Multi-Scale Imaging (MSI) is the back-bone of many Leginon applications. The setup and
6
calibrations required for an imaging sequence can be understood better when it is compared to
7
a low dose mode on a typical electron microscope such as FEI Tecnai.
8
9
10
11
h2. Low Dose Kit
12
13
14
15
A typical low dose kit contains three modes: Exposure, Focus, and Search
16
17
18
* Exposure Mode for final image acquisition: The image shift at (0, 0) and beam shift
19
to center on the detector. The beam intensity, diameter, and the detector exposure time
20
are matched to provide optimal low-dose exposure that does not irradiate other
21
area.
22
23
24
* Focus Mode for focusing the microscope to the height of the specimen at an area
25
other than that of the final exposure: The image shift is set to non-zeros. The beam
26
shift is compansated so that the beam is centered on the detector. As a result, the
27
image of a remote area is imaged on the detector. For example, FEI Tecnai's Focus mode
28
allows the user to set S1 and S2 focus location by defining the radius of the offset and
29
an angle of rotation for S1 with S2 always 180 degree opposite to S1. The origin of the
30
image offset is at (0,0). Since focusing is normally done at a magnification equal to or
31
higher than Exposure, there is only a very small magnification dependent image
32
shift.
33
34
35
* Search Mode for searching for and centering a potential target. The Search mode is
36
usually set at a low magnification for large field of view and minimal dose. In
37
operation, the user looks at an image in the viewing screen, finds a potential target
38
object, and then moves the specimen stage with the goniometer to locate a potential
39
target by moving it to the center of the viewing screen, which is also the center of the
40
detector. If aligned properly, this will also be the center of the image taken at the
41
Exposure mode.
42
43
44
!http://emg.nysbc.org/software/leginon/images/images/lowdose.png!This figure shows the steps of targeting an object for low dose image collection.
45
First, move the object (gray star) to the center of the viewing screen in Search Mode
46
(yellow circle). If properly aligned, then Focus at the set offset radius and angles,
47
and at last acquire final image at the center in Exposure mode (Red square).
48
49
50
Because of the large difference in magnification, it is often necessary to adjust
51
the image shift of the Search Mode to achieve proper alignment to Exposure mode.
52
53
54
55
56
57
58
59
h2. Setup of the low-dose kit
60
61
62
63
Although not obvious to everyone, the low dose kit does require set up. These
64
include:
65
66
67
* Set the magnification, spot size etc. desired for each mode.
68
69
70
* Align image shift of the Search mode to the Exposure mode so that object at the
71
center of the viewing screen is at the center of the Exposure image.
72
73
74
* Adjust the radius and angle of the two Focus location to optimize their image
75
shift relative to that of the Exposure mode.
76
77
78
* Align the beam for each modes to the optical (detector) axis.
79
80
81
* Adjust the intensity of the beam for proper exposure in each mode.
82
83
84
85
86
87
88
89
!http://emg.nysbc.org/software/leginon/images/images/bad_align_lowdose.png!This figure illustrates the consequence of misalignment of image shift of the lowere
90
magnification Search Mode to the Exposure Mode. The yellow circle represents the field of
91
the view in the search mode, the gray star an object that the user centers as a target for
92
final exposure (red square). When the image shift in the Search mode is not aligned to the
93
exposure, the object centered in the field of the view in Search mode is not centered in
94
the Exposure mode image. Note that in this case, the image shift offset origin of the two
95
Focus modes (black) is such that it is aligned with Exposure mode.
96
97
98
99
100
101
102
103
h2. Leginon MSI analog of the Low Dose kit
104
105
106
107
Leginon separates the function of the low dose kit idea into three parts: defining the
108
presets, choosing Acquisition/Focus targets, and moving to a target on an image.
109
110
111
112
h2. Low-Dose-Kit Equivalent Presets
113
114
115
116
The presets in Leginon can be set up like the three modes in the low dose kit with two
117
exceptions: First, there is only one Focus preset, and second, the focus preset has the
118
same image shift as the Exposure preset.
119
120
121
122
|_.Magnification (example):|_.Preset name:|_.Image Shift (x,y):|
123
|550|Search|Aligned|
124
|100000|Focus|0,0*|
125
|50000|Exposure|0,0|
126
127
128
129
130
*Focus Preset is also at image shift (0,0) since the target selected for it is
131
separated from that for Exposure Preset.
132
133
134
135
136
137
h2. Acquisition and Focuse Targets
138
139
140
141
Leginon users can select more than one type of targets on a single image. For a
142
low-dose-kit analog, we can define one or many Acquisition targets (used for acquiring
143
images at Exposure Preset) on an image acquired in the Search Preset. In addition, we
144
define one Focus target on the same image (for focusing at Focus Preset, of
145
course).
146
147
148
Because the focus targets can be selected independently from the those for the final
149
exposure, the image shift for the Focus preset is (0,0) rather than microns away from the
150
origin. At its current state, Leginon only focus at one position per low mag Search
151
image.
152
153
154
155
156
157
h2. Moving to the Targets
158
159
160
161
When a person operates in low-dose mode, he/she finds a target on the image in the
162
viewing screen and move the stage according to its relative position to the viewing screen
163
to reach the target, i.e., center the target on the screen in Search mode. Leginon moves
164
to the targets using a similar logic. When a target is selected from an image acquired
165
using the Search Preset in Leginon, in order to acquire a new image in Focus or Exposure
166
mode at this target, the program move the target relative to the center of the CCD
167
detector using the knowledge (i.e., calibrations) of the Search Preset at which the old
168
image is acquired, not that of the Focus or Exposure Preset with which the new image will
169
be acquired.
170
171
172
!http://emg.nysbc.org/software/leginon/images/images/MSI_lowdose.png!This
173
four paneled figure shows how Leginon performs a low-dose kit type of image acquisition
174
sequence where the focus target (blue) is first moved to the center of the CCD and then
175
the exposure target(green). Compare this with the one on <link linkend="low_dose">low dose
176
kit</link>.
177
178
179
As a consequence, let's say that we decide to move to the target by moving the stage,
180
we will want the Exposure node to use the following settings: Preset-Exposure; Move
181
Type-Stage. However, the required stage position matrix calibration is at that of the
182
Search Preset.
183
184
185
186
187
188
h2. Setup of the Presets in a low-dose equivalent MSI
189
190
191
192
Set up of the presets is therefore not very different from that for the low dose
193
kit.
194
195
196
* Set up magnification, spot size, beam intensity for each preset.
197
198
199
* Set the image shift (0,0) for Focus and Exposure preset.
200
201
202
* Align the image shift of Search preset to that of Exposure/Focus.
203
204
205
* Align the beam shift for each mode to the optical/detector axis.
206
207
208
209
210
211
212
213
h2. Calibrations required
214
215
216
217
For the above imaging sequence, the calibration requirements with manual focusing and
218
target selection are
219
220
221
* Pixel size calibration for 550x, 50,000x, and 100,000x
222
223
224
* Bright/Dark reference images for the CCD configuration used for each
225
preset.
226
227
228
* Image shift matrix calibration for 550x to perform image shift alignment of Search
229
Preset in Leginon.
230
231
232
* Stage position matrix calibration for 550x to move the target selected.