Running "Manual" Application » History » Version 5

Amber Herold, 06/23/2010 02:33 PM

1 1 Amber Herold
h1. Running "Manual" Application
2
3
4
5
*  Leginon/Manual> Open the "Settings" tool !http://emg.nysbc.org/software/leginon/images/icons/settings.png! and configure the settings:
6
7
8 4 Anchi Cheng
Important-Use low dose kit option should be set to yes if using the low dose kit on Tecnai. Low Dose Pause Time need to be set to 5 sec or more to complete the mode change.
9 1 Amber Herold
10
11 5 Amber Herold
Recommendations-
12 4 Anchi Cheng
> > > Acquisition/Camera Configuration- use one of the preconfigured selection for binning and size
13 1 Amber Herold
14
15 3 Anchi Cheng
> > > Correct image=yes if dark and bright images of the camera configuration exist.
16 1 Amber Herold
17
18 3 Anchi Cheng
> > > Save image to the database=yes.
19 1 Amber Herold
20
21
*  Leginon/manual/Settings> Save and close the settings by pressing "OK" button.
22
23
24
*  Adjust the exposure time in the setting and/or beam size/spot size on the scope to achieve the desired dose and coverage in exposure mode, if low-dose kit is used.
25
26
27
*  Microscope> Search the grid for interested area and focus as normally performed.
28
29
30
*  Microscope> If low dose mode is used, leave TEM in Focus model. Leave the screen at up position if "Main Screen/Up before acquire" is not checked in the settings. No need to put on beam blank.
31
32
33
*  Leginon/manual> Click on the "Acquire" tool !http://emg.nysbc.org/software/leginon/images/icons/acquire.png! to acquire the image.
34
35
36
*  Microscope> If low dose mode is used, Leginon will force the pre-specimen shutter close, switch the scope to exposure mode to acquire the image and then switch back to the original mode.
37 2 Amber Herold
38
______
39
40
[[Startup|< Startup]] | [[Advanced Features|Advanced Features >]]