h1. Parameters needed 

 Need to enter: 
 Checkbox whether to do it at all or not. 

 Slit Width: Suggest to use 15 eV, since calibration was performed at that width. Generally we collect at 20 eV, but there is no problem to switch to 15 eV for the measurement 

 Exposure time: Default is 500 ms, which gives enough signal for the method to work. Can experiment with faster times if desired. 

 Mean free path: We measured mean free path to be 395 nm at NYSBC at 300 kV. It should depend only on the voltage. 

 Collect every N images: Since it adds overhead, you may want to do it only every 5 or 10 images. 

 h1. Available in 3.4 version: Beta ONLY: 

 h1.    Objective Scattering Thickness estimate 

 Need to enter 

 Checkbox whether to calculate it or not 
 Exposure Counts over vacuum 
 Mean free path for objective scattering: depends on instrument and aperture diameter. 
 * 300 is a good intitial guess if unknown  
 Values determined at NYSBC: 
 * titan Krios 300 keV 20 eV slit, 100 um objective aperture: 322 nm 
 * titan Krios, 300 keV no EF, 100 um objective aperture: 3329 nm 
 * Tecnai F20 200 keV 100 um objective aperture: 392 nm 
 * Tecnai F20 200 keV 70 um objective aperture: 302 nm 
 * T12    120 keV 100 um objective aperture: 319 nm 
 * T12 120 keV 70 um objective aperture: 247 nm