Matrix Calibrator » History » Version 2
Amber Herold, 04/27/2010 02:44 PM
1 | 1 | Amber Herold | h1. Matrix Calibrator |
---|---|---|---|
2 | |||
3 | |||
4 | Matrix calibrator performs four different calibrations that based on linear transformation |
||
5 | conversion: Image shift, beam shift, stage position, and image/beam shift |
||
6 | |||
7 | <highlights> |
||
8 | |||
9 | Required bindings |
||
10 | |||
11 | |||
12 | Required bindings to use preset instrument configuration set by presets manager |
||
13 | |||
14 | <simplelist type="vert"> |
||
15 | <member> PresetsManagerNode - (PresetChangedEvent) -> MatrixCalibratorNode</member> |
||
16 | </simplelist> |
||
17 | </highlights> |
||
18 | |||
19 | |||
20 | |||
21 | Matrix calibration is made by making N sets of measurements (specified by "N Average"). |
||
22 | Each measurement set acquires three images, first at a given origin, second with an x-axis |
||
23 | movement in the specified "Parameter" by the specified "Shift Fraction" of the image, and |
||
24 | third with an y-axis movement by the same shift fraction. The resulting shifts in the acquired |
||
25 | images are obtained by cross correlation. A transformation matrix is then generated for the |
||
26 | measurement set. The origin is shifted by the "Interval" specified in the node, in meters |
||
27 | before the next set of measurements is taken. At the end, the N matrixes obtained are averaged |
||
28 | and saved in the database at the specific magnification and movement type and can be applied |
||
29 | to any camera configuration. "Tolerance", expressed in fraction of image, is used as an error |
||
30 | check. The calibration is considered failed when the measured movement is much different from |
||
31 | that calculated from pixel calibration |
||
32 | |||
33 | |||
34 | Because the calibration is based on correlation of the acquired images, well contrasted |
||
35 | images with the intended movement clearly shown but not by such a large amount that the |
||
36 | apparent feature is altered is the key for sucessful calibrations. |
||
37 | |||
38 | |||
39 | Image shift calibration is useful only on FEI microscope where image shift/beam shift |
||
40 | compensation is calibrated and stored at the scope. Without such stored calibration, Image |
||
41 | shift/Beam shift calibration is required instead. |
||
42 | |||
43 | |||
44 | |||
45 | |||
46 | |||
47 | h2. Toolbar |
||
48 | |||
49 | |||
50 | |||
51 | |||
52 | |||
53 | * Settings = Adjust the manual image acquisition parameters. |
||
54 | |||
55 | |||
56 | * Acquire Image = Acquire a current image. |
||
57 | |||
58 | |||
59 | * Stage Position | Image Shift | Beam Shift = Choose the type of calibration to |
||
60 | complete |
||
61 | |||
62 | |||
63 | * Parameter Settings = Adjust the parameters (and limits) for the calibration |
||
64 | process |
||
65 | |||
66 | |||
67 | * Calibrate = Start the calibration. |
||
68 | |||
69 | |||
70 | * Abort = Stop the Continuous Acquire function. |
||
71 | |||
72 | |||
73 | * Edit Current Calibration = Edit the matrix when the calibration is completed. Not |
||
74 | recommended for beginner. |
||
75 | |||
76 | |||
77 | |||
78 | |||
79 | |||
80 | |||
81 | |||
82 | |||
83 | |||
84 | h2. Settings |
||
85 | |||
86 | |||
87 | |||
88 | |||
89 | |||
90 | * Use "phase | cross" correlation = Select the type of correlation used to |
||
91 | calculate the calibration. |
||
92 | |||
93 | |||
94 | * Override Preset = use this node's <link linkend="camconfig">Instrument and Camera |
||
95 | Configuration</link> to acquire the images in this node |
||
96 | |||
97 | |||
98 | |||
99 | |||
100 | |||
101 | |||
102 | |||
103 | |||
104 | |||
105 | h2. Parameter Settings |
||
106 | |||
107 | |||
108 | |||
109 | |||
110 | |||
111 | * Tolerance = percent error allowed when calculating the calibration |
||
112 | |||
113 | |||
114 | * Shift Fraction = amount sequential images should shift relative to each |
||
115 | other |
||
116 | |||
117 | |||
118 | * Average = the number of sequential images that are acquired in order to calculate |
||
119 | the calibration |
||
120 | |||
121 | |||
122 | * Interval "2e-6" = the interval (m) that is used to move the stage/beam/image |
||
123 | position to take images required for the image sequence |
||
124 | |||
125 | |||
126 | * Use current positon as starting point = When enabled (default), the current stage |
||
127 | postion is used to begin the calibration |
||
128 | |||
129 | |||
130 | * Base (x,y) = If "Use current position as starting point" is not enabled, this |
||
131 | position will be used to start the calibration |
||
132 | |||
133 | |||
134 | |||
135 | |||
136 | |||
137 | |||
138 | |||
139 | |||
140 | |||
141 | h2. Image Control Panel |
||
142 | |||
143 | |||
144 | This panel controls which image is shown in the Image Display area. |
||
145 | |||
146 | |||
147 | |||
148 | |||
149 | * Image = shows the corrected images as they are acquired |
||
150 | |||
151 | |||
152 | * Correlation = shows the correlation image between sequential images in a |
||
153 | calibration |
||
154 | |||
155 | |||
156 | * Peak = shows an orange crosshair over the Peak that has been identified from the |
||
157 | correlation image |
||
158 | 2 | Amber Herold | |
159 | |||
160 | |||
161 | |||
162 | ______ |
||
163 | |||
164 | [[Matlab Target Finder|< Matlab Target Finder]] | [[Mosaic Click Target Finder|Mosaic Click Target Finder >]] |
||
165 | |||
166 | ______ |