Pre-MSI Set-up » History » Version 28
Anchi Cheng, 05/03/2017 02:15 AM
1 | 1 | Amber Herold | h1. Pre-MSI Set-up |
---|---|---|---|
2 | |||
3 | |||
4 | |||
5 | h2. Design Presets |
||
6 | |||
7 | |||
8 | |||
9 | 5 | Amber Herold | To start your own MSI experiment, you must decide what presets you want to use. Presets are used to define scope and camera parameters with which images are acquired. |
10 | 1 | Amber Herold | |
11 | 5 | Amber Herold | MSI presets can be customized to your need of data collection. By matching nodes/subnodes in MSI with presets and move types, the behavior of the node is |
12 | 1 | Amber Herold | defined. |
13 | |||
14 | 28 | Anchi Cheng | For easy reference and for preloading node settings in the our distribution, preset names in MSI has been standardized at NRAMM. You can use other names when you become advanced user and are comfortable changing settings. They are abbreviated to 2 letter codes to reduce the length of the filenames that contains every preset used in its family history. |
15 | 1 | Amber Herold | |
16 | 4 | Amber Herold | Example MSI preset nomenclature: |
17 | 1 | Amber Herold | |
18 | |_.Preset name:|_.abbrev for:|_.in the context of:| |
||
19 | |gr|grid|em grid| |
||
20 | |sq|square|em grid square| |
||
21 | |hl|hole|quantifoil hole| |
||
22 | |fc|focus|focus image to be checked with fft| |
||
23 | |fa|focus-auto|automatic focusing| |
||
24 | |en|exposure-near|close-to-zero defocus exposure image in a focal pair| |
||
25 | |ef|exposure-far|far-from-zero defocus exposure image in a focal pair| |
||
26 | |||
27 | |||
28 | |||
29 | 8 | Anchi Cheng | Example MSI preset parameters (based on 120 or 200 kV high tension, 100 um C2 aperture and 100 um objective aperture on a 4k ccd with pixel size at 1.6 angstrom at scope nominal magnification of 50,000x): |
30 | 1 | Amber Herold | |
31 | |_.Magnification:|_.Preset name:|_.Image Shift (x,y):|_.Dimension:|_.Binning:|_.Beam Coverage:|_.Exposure Time (ms):|_.Spot Size:|_.Defocus (m):| |
||
32 | |120|gr|Aligned|512|8|max|20|4|0.0| |
||
33 | |550|sq|Aligned|1024|4|1x CCD size|100|4|-2e-3| |
||
34 | |5000|hl|Aligned|512|8|1x CCD|20|4|-1.5e-4| |
||
35 | |50000|fc|0,0|512|1|<~ 1x CCD|300|4|-2e-6| |
||
36 | |50000|fa|0,0|1024|4|>2x CCD|50|4|-2e-6| |
||
37 | |50000|en|0,0|4096|1|2x CCD|170 (10e/A^2)|4|-1e-6| |
||
38 | |50000|ef|0,0|4096|1|2x CCD|170 (10e/A^2)|4|-2e-6| |
||
39 | |||
40 | |||
41 | |||
42 | The preset parameters in this example are chosen to give the following properties: |
||
43 | |||
44 | |||
45 | |||
46 | |||
47 | 7 | Anchi Cheng | * *en* and *ef* are for final exposures and therefore have highest resolution and dimension. |
48 | 1 | Amber Herold | |
49 | |||
50 | 5 | Amber Herold | * *en* stands for near focus exposure and therefore has lower defocus value than *ef*,the far-from-focus exposure. |
51 | 1 | Amber Herold | |
52 | |||
53 | 5 | Amber Herold | * *fa* is for high resolution autofocus and therefore need to be high mag. The dimension and binning is a result of compromising speed, S/N ratio and sensitivity to |
54 | 1 | Amber Herold | details. |
55 | |||
56 | |||
57 | 5 | Amber Herold | * *fc* is for manual focus and ice melting and therefore more intense than *fa*. To get good FFT for checking Thon ring behavior, it is not binned. The small dimension is for increasing speed. |
58 | 1 | Amber Herold | |
59 | |||
60 | 5 | Amber Herold | * *hl* is for intermediate targeting. It needs to cover the error range of modeled stage position to allow final targeting by image shift only, hence the magnification. The binning is for speed. |
61 | 4 | Amber Herold | |
62 | 1 | Amber Herold | |
63 | 8 | Anchi Cheng | * *sq* is for grid square targeting. The image produced by the preset ideally need to cover most if not all of the square but with enough resolution to evaluate the content. However, the first priority is to *choose a magnification at which no aperture used during the data collection will limit much of the illuminated area on the CCD*. If too much of the area is obstructed, target adjustment critical to MSI operation will fail. The coverage of the grid square can always be achieved by choosing multiple targets on it. |
64 | 1 | Amber Herold | |
65 | |||
66 | 6 | Anchi Cheng | * *gr* is for producing grid atlas. It is selected based on the size number of tiles needed to cover the whole grid in an acceptable time and the amount of distortion common to very low mag imaging. Objective aperture can be removed and a larger C2 aperture used while using this preset to acquire grid atlas. It is never used after the atlas is acquired. |
67 | 1 | Amber Herold | |
68 | |||
69 | |||
70 | |||
71 | |||
72 | In designing your own preset, follow the above properties using a mag that gives the required coverage with your scope, camera, and grid mesh. If you have a 2k or 1k camera, binning may not be as necessary as for a 4k camera whose data acquisition time is 10-30 sec without binning. |
