Project

General

Profile

New User Features 30 » History » Version 2

Anchi Cheng, 12/16/2013 07:01 PM

1 1 Anchi Cheng
h1. New User Features 3.0
2
3
h2. Instrument function:
4
5
# Gatan K2 Summit support
6
# Direct Detector Frame alignment (Appion) (K2 or DE)
7
# Microcondenser mode setting in presets manager
8
9
h2. Target Navigation:
10
11
# Option of using parent image move type to move to target during target adjustment.
12
** This improves the targeting accuracy and minimize image-shift-induced beam-tilt in image-shifted final exposure if queuing is used in Exposure Targeting and Navigator iterative move is used in acquiring its parent image.
13
** Activate it at /Leginon/Exposure/Settings/Advanced Settings:  Use ancestor image mover in adjustment
14
15
h2. Automatic Target Finding:
16
17
# Limit number of focusing performed on a series of images.
18
** For example, focus once every 5 hl images that processed by Exposure Targeting node.
19
** Iterator is reset at new parent so that hl image from a new square will always be focused.
20
** Activate at /Leginon/Exposure Targeting/Settings/Advanced Settings/Hole Finder(or Raster Finder) Settings: Focus every nnn image.  Default=1 meaning no skipping.
21
# Option to threshold very low intensity part of the image (grid bar or thick contaminant before template correlation in JAHCFinder used in MSI-T application
22
** This has the effect of reducing false-positive in template-based hole finding.
23
** Enter the threshold value at /Leginon/Exposure(or Hole) Targeting/Template Correlation Settings: Fill image values below n with mean before correlation.
24
25
h2. Focusing
26
27
# Repeat same beam-tilt-based focusing step with the exactly same setting except not to drift checking will improve the focusing when the first adjustment is large
28
** The first focusing step brings the defocus close to the range ideal for the algorithm for the repeat.
29
** A bug fix makes this approach reliable and is recommended if defocus is really critical and speed of data acquisition does not out-weight it.
30
31
h2.  RCT Acquisition:
32
33
# Tilt Test Tool !http://emg.nysbc.org/software/leginon/images/icons/alpha.png!
34
** Usage: For testing if the grid bars will block the intended targets when tilted.
35
**# Send the preset using for defining the RCT targets to scope with Presets Manager.
36
**# Each click of the Tilt Test Tool will take the goniometer to the next tilt defined in RCT settings and acquire an image.