Run Leginon Client on the Instrument Computers » History » Version 14
Anchi Cheng, 08/12/2014 02:56 PM
1 | 13 | Anchi Cheng | h1. Run Leginon Client on the Instrument Computers |
---|---|---|---|
2 | 1 | Amber Herold | |
3 | 13 | Anchi Cheng | Leginon Client is run on the microscope/camera-controlling computers (and the robot controlling computer, if applicable) so that the main Leginon program on the remote computer can launch the necessary node at the right place. For example, in order for the remote computer to get and set microscope parameters, the Instrument node need to be launched on the microscope-controlling computer. Before starting Leginon Client, check that your camera controlling software is set up properly for Leginon interaction: |
4 | 1 | Amber Herold | |
5 | |||
6 | 3 | Anchi Cheng | h2. Using a Gatan Camera |
7 | 1 | Amber Herold | |
8 | 4 | Amber Herold | # Open Digital Micrograph (or check that the Tietz software is closed). |
9 | |
||
10 | # insert the Gatan CCD in Digital Micrograph. |
||
11 | 1 | Amber Herold | |
12 | 3 | Anchi Cheng | |
13 | 1 | Amber Herold | h2. Using a Tietz Camera |
14 | 3 | Anchi Cheng | |
15 | 4 | Amber Herold | # Close Tietz TCL program |
16 | |
||
17 | # Put camera switch box to select CCD. |
||
18 | 1 | Amber Herold | |
19 | 13 | Anchi Cheng | h2. Using a Direct Electron DE-12/DE-20 Camera |
20 | 8 | Anchi Cheng | |
21 | 7 | Anchi Cheng | # Start DirectElectronAPI.exe and MicroManager |
22 | |
||
23 | # Set the desired *Frames Per Second* and cool down the camera to the desired temperature in MicroManager |
||
24 | |
||
25 | # Close MicroManager but leave DirectElectronAPI.exe running |
||
26 | 1 | Amber Herold | |
27 | 12 | Anchi Cheng | h2. Using FEI Falcon/Gatan Orius combination |
28 | |||
29 | # Start TUI |
||
30 | # Start Digital Micrograph |
||
31 | # Start TIA |
||
32 | # Set Camera selection in TUI/CCD/TV Camera to BM-Falcon |
||
33 | 1 | Amber Herold | |
34 | 3 | Anchi Cheng | h2. Start Leginon Client |
35 | 1 | Amber Herold | |
36 | |||
37 | 3 | Anchi Cheng | * scope> Start Leginon Client (Either with the shortcut on the Desktop or follow the Start Menu-\Start\Programs\Leginon>Leginon Client) |
38 | 1 | Amber Herold | |
39 | 11 | Anchi Cheng | Leginon Client GUI is fairly limited and won't do anything if the main processing host does not put any nodes on it. |
40 | 10 | Anchi Cheng | !leginonclientGUI.png! |
41 | 9 | Anchi Cheng | |
42 | 1 | Amber Herold | |
43 | * During Leginon run an Application Error Message may pop up saying |
||
44 | |||
45 | 3 | Anchi Cheng | <pre> |
46 | 1 | Amber Herold | |
47 | 3 | Anchi Cheng | Exception EAccessViolation in module adaExp.exe at .... |
48 | 1 | Amber Herold | |
49 | |||
50 | 3 | Anchi Cheng | Access violation at address xxxx in module 'adaExp.exe'. Read of address xxxx" |
51 | 1 | Amber Herold | |
52 | 3 | Anchi Cheng | </pre> |
53 | 1 | Amber Herold | |
54 | 2 | Amber Herold | Do NOT touch this message box, or leginon client will need restarting. |
55 | |||
56 | 6 | Anchi Cheng | h2. If the camera is controlled separately from the scope host, Start Leginon Client there, too. |
57 | |||
58 | See [[Using_Leginon_on_a_system_where_the_microscope_and_camera_are_controlled_by_different_computers]]. |
||
59 | 2 | Amber Herold | ______ |
60 | |||
61 | 14 | Anchi Cheng | [[Test Network Connection Between Remote and Instrument Computers|< Test Network Connection Between Remote and Instrument Computers]] | [[Start Leginon at your main working station remotely|Start Leginon at your main working station remotely >]] |