Leginon Client is run on the microscope/camera-controlling computers (and the robot controlling computer, if applicable) so that the main Leginon program on the remote computer can launch the necessary node at the right place. For example, in order for the remote computer to get and set microscope parameters, the Instrument node need to be launched on the microscope-controlling computer.  Before starting Leginon Client, check that your camera controlling software is set up properly for Leginon interaction:
4 1 Amber Herold
5
6 3 Anchi Cheng
h2. Using a Gatan Camera
7 1 Amber Herold
8 4 Amber Herold
#  Open Digital Micrograph (or check that the Tietz software is closed).
9
 
10
#  insert the Gatan CCD in Digital Micrograph.
11 1 Amber Herold
12 3 Anchi Cheng
13 1 Amber Herold
h2. Using a Tietz Camera
14 3 Anchi Cheng
15 4 Amber Herold
#  Close Tietz TCL program
16
 
17
#  Put camera switch box to select CCD.
18 1 Amber Herold
19 13 Anchi Cheng
h2. Using a Direct Electron DE-12/DE-20 Camera
20 8 Anchi Cheng
21 7 Anchi Cheng
#  Start DirectElectronAPI.exe and MicroManager
22
 
23
#  Set the desired *Frames Per Second* and cool down the camera to the desired temperature in MicroManager
24
 
25
#  Close MicroManager but leave DirectElectronAPI.exe running
26 1 Amber Herold
27 12 Anchi Cheng
h2. Using FEI Falcon/Gatan Orius combination
28
29
# Start TUI
30
# Start Digital Micrograph
31
# Start TIA
32
# Set Camera selection in TUI/CCD/TV Camera to BM-Falcon
33 1 Amber Herold
34 3 Anchi Cheng
h2. Start Leginon Client
35 1 Amber Herold
36
37 3 Anchi Cheng
*  scope> Start Leginon Client (Either with the shortcut on the Desktop or follow the Start Menu-\Start\Programs\Leginon>Leginon Client)
38 1 Amber Herold
39 11 Anchi Cheng
Leginon Client GUI is fairly limited and won't do anything if the main processing host does not put any nodes on it.
40 10 Anchi Cheng
!leginonclientGUI.png!
41 9 Anchi Cheng
42 1 Amber Herold
43
* During Leginon run an Application Error Message may pop up saying
44
45 3 Anchi Cheng
<pre>
46 1 Amber Herold
47 3 Anchi Cheng
Exception EAccessViolation in module adaExp.exe at ....
48 1 Amber Herold
49
50 3 Anchi Cheng
Access violation at address xxxx in module 'adaExp.exe'. Read of address xxxx"
51 1 Amber Herold
52 3 Anchi Cheng
</pre>
53 1 Amber Herold
54 2 Amber Herold
Do NOT touch this message box, or leginon client will need restarting.
55
56 6 Anchi Cheng
h2. If the camera is controlled separately from the scope host, Start Leginon Client there, too.
57
58
See [[Using_Leginon_on_a_system_where_the_microscope_and_camera_are_controlled_by_different_computers]].
59 2 Amber Herold
______
60
61
[[Test Network Connection Between Remote and Microscope Computers|< Test Network Connection Between Remote and Microscope Computers]] | [[Start Leginon at your main working station remotely|Start Leginon at your main working station remotely >]]
62
63 1 Amber Herold
______