Need to enter:
4
Checkbox whether to do it at all or not.
5
6
Slit Width: Suggest to use 15 eV, since calibration was performed at that width. Generally we collect at 20 eV, but there is no problem to switch to 15 eV for the measurement
7
8
Exposure time: Default is 500 ms, which gives enough signal for the method to work. Can experiment with faster times if desired.
9
10
Mean free path: We measured mean free path to be 395 nm at NYSBC at 300 kV. It should depend only on the voltage.
11
12
Collect every N images: Since it adds overhead, you may want to do it only every 5 or 10 images.
13 2 William Rice
14 4 Anchi Cheng
h1. Available in 3.4 version:
15 3 William Rice
16 2 William Rice
h1.  Objective Scattering Thickness estimate
17
18
Need to enter
19
20
Checkbox whether to calculate it or not
21
Exposure Counts over vacuum
22 5 Anchi Cheng
Mean free path for objective scattering (also known as ALS coefficient: depends on instrument and aperture diameter.
23 2 William Rice
* 300 is a good intitial guess if unknown 
24
Values determined at NYSBC:
25
* titan Krios 300 keV 20 eV slit, 100 um objective aperture: 322 nm
26
* titan Krios, 300 keV no EF, 100 um objective aperture: 3329 nm
27
* Tecnai F20 200 keV 100 um objective aperture: 392 nm
28
* Tecnai F20 200 keV 70 um objective aperture: 302 nm
29
* T12  120 keV 100 um objective aperture: 319 nm
30
* T12 120 keV 70 um objective aperture: 247 nm