Queuing Example 3 - Both Exposure and Hole Targeting » History » Version 1
Amber Herold, 04/22/2010 05:07 PM
1 | 1 | Amber Herold | h1. Queuing Example 3 - Both Exposure and Hole Targeting |
---|---|---|---|
2 | |||
3 | |||
4 | |||
5 | |||
6 | MSI-Tomography application is an example of such queuing set up. It is also the useful |
||
7 | when energy filter is used. It is chosen so that the microscope stays at the same presets as |
||
8 | long as possible to obtain all the images so that LM preset sq will not need to be used once |
||
9 | tomography data collection starts. |
||
10 | |||
11 | |||
12 | |||
13 | h2. Configuration |
||
14 | |||
15 | |||
16 | |||
17 | |||
18 | |||
19 | * /Square/Toolbar/Setup/Wait for a node to process image = no |
||
20 | |||
21 | |||
22 | The "no" setting saves time by acquiring images while holefinder and YOU are picking |
||
23 | targets. |
||
24 | |||
25 | |||
26 | * /Hole Targeting/Toolbar/Setup/Queue up targets=yes. |
||
27 | |||
28 | |||
29 | * /Hole Targeting/Toolbar/Setup/Allow user verification of the selected targets=yes, |
||
30 | if desired. |
||
31 | |||
32 | |||
33 | * /Hole/Toolbar/Setup/Wait for a node to process image = no. |
||
34 | |||
35 | |||
36 | * /Exposure Targeting/Toolbar/Setup/Queue up targets=yes. |
||
37 | |||
38 | |||
39 | * /Exposure Targeting/Toolbar/Setup/Allow user verification of the selected |
||
40 | targets=yes, if desired. |
||
41 | |||
42 | |||
43 | |||
44 | |||
45 | |||
46 | |||
47 | |||
48 | h2. Operation |
||
49 | |||
50 | |||
51 | |||
52 | |||
53 | |||
54 | * Select square targets on grid atlas and submit as usual. |
||
55 | |||
56 | |||
57 | * All sq images will be acquired while "Hole Targeting" finds the targets on each sq |
||
58 | image since waiting in "Square" node is off. Note that since the square node acquires |
||
59 | images continuously, both hole finding and acquisition may compete for processing and |
||
60 | memory time. If it affects your interaction with the program, just wait for the |
||
61 | acquisition to complete. |
||
62 | |||
63 | |||
64 | * If you have just acquired many (such as 50) sq images in LM mode, the microscope |
||
65 | has stayed in LM for a long time. It is advisable to set the instrument to one of the |
||
66 | preset in HM mode as soon as the acquisitions are completed and check for possible image |
||
67 | shift inconsistency between presets. If energy filter need to be on for higher |
||
68 | magnification imaging, do so now. |
||
69 | |||
70 | |||
71 | * If user verification is turned on in "Hole Targeting", the targets should be |
||
72 | submitted using "submit"<inlinegraphic |
||
73 | fileref="http://emg.nysbc.org/software/leginon/images/icons/play.png" format="PNG"/> |
||
74 | tool. |
||
75 | |||
76 | |||
77 | * The Queue is submitted and processed only when "Submit Queued Target"<inlinegraphic |
||
78 | fileref="http://emg.nysbc.org/software/leginon/images/icons/send_queue_out.png" |
||
79 | format="PNG"/> tool is clicked. |
||
80 | |||
81 | |||
82 | * Repeat the same target selection and submission process at "Exposure Targeting" |
||
83 | node. |
||
84 | |||
85 | |||
86 | * In case when leginon crashes during queue processing, the data acquisition should |
||
87 | be resumed by "Submit Queued Target"<inlinegraphic |
||
88 | fileref="http://emg.nysbc.org/software/leginon/images/icons/send_queue_out.png" |
||
89 | format="PNG"/> in "Exposure Targeting" node first. If nothing happens, it means the |
||
90 | queue in that node is empty. In that case, repeat it in "Hole Targeting" to |
||
91 | resume. |
||
92 | |||
93 | |||
94 | |||
95 | |||
96 | |||
97 | |||
98 | |||
99 | h2. Interruption |
||
100 | |||
101 | |||
102 | |||
103 | Once the queue is submitted there are several modes of interruption |
||
104 | |||
105 | |||
106 | * If leginon crashes during queue processing, the data queue acquisition can be |
||
107 | resumed by "Submit Queued Target" in first the "Exposure Targeting Q" node, then if |
||
108 | nothing happens, in the "Hole Targeting" node. |
||
109 | |||
110 | |||
111 | * Queued targets from "Hole Targeting" can be paused in "Hole" node with "pause" |
||
112 | button and is <link linkend="MSI_pause">recommanded for the liquid nitrogen refill of |
||
113 | the side-entry cryo stage</link>. |
||
114 | |||
115 | |||
116 | * The "Abort" button<inlinegraphic |
||
117 | fileref="http://emg.nysbc.org/software/leginon/images/icons/stop.png" format="PNG"/> |
||
118 | in "Hole" node aborts the acquisition of the remaining targets from the same parent "hl" |
||
119 | image and proceeds to process targets from next square in the queue. |
||
120 | |||
121 | |||
122 | * The "Abort Queue" button<inlinegraphic |
||
123 | fileref="http://emg.nysbc.org/software/leginon/images/icons/stop_queue.png" |
||
124 | format="PNG"/> in "Hole" node aborts all remaining targets in the queue. |
||
125 | |||
126 | |||