1 | 1 | Amber Herold | h1. Manual Acquisition |
---|---|---|---|
2 | |||
3 | |||
4 | |||
5 | Manual Acquisition is a node that allows the electron microscopist to acquire CCD images |
||
6 | manually while at the microscope. This node essentially creates a button to acquire images |
||
7 | manually with Leginon and gives all the benefits of the Leginon database and image storage. |
||
8 | |||
9 | |||
10 | |||
11 | Required bindings: None |
||
12 | |||
13 | |||
14 | |||
15 | |||
16 | |||
17 | h2. Toolbar |
||
18 | |||
19 | |||
20 | |||
21 | |||
22 | |||
23 | * Settings = Adjust the manual image acquisition parameters. |
||
24 | |||
25 | |||
26 | * Grids = Select the current grid.(Not used without Project Database) |
||
27 | |||
28 | |||
29 | * Acquire = Acquire a current image. |
||
30 | |||
31 | |||
32 | * Continuous Acquire = Click this to begin acquiring images continuously. |
||
33 | |||
34 | |||
35 | * Abort = Stop the Continuous Acquire function. |
||
36 | |||
37 | |||
38 | |||
39 | |||
40 | |||
41 | |||
42 | |||
43 | |||
44 | |||
45 | h2. Settings |
||
46 | |||
47 | |||
48 | |||
49 | |||
50 | |||
51 | * Camera Configuration = Select the parameters that describe how the image is |
||
52 | acquired. |
||
53 | |||
54 | - Dimension = image x,y pixel dimension |
||
55 | |||
56 | - Offset = x,y offset from top left corner of full CCD image |
||
57 | |||
58 | - Binning = x,y image binning |
||
59 | |||
60 | - pre-configured image sizes |
||
61 | |||
62 | - Custom = customize image dimension, offset, and binning |
||
63 | |||
64 | - Exposure time (ms) = CCD and specimen expsosure time |
||
65 | |||
66 | |||
67 | * Correct image = When enabled, the raw CCD image will be corrected with Bright and |
||
68 | Dark CCD reference images |
||
69 | |||
70 | |||
71 | * Save image to database = When enabled, the acquired CCD image will be |
||
72 | automatically stored in the database. All microscope and camera information will be |
||
73 | stored for each image. |
||
74 | |||
75 | |||
76 | * Loop pause "30" seconds = The amount of time the Continuous Acquire function |
||
77 | waits before acquiring the next image. |
||
78 | |||
79 | |||
80 | * Label = Enter a label to be inserted in the image file name (in addition to the |
||
81 | normal Leginon file name).* |
||
82 | |||
83 | |||
84 | * Main Screen: Up before acquire = When enabled, Leginon will raise the |
||
85 | microscope's main viewing screen. |
||
86 | |||
87 | |||
88 | * Main Screen: Down after acquire = When enabled, Leginon will lower the |
||
89 | microscope's main viewing screen after an image has been acquired. |
||
90 | |||
91 | |||
92 | * Low Dose: |
||
93 | |||
94 | - Use low dose = When enabled, Leginon will switch the microscope from focus to |
||
95 | exposure mode of the Low Dose kit to acquire the image and then return to focus |
||
96 | mode. |
||
97 | |||
98 | - Low Dose pause "4" seconds = Wait this amount of time before acquiring an image |
||
99 | when using Low Dose on the microscope and acquiring images with the Acquire feature in |
||