Project

General

Profile

Running the application » History » Version 18

Anchi Cheng, 07/15/2010 09:40 PM

1 1 Amber Herold
h1. Running the application
2
3
4
5
h2. Import Notes about Image Intensity Recorded through
6
Tomography Node
7
8
9
10 14 Amber Herold
Tomography node saves the images in a different format from other Acquisition nodes. By
11 1 Amber Herold
default, the flat-field correct CCD counts are multiplied by 10 and converted to signed
12
16-bit integer before the image is displayed and saved. This makes CCD counts of 3276.8 or
13
larger overflow to negatives. Other Leginon Acquisition images are saved as float without
14
manipulation.
15
16
17
To avoid this problem, find out what exposure time corresponds to the fractionated dose
18
from your tilt angle step and range and total dose and take an image at tomo preset with
19
such an exposure in Navigation node. You will need to reduce the total dose if a good
20
fraction of the counts are larger than 3200 even though it would not appear to be saturated
21
in the float scale without the 10x factor. Alternatively, change the scale factor in
22
Tomography node.
23
24
25
26
27
28
h2. Multiscale Imaging
29
30
31
32
*  Preset image shift alignment/beam shift alignment are the same as in MSI
33
application
34
35
36 5 Amber Herold
*  New dark/bright references should be reacquired for "tomo" preset that acquires the
37 1 Amber Herold
final data. It is best to do this at the same dose per tomography image calculated from
38
the total dose, the tilt parameters, and the dose measurement.
39
40
41
*  For best focusing result, perform autofocus at the same magnification as the
42
tomography data collection, align microscope well at the eucentric focus and the
43
rotation center and save them before data collection.
44
45
46
47
48
49
50
51
h2. Using Tomography Preview
52
53
54
55
*  Preview targets (pink) can be selected when selecting targets in "Tomography
56
Targeting"
57
58
59
*  When the targets are processed, targets that are of the type "preview" are
60
processed before focus and acquisition targets.
61
62
63
*  Tomography Preview node acquires a image at the preview target using "preview"
64
preset which should be set at minimal dose.
65
66
67
68
69
70
71
72
h2. Dose Measurement
73
74
75
76
If "Measure Dose before collection" is checked in Tomography node, the stage will be
77
moved to the reference target and a dose image of the "tomo" preset will be acquired (center
78
512x512 of whateven binning of the preset) before each tilt series if the interval between
79
the series is longer than the limit time set in the settings of Dose Measurement node. The
80
measured value will then be used to recalculate the proper exposure time for tomography
81
imaging.
82
83
84
For this function to behave properly, the followings should be done during
85
operation:
86
87
88
* One, and only one, "reference" target should be selected in either "Square
89
Targeting" or "Hole Targeting" or "Tomography Targeting" node. The reference target
90
should be of either a broken square or a empty hole if no broken square can be
91
found.
92
93
94
* "Measure Dose" before collection should be selected in Tomography node.
95
96
97
* "Exposure time max/min" in Tomography node should be in a range that can accommodate
98
the electron beam fluctuation over time.
99
100
101
102
103
104
105
106
h2. Align Zero Loss Peak
107
108
109
110
This function applies only to Gatan energy filter EFTEM. If "Align ZLP before
111
collection" is checked in Tomography node, the stage will be moved to the reference target
112
and starts the procedure to align zero loss peak before each tilt series if the interval
113
between the series is longer than the limit time set in the settings of Dose Measurement
114
node.
115
116
117
For this function to behave properly, the followings should be done during
118
operation:
119
120
121
* One, and only one, "reference" target should be selected in either "Square
122
Targeting" or "Hole Targeting" or "Tomography Targeting" node. The reference target
123
should be of either a broken square or a empty hole if no broken square can be found.
124
This is the same reference target used for dose measurement.
125
126
127
* "Align ZLP" before collection should be selected in Tomography node.
128
129
130
131
132
133
134
135
h2. Low Magnification Model Fitting
136
137
138
139
The fitting of optical axis offset does not always works if the offset is so large that
140
the feature moves out of view with even a small tilt. In such a case, it is worth first
141
collect a tomography series at a lower magnification to define roughly the model.
142
143
144
At beginning of each session, or forced by the user, the model is initialized. By
145
default, at the initialization, Tomography node uses past fitting results that show good
146
agreement with the experimental data at the magnification of the preset used. If a good
147
model is not found, that from lower magnifications will be used. It is possible to force the
148
node to use a model fitted at a particular magnification by selecting it in
149
Tomography/Settings/Model.
150
151
152
Therefore, we recommend that, in case of fitting failure on good contrast images, the
153
followings should be done:
154
