RCT run protocol from a user » History » Version 3

Anchi Cheng, 10/17/2012 05:59 PM

1 1 Anchi Cheng
h1. RCT run protocol from a user
2
3
4
h1. Dump from other wiki
5
6
h2. Setup
7
8
h3. Days before
9
10
* Pump down stage
11
*   Cryo-cycle for 4 hours (240 min)
12
*   Prepare grids 
13
14
h3. Scope alignments
15
16
2.2.1 no sample
17
18
    gun shift/tilt
19
    condenser aperture
20
    condenser stig.
21
    check if cold trap is in
22
    adjust eye pieces 
23
24
2.2.2 with sample
25
26
    z-height
27
    focus both M and SA modes
28
    objective stig.
29
    objective aperture at hi mag, use focus to shrink beam, not intensity
30
    at zero defocus
31
        beam tilt pivot points X,Y
32
        rotation center 
33
34
h3. Presets and Magnifications
35
36
|preset|mag|camera config|defocus|notes|
37
|gr|120X|512x8|0|do not cycle through|
38
|sq|1700X|1024x4|-3.0e-4|for square_centering|
39
|cs|1700X|1024x4|-3.0e-4|for rct_targeting, copy of sq|
40
|hl|5000X|512x8|-5.0e-5|image shift alignment and z-focus|
41
|en|50kX-100kX|4096x1|-1.5e-6 to -3.5e-6| |
42
|fa|50kX-100kX|512x4|-1.5e-6|beam focus and drift checks|
43
44
h3. Leginon
45
46
2.4.1 pre-Altas
47
48
    beam shift/intensity for each preset
49
    align image shifts btw presets
50
    matrix calibrate, stage position for gr preset
51
    optional: fine tune rotation center in Leginon, available in any focus node using the fa preset
52
    on rare occasions: check optical axis using Leginon tool, this is only required if the square is shifting alot when you tilt the stage 
53
54
2.4.2 post-Atlas
55
56
    matrix calibrate, stage position for sq preset
57
        error ~50%
58
        shift ~25%
59
        num position ~2-3
60
        interval ~1-3e-6 
61
    RCT node calibration
62
        make sure the number of features is between 700 and 1500
63
        take a picture both with and without the grid bar
64
        take a picture both untilted and tilted 
65
    calibrate dose
66
    RCT node setup
67
    re-align presets
68
    re-center objective aperture 
69
70
h2. Running Leginon RCT
71
72
    Start a square:
73
        Square Targeting (if you have an atlas)
74
            Go to "Square Targeting"
75
            Hit the reload button
76
            Select a new square with the green cross
77
            Click the play |> button 
78
        Simulate Square (if you do NOT have an atlas)
79
            Go to "Square"
80
            Hit the simulate target button 
81
    Square centering
82
        When Leginon is finished, a question mark will appear next to the "Square Centering" node
83
        Click on the "Square Centering" node
84
        Select one z-focus position (blue cross) and one imaging position (green position) close to the center of the square
85
        Click the play |> button 
86
    RCT Targeting
87
        When Leginon is finished setting the z-height, a question mark will appear next to the "RCT Targeting" node
88
        Click on the "RCT Targeting" node
89
        Now you select all the holes to image (not too close to the edge) and a single focus position
90
        Optional: obtain a picture of the tilted square
91
            Click on the "Navigation" node
92
            Open up VNC viewer of the tecnai
93
            In the VNC window, set the tilt angle to the desire angle
94
            In the "Navigation" node, collect an image by clicking on the acquire button (camera logo) at the top
95
            Important, in the VNC window, return the tilt angle to zero degrees, otherwise all your images will fail
96
            Now select you targets in the "RCT Targeting" node and view the tilted image in the "Navigation" node. Jump back and forth between the two nodes until you select all holes present in both images. 
97
        When finished, hit the play |> button 
98
    Focusing
99
        If manual focusing is enabled, wait for the manual focus window to pop up
100
        focus the microscope, reset the defocus, and click the red stop button. 
101
    Watch your images in the "RCT" node 
102
103
h2. Nitrogen fill
104
105
Since RCT only focuses once per square, when you fill the dewar, you must refocus to account for changes in the z-height and focus during refill and there is no check for drift calm down so you must manually wait for drift to subside.
106
107
Steps in Nitrogen fill:
108
109
    Pause RCT node
110
    Wait until Leginon is paused
111
    Fill holder and cold trap dewars
112
    Turn off lights
113
    Check if objective aperture is inserted
114
    RESET TIMER
115
    Focus
116
        simulate z-focus
117
        send hole preset to scope
118
        navigate to good focus spot in navigator
119
        simulate focus in the focus node
120
        repeat focus until drift subsides 
121
    If there is an empty hole: you can measure dose or collect bright fields
122
    Declare drift
123
    Hit play button on RCT node
124
    Check images to make sure they are good
125
    Uncheck user verification of targets 
126
127
    This page was last modified 16:45, 16 July 2010.
128 2 Anchi Cheng
129
130
__________________
131
132 3 Anchi Cheng
[[RCT calibration| < RCT application calibration]]