New User Features 30 » History » Version 2
Anchi Cheng, 12/16/2013 07:01 PM
| 1 | 1 | Anchi Cheng | h1. New User Features 3.0 |
|---|---|---|---|
| 2 | |||
| 3 | h2. Instrument function: |
||
| 4 | |||
| 5 | # Gatan K2 Summit support |
||
| 6 | # Direct Detector Frame alignment (Appion) (K2 or DE) |
||
| 7 | # Microcondenser mode setting in presets manager |
||
| 8 | |||
| 9 | h2. Target Navigation: |
||
| 10 | |||
| 11 | # Option of using parent image move type to move to target during target adjustment. |
||
| 12 | ** This improves the targeting accuracy and minimize image-shift-induced beam-tilt in image-shifted final exposure if queuing is used in Exposure Targeting and Navigator iterative move is used in acquiring its parent image. |
||
| 13 | ** Activate it at /Leginon/Exposure/Settings/Advanced Settings: Use ancestor image mover in adjustment |
||
| 14 | |||
| 15 | h2. Automatic Target Finding: |
||
| 16 | |||
| 17 | # Limit number of focusing performed on a series of images. |
||
| 18 | ** For example, focus once every 5 hl images that processed by Exposure Targeting node. |
||
| 19 | ** Iterator is reset at new parent so that hl image from a new square will always be focused. |
||
| 20 | ** Activate at /Leginon/Exposure Targeting/Settings/Advanced Settings/Hole Finder(or Raster Finder) Settings: Focus every nnn image. Default=1 meaning no skipping. |
||
| 21 | # Option to threshold very low intensity part of the image (grid bar or thick contaminant before template correlation in JAHCFinder used in MSI-T application |
||
| 22 | ** This has the effect of reducing false-positive in template-based hole finding. |
||
| 23 | ** Enter the threshold value at /Leginon/Exposure(or Hole) Targeting/Template Correlation Settings: Fill image values below n with mean before correlation. |
||
| 24 | |||
| 25 | h2. Focusing |
||
| 26 | |||
| 27 | # Repeat same beam-tilt-based focusing step with the exactly same setting except not to drift checking will improve the focusing when the first adjustment is large |
||
| 28 | ** The first focusing step brings the defocus close to the range ideal for the algorithm for the repeat. |
||
| 29 | ** A bug fix makes this approach reliable and is recommended if defocus is really critical and speed of data acquisition does not out-weight it. |
||
| 30 | |||
| 31 | h2. RCT Acquisition: |
||
| 32 | |||
| 33 | # Tilt Test Tool !http://emg.nysbc.org/software/leginon/images/icons/alpha.png! |
||
| 34 | ** Usage: For testing if the grid bars will block the intended targets when tilted. |
||