outliers. Right click on a shown bad pixel removes it from the list.

Medium mag preset image and beam shift
refinement

Preset image shift alignment can be done follow the same procedure as the section "Grid preset image shift refinement" but is more
efficient using the automated mode as follows.

  • Leginon/Presets Manager> Send one of the exposure preset as the reference
    (e.g. en) "To Scope"<inlinegraphic
    fileref="http://emg.nysbc.org/software/leginon/images/icons/instrumentset.png"
    format="PNG"/>
  • scope> Move the grid to a position on the "align" square, either on a
    recognizable object or burn a good hole through ice at the center of the CCD as
    preparation for lower mag preset image shift correction.
  • Leginon/Presets Manager> Click on the tool "Align presets to each other"
    <inlinegraphic
    fileref="http://emg.nysbc.org/software/leginon/images/icons/alignpresets.png"
    format="PNG"/> . The tool automatically select the reference preset that was
    selected in the Presets Manager mainpanel and choose a preset at closest magnification
    as the starting aligning preset.
  • Leginon/Presets Manager/Align Presets> Click "Start" to acquire the first
    image pair.
  • Check, adjust, and save the preset(s) to be aligned.
The left panel displays and navigate by "Stage Position" of the reference preset
while the right panel displays and navigate by "Image Shift" of the preset to be
aligned
  • Leginon/Presets Manager/Align Presets> Changing Selection of "Similar
    look across mags" or "Full CCD" mode at the top left corner of the window will
    cause reacquisition of the image pair in the automated mode.
  • Leginon/Presets Manager/Align Presets> If the recongnizable object can
    be found in both panels but not at the center on the reference left panel,
    navigate* it to the center using its "Click Tool". Wait for the move to complete
    and both images refreshed.
  • Leginon/Presets Manager/Align Presets> Navigate* the object on the right
    panel to the center using its "Click Tool". The new image shift is automatically
    saved after each move.
  • Repeat 3 if necessary.
  • Leginon/Presets Manager/Align Presets> Click "Continue" to save the
    alignment to all presets of the same magnification and continue to the next
    preset pair.
  • Repeat the process if so desired.

Go to MSI Operation section

The setup that requires you to be next to the microscope is completed. Go to the section
on MSI Operation.

Hole Targeting Set-up

We can use any image that is already loaded into Hole Targeting to adjust these
parameters. The values in this node vary so much from grid to grid that the best way to
find these parameters is by going through a step-by-step trial-and-error process. To see the
final acquisition and focus targets, enable only the Original, acquisition, and focus
display selections. Refer to Introduction Chapter on [[MSI set-up in more details

|how the various tools behave in the control panel]].

The loaded image normally comes from the image acquired through "Square" node when you
submit targets on the grid atlas in "Square Targeting" node. You can also do a practice run
using the default image comes with Leginon. Leave the filename entry blank when you click
"Load" in the hole finding settings window for the step "Original". The default values in
the node gives an example of reasonable target finding with such default image.

Setting up Hole Targeting Depends on which MSI flavor you are running:

  • MSI-Edge uses template matching of the
    edge image
  • MSI-T uses template matching of the original
    image
  • MSI-Raster creates raster points from a given
    origin
  • MSI-Tomography does not use automatic
    target finding. Therefore, it does not requires set up.

Exposure Targeting Set-up

Like hole targeting, we can use any image that is already loaded into Exposure Targeting
to adjust these parameters.

Setting up Exposure Targeting Depends on which MSI flavor you are running.

  • MSI-Edge uses template matching of the
    edge image
  • MSI-T uses template matching of the original
    image
  • MSI-Raster creates raster points from a given
    origin
  • MSI-Tomography does not use automatic
    target finding. Therefore, it does not requires set up.

Autofocus Test

There are two focusers in MSI. "Z Focus" node usually moves stage in Z direction to
Eucentric height at lower mags. "Focus" node usually just correct the defocus in the way
equivalent to turning the focus adjustment knob on the scope at a higher mag. The defocus
measurments in both nodes are based on the same principle of beam-tilt induced image shift.
A correlation (normally phase) is used to obtain the image shift. Therefore, it is important
to test and preform autofocus measurement at area with recognizable feature that does not
move significantly during the defocus measurement.

The test use "Manual_after" focus step in the focus sequence for checking. Two
procedures can be used to reach a area for good correlation measurements:

  • Using Simulate Target tool
  • Leginon/Presets Manager> Send the autofocus preset "To Scope"
  • scope> Move to an area that has recognizable feature on the carbon
    support at the center.
  • Leginon/(Z) Focus/Focus Sequence<inlinegraphic
    fileref="http://emg.nysbc.org/software/leginon/images/icons/focus_sequence.png"
    format="PNG"/>> Enable desired autofocus and "Manual_after" focus
    steps.
  • Leginon/(Z) Focus/Toolbar> Click Simulate Target<inlinegraphic
    fileref="http://emg.nysbc.org/software/leginon/images/icons/simulatetarget.png"
    format="PNG"/> and follow the autofocus
  • Submiting focus target from Hole Targeting

*To use this procedure, there has to be a grid mosaic, and the Hole Targeting
should allow user verification of targets

  • Leginon/(Z) Focus/Focus Sequence<inlinegraphic
    fileref="http://emg.nysbc.org/software/leginon/images/icons/focus_sequence.png"
    format="PNG"/>> Enable desired autofocus and "Manual_after" focus
    steps.
  • Leginon/Square Targeting> Select a test square and submit it<inlinegraphic
    fileref="http://emg.nysbc.org/software/leginon/images/icons/play.png"
    format="PNG"/>
  • Leginon/Hole Targeting> Submit a focus target to test autofocus in Z Focus
    node and an acquisition target for the next stage. Both targets should be in an
    area that has recognizable feature with carbon support.
  • Leginon/Z focus> follow the autofocus test in this node.
  • Leginon/Exposure Targeting> Submit a focus target at a location that has
    recognizable feature with carbon support to test autofocus in Focus node.
  • Leginon/Focus> follow autofocus test in this node.

Checking the performance of autofocus by the following
criteria:

  • During autofocusing
  • Display the image and correlection peak: The observed movement of images
    upon beam tilt should be consistent with the correlation peak and the edge of
    the beam should not be visible in any of the beam-tilt images.
  • Display the correlation image: The correlation peak at its correct position
    should be significantly higher than background.
  • During manual focusing

Manual Focusing can be activated either by enabling "Manual_before/Manual_after"
steps or directly when leginon is idle using (Z) Focus/Toolbar/MF. To activate manual
focus directly, it is necessary to send to scope the preset of the choice (normally
fc) first, though.

  • MF> The FFT should show visible and isotropic Thon ring according to the
    fc preset defocus.
  • MF/Toolbar> Make sure the move type is "Defocus" and the set value in the
    box is 0 m then click "Set instrument" botton<inlinegraphic
    fileref="http://emg.nysbc.org/software/leginon/images/icons/instrumentset.png"
    format="PNG"/>. This will set the defocus to the current 0.
  • MF> The perfect autofocus result is that the FFT indicates that the scope
    is at Gaussian focus.
  • For relative movement, press "+" or "-". The step size is determined in
    MF/Toolbar/Settings>
  • To reset defocus to zero at the new focus, press the so-named
    button<inlinegraphic
    fileref="http://emg.nysbc.org/software/leginon/images/icons/instrumentsetnew.png"
    format="PNG"/>.
  • Clicking the Stop button<inlinegraphic
    fileref="http://emg.nysbc.org/software/leginon/images/icons/stop.png"
    format="PNG"/> exits the maunal check.