h2. Phase Plate Aligner Tool Bar
2
3
!phaseplatetools.png!
4
5
From left to right:
6
7
h3. Settings
8
9
Required setting change before usage:
10
* Deactivate "Bypass Conditioner" to activate this node to respond to request.
11
* Enter current phase plate slot (1-6) and Current Phase Patch Position, and click "Update" to record it.
12
13
h3. Run Test
14
15
It goes through all the steps that it does involving moving to the next phase plate patch position.
16
17
# Enter current phase plate slot and patch position in the settings.
18
# Make sure your have a valid reference target in this session, or manually move the stage to an open area.
19
# Send the on-plane condition preset you want to run the position advance test to scope with Presets Manager.
20
# Left click this tool
21
22
h3. Stop Test
23
24
This does nothing now.
25
26
h3. Acquire and Save Current Position as Reference
27
28
It saves an image as parent image with its center as reference target.  It makes it easier to choose reference at higher magnification and allow navigator iterated move to be applied.
29
30
# Move manually to the position that you would like to use.
31
# Send a medium mag preset (such as sq or hl) that is centered at this position.
32
# Left click this tool.
33
34
h3. Reset Counter
35
36
Reset counting without moving to the next patch.
37
38
h3. Patch State Mapping
39
40
Some area of the phase plate slot may not be usable. This allows them to be skipped during automated advancing.
41
42
# Enter the phase plate slot number you would like to map as current phase plate slot in the settings.
43
# Left clieck this tool.
44
# Choose and set to "checked" each position you would like to skip.
45
# *Important* Due to a gui bug, make sure you choose a position other than your last that you have changed state before clicking "OK" and exit the dialog.