RCT calibration » History » Version 3

Neil Voss, 08/13/2012 06:02 PM

1 1 Anchi Cheng
h1. RCT application calibration
2
3
RCT type of data collection need the normal MSI calibration.  In addition, the followings are also required.
4
5 2 Anchi Cheng
* image shift matrix and modeled-stage calibration (scale and rotation only) at the 1700x preset used for select RCT targets
6 1 Anchi Cheng
* recently saved rotation center in the database (You can do so in "Beam Tilt" node in "Calibration" application.
7
8
9 3 Neil Voss
h1. Dump from other wiki
10
11
h2. Setup
12
13
h3. Days before
14
15
* Pump down stage
16
*   Cryo-cycle for 4 hours (240 min)
17
*   Prepare grids 
18
19
h3. Scope alignments
20
21
2.2.1 no sample
22
23
    gun shift/tilt
24
    condenser aperture
25
    condenser stig.
26
    check if cold trap is in
27
    adjust eye pieces 
28
29
2.2.2 with sample
30
31
    z-height
32
    focus both M and SA modes
33
    objective stig.
34
    objective aperture at hi mag, use focus to shrink beam, not intensity
35
    at zero defocus
36
        beam tilt pivot points X,Y
37
        rotation center 
38
39
h3. Presets and Magnifications
40
41
preset   ↓	mag   ↓	ccd   ↓	defocus   ↓	notes   ↓
42
gr 	120X 	512x8 	0 	do not cycle through
43
sq 	1700X 	1024x4 	-3.0e-4 	for square_centering
44
cs 	1700X 	1024x4 	-3.0e-4 	for rct_targeting, copy of sq
45
hl 	5000X 	512x8 	-5.0e-5 	image shift alignment and z-focus
46
en 	50kX-100kX 	4096x1 	-1.5e-6 -- -3.5e-6
47
fa 	50kX-100kX 	512x4 	-1.5e-6 	beam focus and drift checks
48
49
h3. Leginon
50
51
2.4.1 pre-Altas
52
53
    beam shift/intensity for each preset
54
    align image shifts btw presets
55
    matrix calibrate, stage position for gr preset
56
    optional: fine tune rotation center in Leginon, available in any focus node using the fa preset
57
    on rare occasions: check optical axis using Leginon tool, this is only required if the square is shifting alot when you tilt the stage 
58
59
2.4.2 post-Atlas
60
61
    matrix calibrate, stage position for sq preset
62
        error ~50%
63
        shift ~25%
64
        num position ~2-3
65
        interval ~1-3e-6 
66
    RCT node calibration
67
        make sure the number of features is between 700 and 1500
68
        take a picture both with and without the grid bar
69
        take a picture both untilted and tilted 
70
    calibrate dose
71
    RCT node setup
72
    re-align presets
73
    re-center objective aperture 
74
75
h2. Running Leginon RCT
76
77
    Start a square:
78
        Square Targeting (if you have an atlas)
79
            Go to "Square Targeting"
80
            Hit the reload button
81
            Select a new square with the green cross
82
            Click the play |> button 
83
        Simulate Square (if you do NOT have an atlas)
84
            Go to "Square"
85
            Hit the simulate target button 
86
    Square centering
87
        When Leginon is finished, a question mark will appear next to the "Square Centering" node
88
        Click on the "Square Centering" node
89
        Select one z-focus position (blue cross) and one imaging position (green position) close to the center of the square
90
        Click the play |> button 
91
    RCT Targeting
92
        When Leginon is finished setting the z-height, a question mark will appear next to the "RCT Targeting" node
93
        Click on the "RCT Targeting" node
94
        Now you select all the holes to image (not too close to the edge) and a single focus position
95
        Optional: obtain a picture of the tilted square
96
            Click on the "Navigation" node
97
            Open up VNC viewer of the tecnai
98
            In the VNC window, set the tilt angle to the desire angle
99
            In the "Navigation" node, collect an image by clicking on the acquire button (camera logo) at the top
100
            Important, in the VNC window, return the tilt angle to zero degrees, otherwise all your images will fail
101
            Now select you targets in the "RCT Targeting" node and view the tilted image in the "Navigation" node. Jump back and forth between the two nodes until you select all holes present in both images. 
102
        When finished, hit the play |> button 
103
    Focusing
104
        If manual focusing is enabled, wait for the manual focus window to pop up
105
        focus the microscope, reset the defocus, and click the red stop button. 
106
    Watch your images in the "RCT" node 
107
108
h2. Nitrogen fill
109
110
Since RCT only focuses once per square, when you fill the dewar, you must refocus to account for changes in the z-height and focus during refill and there is no check for drift calm down so you must manually wait for drift to subside.
111
112
Steps in Nitrogen fill:
113
114
    Pause RCT node
115
    Wait until Leginon is paused
116
    Fill holder and cold trap dewars
117
    Turn off lights
118
    Check if objective aperture is inserted
119
    RESET TIMER
120
    Focus
121
        simulate z-focus
122
        send hole preset to scope
123
        navigate to good focus spot in navigator
124
        simulate focus in the focus node
125
        repeat focus until drift subsides 
126
    If there is an empty hole: you can measure dose or collect bright fields
127
    Declare drift
128
    Hit play button on RCT node
129
    Check images to make sure they are good
130
    Uncheck user verification of targets 
131
132