Running "Manual" Application » History » Version 4
Anchi Cheng, 05/22/2010 11:45 PM
1 | 1 | Amber Herold | h1. Running "Manual" Application |
---|---|---|---|
2 | |||
3 | |||
4 | |||
5 | * Leginon/Manual> Open the "Settings" tool !http://emg.nysbc.org/software/leginon/images/icons/settings.png! and configure the settings: |
||
6 | |||
7 | |||
8 | 4 | Anchi Cheng | Important-Use low dose kit option should be set to yes if using the low dose kit on Tecnai. Low Dose Pause Time need to be set to 5 sec or more to complete the mode change. |
9 | 1 | Amber Herold | |
10 | |||
11 | 4 | Anchi Cheng | Recommandations- |
12 | > > > Acquisition/Camera Configuration- use one of the preconfigured selection for binning and size |
||
13 | 1 | Amber Herold | |
14 | |||
15 | 3 | Anchi Cheng | > > > Correct image=yes if dark and bright images of the camera configuration exist. |
16 | 1 | Amber Herold | |
17 | |||
18 | 3 | Anchi Cheng | > > > Save image to the database=yes. |
19 | 1 | Amber Herold | |
20 | |||
21 | * Leginon/manual/Settings> Save and close the settings by pressing "OK" button. |
||
22 | |||
23 | |||
24 | * Adjust the exposure time in the setting and/or beam size/spot size on the scope to achieve the desired dose and coverage in exposure mode, if low-dose kit is used. |
||
25 | |||
26 | |||
27 | * Microscope> Search the grid for interested area and focus as normally performed. |
||
28 | |||
29 | |||
30 | * Microscope> If low dose mode is used, leave TEM in Focus model. Leave the screen at up position if "Main Screen/Up before acquire" is not checked in the settings. No need to put on beam blank. |
||
31 | |||
32 | |||
33 | * Leginon/manual> Click on the "Acquire" tool !http://emg.nysbc.org/software/leginon/images/icons/acquire.png! to acquire the image. |
||
34 | |||
35 | |||
36 | * Microscope> If low dose mode is used, Leginon will force the pre-specimen shutter close, switch the scope to exposure mode to acquire the image and then switch back to the original mode. |
||
37 | 2 | Amber Herold | |
38 | ______ |
||
39 | |||
40 | [[Startup|< Startup]] | [[Advanced Features|Advanced Features >]] |
||