Operation » History » Version 7
Anchi Cheng, 07/15/2010 06:32 PM
1 | 1 | Amber Herold | h1. Operation |
---|---|---|---|
2 | |||
3 | |||
4 | |||
5 | h2. Enter Grid Information through Project Web Form |
||
6 | |||
7 | 4 | Anchi Cheng | # Go to http://your_host/myamiweb/project/project.php on a web browser. |
8 | # Select the Grid Tray Panel. |
||
9 | # Click on upload grids/tray |
||
10 | 1 | Amber Herold | |
11 | * Follow the format and instruction to create an upload file with grid information with labels unique to the project and location unique to the grid box to be inserted. |
||
12 | |||
13 | h2. Set up robot and grids in its tray matching your entry |
||
14 | |||
15 | 4 | Anchi Cheng | *Leginon does not know if you switch grids and pretend!* |
16 | 1 | Amber Herold | |
17 | h2. Start Leginon |
||
18 | |||
19 | * This include general scope alignment and preset alignement |
||
20 | |||
21 | 7 | Anchi Cheng | h2. [[1st Pass application|1st Pass Screening]] |
22 | 1 | Amber Herold | |
23 | 4 | Anchi Cheng | # Starting node is the Robot node |
24 | 5 | Anchi Cheng | # Enter in the settings the stage z value that will give the approximate eucentric height for the holder used. |
25 | 4 | Anchi Cheng | # Select the robot tray |
26 | # Select grids in the Robot node and click "Start". You might need to click "Continue". too. It is a bug that is hard to remove without causing other problems. Right click removes a selection. Middle click displays the selected grid label |
||
27 | *The grids are imaged in the order of the grid selection.* |
||
28 | # Activate user verification in "Square Targeting" and "Mid Mag Survey Targeting" nodes |
||
29 | # Use the first grid to optimize square finding and raster sizing and angles and then remove the user verification for full automation. |
||
30 | 1 | Amber Herold | |
31 | 4 | Anchi Cheng | * For "Square Targeting", optimize to find all grid squares with intact support film and proper stain. |
32 | * "Square Target Filtering" is default to find the center-most of the selected targets pass into it. |
||
33 | * "Mid Mag Survey Targeting" is best set to spread out the targeting position. |
||
34 | 1 | Amber Herold | |
35 | 7 | Anchi Cheng | h2. [[Evaluation application|Evaluation]] |
36 | 1 | Amber Herold | |
37 | 4 | Anchi Cheng | # Start the application under the same Leginon session. Microscope and robot not required by this application. |
38 | # Choose "hl" or "sq" preset as the child and "gr" preset as ancestor. |
||
39 | # Click next to get the first image. |
||
40 | # Choose targets and then click on "Transform" tool. |
||
41 | *Do not forget to transform if you modify the targets.* |
||
42 | # Continue to "next" image until at the end. |
||
43 | # Select targets at the other child preset, too. |
||
44 | 2 | Amber Herold | |
45 | 7 | Anchi Cheng | h2. [[2nd_Pass_application|2nd Pass Screening]] |
46 | 2 | Amber Herold | |
47 | 4 | Anchi Cheng | * Starting the sequence at "2nd Pass Targeting" under the same session with mircoscope and robot ready. |
48 | * If no modification is needed after evaluation step, just "Submit" the targets to initiate the process which will pause to wait for confirmation that the robot is ready to start. |
||
49 | 2 | Amber Herold | |
50 | |||
51 | ______ |
||
52 | 1 | Amber Herold | |
53 | [[2nd Pass application|< 2nd Pass application]] |
||