RCT calibration » History » Version 3
Neil Voss, 08/13/2012 06:02 PM
1 | 1 | Anchi Cheng | h1. RCT application calibration |
---|---|---|---|
2 | |||
3 | RCT type of data collection need the normal MSI calibration. In addition, the followings are also required. |
||
4 | |||
5 | 2 | Anchi Cheng | * image shift matrix and modeled-stage calibration (scale and rotation only) at the 1700x preset used for select RCT targets |
6 | 1 | Anchi Cheng | * recently saved rotation center in the database (You can do so in "Beam Tilt" node in "Calibration" application. |
7 | |||
8 | |||
9 | 3 | Neil Voss | h1. Dump from other wiki |
10 | |||
11 | h2. Setup |
||
12 | |||
13 | h3. Days before |
||
14 | |||
15 | * Pump down stage |
||
16 | * Cryo-cycle for 4 hours (240 min) |
||
17 | * Prepare grids |
||
18 | |||
19 | h3. Scope alignments |
||
20 | |||
21 | 2.2.1 no sample |
||
22 | |||
23 | gun shift/tilt |
||
24 | condenser aperture |
||
25 | condenser stig. |
||
26 | check if cold trap is in |
||
27 | adjust eye pieces |
||
28 | |||
29 | 2.2.2 with sample |
||
30 | |||
31 | z-height |
||
32 | focus both M and SA modes |
||
33 | objective stig. |
||
34 | objective aperture at hi mag, use focus to shrink beam, not intensity |
||
35 | at zero defocus |
||
36 | beam tilt pivot points X,Y |
||
37 | rotation center |
||
38 | |||
39 | h3. Presets and Magnifications |
||
40 | |||
41 | preset ↓ mag ↓ ccd ↓ defocus ↓ notes ↓ |
||
42 | gr 120X 512x8 0 do not cycle through |
||
43 | sq 1700X 1024x4 -3.0e-4 for square_centering |
||
44 | cs 1700X 1024x4 -3.0e-4 for rct_targeting, copy of sq |
||
45 | hl 5000X 512x8 -5.0e-5 image shift alignment and z-focus |
||
46 | en 50kX-100kX 4096x1 -1.5e-6 -- -3.5e-6 |
||
47 | fa 50kX-100kX 512x4 -1.5e-6 beam focus and drift checks |
||
48 | |||
49 | h3. Leginon |
||
50 | |||
51 | 2.4.1 pre-Altas |
||
52 | |||
53 | beam shift/intensity for each preset |
||
54 | align image shifts btw presets |
||
55 | matrix calibrate, stage position for gr preset |
||
56 | optional: fine tune rotation center in Leginon, available in any focus node using the fa preset |
||
57 | on rare occasions: check optical axis using Leginon tool, this is only required if the square is shifting alot when you tilt the stage |
||
58 | |||
59 | 2.4.2 post-Atlas |
||
60 | |||
61 | matrix calibrate, stage position for sq preset |
||
62 | error ~50% |
||
63 | shift ~25% |
||
64 | num position ~2-3 |
||
65 | interval ~1-3e-6 |
||
66 | RCT node calibration |
||
67 | make sure the number of features is between 700 and 1500 |
||
68 | take a picture both with and without the grid bar |
||
69 | take a picture both untilted and tilted |
||
70 | calibrate dose |
||
71 | RCT node setup |
||
72 | re-align presets |
||
73 | re-center objective aperture |
||
74 | |||
75 | h2. Running Leginon RCT |
||
76 | |||
77 | Start a square: |
||
78 | Square Targeting (if you have an atlas) |
||
79 | Go to "Square Targeting" |
||
80 | Hit the reload button |
||
81 | Select a new square with the green cross |
||
82 | Click the play |> button |
||
83 | Simulate Square (if you do NOT have an atlas) |
||
84 | Go to "Square" |
||
85 | Hit the simulate target button |
||
86 | Square centering |
||
87 | When Leginon is finished, a question mark will appear next to the "Square Centering" node |
||
88 | Click on the "Square Centering" node |
||
89 | Select one z-focus position (blue cross) and one imaging position (green position) close to the center of the square |
||
90 | Click the play |> button |
||
91 | RCT Targeting |
||
92 | When Leginon is finished setting the z-height, a question mark will appear next to the "RCT Targeting" node |
||
93 | Click on the "RCT Targeting" node |
||
94 | Now you select all the holes to image (not too close to the edge) and a single focus position |
||
95 | Optional: obtain a picture of the tilted square |
||
96 | Click on the "Navigation" node |
||
97 | Open up VNC viewer of the tecnai |
||
98 | In the VNC window, set the tilt angle to the desire angle |
||
99 | In the "Navigation" node, collect an image by clicking on the acquire button (camera logo) at the top |
||
100 | Important, in the VNC window, return the tilt angle to zero degrees, otherwise all your images will fail |
||
101 | Now select you targets in the "RCT Targeting" node and view the tilted image in the "Navigation" node. Jump back and forth between the two nodes until you select all holes present in both images. |
||
102 | When finished, hit the play |> button |
||
103 | Focusing |
||
104 | If manual focusing is enabled, wait for the manual focus window to pop up |
||
105 | focus the microscope, reset the defocus, and click the red stop button. |
||
106 | Watch your images in the "RCT" node |
||
107 | |||
108 | h2. Nitrogen fill |
||
109 | |||
110 | Since RCT only focuses once per square, when you fill the dewar, you must refocus to account for changes in the z-height and focus during refill and there is no check for drift calm down so you must manually wait for drift to subside. |
||
111 | |||
112 | Steps in Nitrogen fill: |
||
113 | |||
114 | Pause RCT node |
||
115 | Wait until Leginon is paused |
||
116 | Fill holder and cold trap dewars |
||
117 | Turn off lights |
||
118 | Check if objective aperture is inserted |
||
119 | RESET TIMER |
||
120 | Focus |
||
121 | simulate z-focus |
||
122 | send hole preset to scope |
||
123 | navigate to good focus spot in navigator |
||
124 | simulate focus in the focus node |
||
125 | repeat focus until drift subsides |
||
126 | If there is an empty hole: you can measure dose or collect bright fields |
||
127 | Declare drift |
||
128 | Hit play button on RCT node |
||
129 | Check images to make sure they are good |
||
130 | Uncheck user verification of targets |
||
131 | |||
132 | This page was last modified 16:45, 16 July 2010. |
||
133 | |||
134 | |||
135 | |||
136 | 1 | Anchi Cheng | |
137 | ______ |
||
138 | |||
139 | [[Presets for MSI-RCT applications|< Presets for MSI-RCT applications]] |
||