Presets for MSI-RCT applications » History » Revision 4
Revision 3 (Amber Herold, 06/21/2010 02:59 PM) → Revision 4/7 (Amber Herold, 06/21/2010 02:59 PM)
h1. Presets for MSI-RCT applications The main difference between MSI-RCT applications and other MSI's are that a low magnification image is needed for the target transformation algorithm to reliably find features to match between tilts. Two strategies has be used with success. h2. 1st strategy: The sq preset is set at a magnification in the HM mode so that hysteresis is minimized. and the hl preset is eliminated The "sq" preset is used both at "Square" and "Centered Square" nodes. *Presets and Magnification for a typical MSI-RCT* |_.Preset|_.magnification|_.defocus (m)| |gr|120|0| |sq|1700|-2.5e-4| |fc,fa,en|50000|(what you need)| h2. 2nd strategy: The sq preset is set at the usual LM magnification. The hl preset is set at the low end of HM mode. The "sq" preset is at "Square" node and "hl" preset is used at "Centered Square" nodes. *Presets and Magnification for a typical MSI-RCT* |_.Preset|_.magnification|_.defocus (m)| |gr|120|0| |sq|550|-2e-3| |hl|1700|-2.5e-4| |fc,fa,en|50000|(what you need)| ______ [[RCT node set-up|< RCT node set-up]] ______