||
73 | |||
74 | |||
75 | 12 | Anchi Cheng | h2. Another Recent Preset Design Example (Rotavirus 4.4 A resolution, Cambell et. al., _Structure_(2012)) |
76 | 1 | Amber Herold | |
77 | 8 | Anchi Cheng | Tecnai F20 200 kV high tension, Gun Lens 3 extraction voltage 4300, 50 um C2 aperture and 100 um objective aperture (except when obtaining grid atlas). |
78 | 12 | Anchi Cheng | Camera is a Tietz F415 4k ccd with pixel size at 1.6 angstrom at scope nominal magnification of 50,000x for lower mag MSI and focusing, and a Direct Electron DE-12 4kx3k direct detection camera for final exposure (ed preset): |
79 | 1 | Amber Herold | |
80 | 10 | Anchi Cheng | |_.Magnification:|_.Preset name:|_.Image Shift (x,y):|_.Dimension:|_.Binning:|_.Beam Coverage:|_.Exposure Time (ms):|_.Dose (e/A^2):|_.Spot Size:|_.Defocus (m):|_.C2 (um):| |
81 | 8 | Anchi Cheng | |120|gr|Aligned|512|8|max|20|n/a|4|0.0|100| |
82 | |1700|sq|Aligned|1024|4|1x CCD size|10|n/a|4|-2e-4|50| |
||
83 | |5000|hl|Aligned|1024|4|1x CCD|7|0.004|4|-1.5e-4|50| |
||
84 | |100000|fc|0,0|512|2|<~ 1x CCD|200|164|4|-2.9e-7|50| |
||
85 | |100000|fa|0,0|1024|4|~4x CCD|50|2.5|4|-1.5e-6|50| |
||
86 | |100000|en|0,0|4096|1|1.3 um on the specimen|407|20|4|(-1e-6 to -2e-6)|50| |
||
87 | |29000|ed|0,0|3kx4k|1|1.3 um on the specimen|407|20|4|(-1e-6 to -2e-6)|50| |
||
88 | 1 | Amber Herold | |
89 | * Note: This experiment used direct detection camera for final exposure (ed preset). en preset was not used in the actual experiment. It is here just for reference and to show what we would have used if recorded on CCD based on our GroEL study. During process, the final images for en would have been binned by 2 based on our experience so the en preset could have being collected at 2048 binned by 2. |
||
90 | 12 | Anchi Cheng | |
91 | * Note: no defocal pair was aquired in this experiment. Therefore, no ef preset. |
||
92 | |||
93 | 24 | Anchi Cheng | h2. Preset Design Example for Falcon II on Krios |
94 | |||
95 | Tecnai Titan Krios 300 kV high tension, Gun Lens 3 extraction voltage 4150, 70 um C2 aperture and 100 um objective aperture (except when obtaining grid atlas). |
||
96 | 25 | Anchi Cheng | Falcon II mounted with Gatan Orius as the secondary camera. Pixel size at 1.4 angstrom at scope nominal magnification of 59,000x (ed preset). |
97 | The c-flat grid used in this has 1.2 um holes. One image is taken at each hole. |
||
98 | 1 | Amber Herold | |
99 | 25 | Anchi Cheng | |_.Magnification:|_.Preset name:|_.Camera:|_.Image Shift (x,y):|_.Dimension:|_.Binning:|_.Beam Coverage:|_.Exposure Time (ms):|_.Dose (e/A^2):|_.Spot Size:|_.Defocus (m):|_.C2 (um):| |
100 | |40|gr|Orius|Aligned|1024|4|max|100|n/a|10|0.0|70| |
||
101 | |1700|sq|Falcon|Aligned|1024|4|1x CCD size|100|n/a|10|-5e-4|70| |
||
102 | |3800|hl|Falcon|Aligned|1024|4|12 um|300|n/a|10|-2e-4|70| |
||
103 | |45000|fm|Orius|0,0|1024|2|small|500|100|7|-2.9e-7|70| |
||
104 | |75000|fc|Orius|0,0|1024|2|0.5 um on the specimen|500|164|6|-2.9e-7|70| |
||
105 | |75000|fa|Falcon|0,0|1024|4|1.2 um on the specimen|200|2.5|7|-1.5e-6|70| |
||
106 | |59000|ed|Falcon|0,0|4kx4k|1|1.3 um on the specimen|2000|45|6|(-1e-6 to -2e-6)|70| |
||
107 | |||
108 | * fm is the melting preset. No image is taken with this. The preset is used only to set microscope parameters. The camera is set to Orius as extra protection. |
||
109 | * Frames are saved in ed preset with (1,2,17) set in "Frames to use" field, giving 22.5 e/A^2 total dose in the movie that is used for image processing and volume reconstruction. |
||
110 | 24 | Anchi Cheng | |
111 | 17 | Anchi Cheng | h2. Preset Design Example for small pixel Gatan K2 camera alone |
112 | 12 | Anchi Cheng | |
113 | 27 | Anchi Cheng | {{include(Krios-K2-TSRI_Presets)}} |
114 | 3 | Alex Kadokura | |
115 | 26 | Anchi Cheng | {{include(J2100f_k2_preset_design)}} |
116 | |||
117 | 1 | Amber Herold | h2. Calibrations |
118 | |||
119 | |||
120 | 5 | Amber Herold | If this is the first time calibrations have ever been completed for Leginon, complete all the calibrations listed in [[Leginon "Calibrations" Application|the chapter on calibrations]]. Calibrations are extremely important to operate the TEM in a consistent manner. As long as the calibrations are stable, later users do not need to perform them. Most calibrations are HT and magnification dependent. Therefore, new calibration may be required if your presets use different HT and/or magnification from all previous users. |
121 | 2 | Amber Herold | |
122 | |||
123 | ______ |
||
124 | |||
125 | [[Summary of MSI applications|< Summary of MSI applications]] | [[Initial MSI application preferences|Initial MSI application preferences >]] |
||
126 | |||
127 | ______ |