h1. Run Leginon Client on the Microscope Computer
2
3
4
5
6
Leginon Client is run on the microscope-controlling computer (and the robot controlling
7
computer, if applicable) so that the main Leginon program on the remote computer can launch
8
the necessary node at the right place. For example, in order for the remote computer to get
9
and set microscope parameters, the Instrument node need to be launched on the
10
microscope-controlling computer.
11
12
13
14
h2. Start Leginon Client
15
16
17
18
19
20
*  scope> Start Leginon Client (Either with the shortcut on the Desktop or follow
21
the Start Menu-\Start\Programs\Leginon>Leginon Client)
22
23
24
25
26
27
28
* During Leginon run an Application Error Message may pop up saying
29
30
<blockquote>
31
32
_"Exception EAccessViolation in module adaExp.exe at ...._
33
34
35
_Access violation at address xxxx in module 'adaExp.exe'. Read of address
36
xxxx"_
37
38
</blockquote>
39
40
Do NOT touch this message box, or leginon client will need restarting.
41
42
43
44
45
46
h2. Using a Gatan Camera
47
48
49
50
51
52
*  Open Digital Micrograph (or check that the Tietz software is closed).
53
54
55
*  insert the Gatan CCD in Digital Micrograph.
56
57
58
59
60
61
62
63
64
65
h2. Using a Tietz Camera
66
67
68
69
70
71
*  Close Tietz TCL program
72
73
74
*  Put camera switch box to select CCD.
75 2 Amber Herold
76
77