h1. Microscope Set-up Checklist
2
3
4
5
This checklist is valid for users needing the full leginon capacity such as those of MSI application. "Manual" application users can set up the microscope in Low Dose Mode and follow their usual practice. The list in practical order includes alignments that are critical for Leginon operation and are ones Leginon can not perform, yet. Other alignments not mentioned here can often be refined within leginon. Conditions listed in [] show possible ways to check the alignment. In most cases, alignment is only needed if the check fails. * indicates tips for better results
6
7
8
9
* Check that Low Dose Mode is NOT activated
10
11
12
* Adjust the grid to eucentric height using the Alpha Wobbler in HM mode
13
14
15
* Reset to true focus to provide a good starting point for alignment in both HM and LM mode
16
17
18
*At LM mode, true focus can be reached by adjusting the image to the point of contrast reversal with the defocus knob. A halo-type effect is also produced around the image (when the beam radius is small) at defocus values away from true focus.
19
20
21
* Gun Alignments [Bright point at the center of the lightly defocused, large "spot size" beam. Bright and Coherent Beam]
22
23
24
*Load a previous alignment for a better starting point.
25
26
27
ITERATE
28
29
30
* Condensor aperture centering
31
32
33
34
35
* Condensor stigmation
36
37
38
39
40
* Gun tilt
41
42
43
44
45
* Gun shift
46
47
48
49
50
* Gun tilt pivot point (if tilt and shift are hard to optimize)
51
52
53
54
55
* Spot size dependent gun shift (if the beam jumps a lot between spots)
56
57
58
59
60
* HM Beam Alignment (iterate) [image and beam do not move with change of defocus]
61
62
63
*Check eucenter and reset to Eucentric focus again for best result
64
65
66
*Perform this alignment at 200k+ x may increase the accuracy although doing it at the mag where the autofocus is performed (usually 50k x) reduces variation from day to day.
67
68
69
*Do not click on the "Beam Shift" option in the direct alignment page in Tecnai UI. It resets the beam shift value, and do more harm than good if you want to import presets from other days.
70
71
72
ITERATE
73
74
75
* Beam tilt pivot point x
76
77
78
* Beam tilt pivot point y
79
80
81
* Rotation center (i.e. set proper beam tilt)
82
83
84
85
86
* HM Image/Beam Calibration [image does not move when beam is shifted and vice versa]
87
88
89
*Available on FEI Tecnai series only
90
91
92
For Alignment, follow instruction in Alignment/Image HM TEM/"Image/Beam Calibration"
93
94
95
* In Diffraction Mode in HM range, center the objective aperture.
96
97
98
* Objective lens stigmation correction in HM mode [isotropic Thon ring in FFT]
99
100
101
102
103
104
The followings are optional alignment in LM mode. The alignment is needed only if serious problems are found and the alignment in general won't ever converge as nicely as those in HM mode.
105
106
107
108
* LM Beam Alignment (iterate) [image and beam do not move with change of defocus]
109
110
111
*Reset to Eucentric focus again for best result
112
113
114
*Perform this alignment at one of the higher mags or at the mag for "sq" preset in MSI application is recommended
115
116
117
* Beam tilt pivot point x
118
119
120
* Beam tilt pivot point y
121
122
123
* Rotation center (i.e. set proper beam tilt)
124
125
126
127
128
* LM Image/Beam Calibration [image does not move when beam is shifted and vice versa]
129
130
131
*Available on FEI Tecnai series only
132
133
134
For Alignment, follow instruction in Alignment/Image LM/"Image/Beam Calibration"
135
136
137
* Diffraction lens stigmation correction in LM mode [isotropic details at the edge of a hole]