3 | |||
4 | |||
5 | 6 | Amber Herold | # Leginon/Manual> Open the "Settings" tool !http://emg.nysbc.org/software/leginon/images/icons/settings.png! and configure the settings: |
6 | |
||
7 | 1 | Amber Herold | Important-Use low dose kit option should be set to yes if using the low dose kit on Tecnai. Low Dose Pause Time need to be set to 5 sec or more to complete the mode change. |
8 | 6 | Amber Herold | |
9 | 1 | Amber Herold | Recommendations- |
10 | 6 | Amber Herold | ** Acquisition/Camera Configuration- use one of the preconfigured selection for binning and size |
11 | |
||
12 | ** Correct image=yes if dark and bright images of the camera configuration exist. |
||
13 | |
||
14 | ** Save image to the database=yes. |
||
15 | |
||
16 | # Leginon/manual/Settings> Save and close the settings by pressing "OK" button. |
||
17 | |
||
18 | # Adjust the exposure time in the setting and/or beam size/spot size on the scope to achieve the desired dose and coverage in exposure mode, if low-dose kit is used. |
||
19 | |
||
20 | # Microscope> Search the grid for interested area and focus as normally performed. |
||
21 | |
||
22 | # Microscope> If low dose mode is used, leave TEM in Focus model. Leave the screen at up position if "Main Screen/Up before acquire" is not checked in the settings. No need to put on beam blank. |
||
23 | |
||
24 | # Leginon/manual> Click on the "Acquire" tool !http://emg.nysbc.org/software/leginon/images/icons/acquire.png! to acquire the image. |
||
25 | |
||
26 | # Microscope> If low dose mode is used, Leginon will force the pre-specimen shutter close, switch the scope to exposure mode to acquire the image and then switch back to the original mode. |
||