155
156 3 Amber Herold
#  Tomography/Settings/Image Acquisition> change the preset to "hl".
157
#  Tomography/Settings> adjust Tilt and Exposure parameters to match.
158
#  Acquire the tiltseries images.
159
#  If the tracking is good, change the preset back.
160
#  Tomography/Settings/Model> Initialize with the model of (the mag of "hl"
161 1 Amber Herold
preset).
162 3 Amber Herold
#  Acquire the tomo-series.
163
#  If tracking is good, change back to Initialize with the model of "this preset and
164 1 Amber Herold
lower" mag.
165
166
167
168
169
170
171
172
h2. What is a Good Tilt-Axis Model?
173
174
175
176 5 Amber Herold
The goniometer-tilt-axis-based tracking model developed by Zheng et. al. corrects the
177
specimen height (z-axis) by a change of defocus using measured shift of feature shifts in
178 1 Amber Herold
the images (x and y-axes). The tracking in the x and y directions does not involve the use
179 5 Amber Herold
of such model, but is done by smooth curve fitting or preceding tilts. Therefore, to judge
180 1 Amber Herold
the adequacy of the model, one should check the resulting defocii of the images in the
181
series remain unchanged.
182
183
184
On the other hand, the feature tracking in x and y is likely to fail only if the tilting
185
does not induce a smooth shift of the imaging feature a sudden drop of specimen position at
186
a particular tilt angle often throws off the smooth curve fitting. It is possible to reduce
187
such effect by increasing the number of data points included in the smoothing as set in the
188
model section of the tomography node settings window. Otherwise, the goniometer need to be
189
serviced.
190
191 18 Anchi Cheng
[[Tomography Trouble Shooting]] 
192 1 Amber Herold
193
194
195
196
h2. Failure of xy feature tracking
197
198
199
200 9 Amber Herold
Feature tracking in x and y axes is a 2nd order polynomial fit of preceeding data
201 1 Amber Herold
points. The default uses 5 data points. When a sudden jump occurs in the tracking error, it
202
tend to follow the trend of the last point. If the jump is a temporary clich in the
203
goniometer, this tend to over correct the tracking error and eventually loose track as shown
204
in Figure 5. A possible fix is to increase the number of data points in the fitting. This
205
can be set in the tomography setting "Smooth n tilts for defocus prediction". 4 in defocus
206
prediction is equivalent to 5 points (n+1) for xy tracking.
207
208
209 6 Amber Herold
*Figure 5* 
210
!http://emg.nysbc.org/software/leginon/images/images/tomoxbad.png!
211 1 Amber Herold
212
213
214
215
216
h2. Large tracking error between the first and second tilt
217
images
218
219
220
221
The first image in each tilt group of the tilt series at the "start" angle (normally 0
222
deg) and the second image at tilt of "step" angle from the "start" angle do not use the
223
fitted model.  It is assumed that the eucentric height judged by stage alpha wobbling in the
224
"Tomo Focus" node gives a stage height that the tracking of feature by such a small tilt
225
would be good enough.  In most cases this is a reasonable assumption.  However, we have had
226
experience of goniometer alignment problem where the assumption fails.  The symptom is
227
illustrated in Figure 6 below.  Note that the Feature tracking error is displayed as
228
percentage of the image length.
229
230
231 6 Amber Herold
*Figure 6*
232
!http://emg.nysbc.org/software/leginon/images/images/tomoz0bad.png!
233 1 Amber Herold
234
235 11 Amber Herold
This tilt series was taken with a starting angle of zero and at an image size of < 1
236
um.  As can be seen here, apart from the 2 and minus 2 degree tilts, the tracking error was less
237 8 Amber Herold
than 2 % of the image.  Only the tracking of the feature between 0 and +/- 2 degrees are
238 11 Amber Herold
large.  At close to 20 % error, this made the overlap between plus and minus 2 degrees unacceptable and
239 1 Amber Herold
often cause popular alignment programs to misalign the two half of the series.
240
241
242
The first solution is of course to report it to your microscope service engineer.  When
243
we had this problem, many users noticed that it was difficult to adjust stage to eucentric
244
height manually with alpha wobbler.  Features jumped while the goniometer changed rotation
245
direction.  In addition, different magnitude of tilt range suggests different eucentric
246
heights.  It is not easy to fix this, so it might take a while.
247
248
249
Before the hardware is fixed physically, it is still possible collect tomograms.  The
250
model fitting of the overall curve in the above case gave z0 of +5 um through the whole tilt
251
series (Figure not shown).  Therefore, by moving the stage up by such an amount after the
252
stage-tilt-based autofocusing can bring us to the correct height for tomography.  This can
253
be acheived by saving the "tomo eucentric" focus current to the database, align rotation
254
center for this stage height and focus.  Then change the correction type of the
255
"Beam_Tilt_Fine" focusing step to "Stage Z".
256
257
258
259
260
261
h2. Failure of model-based correction
262
263
264
265
The model used in the defocus correction in Leginon tomography node is a very simplified
266
one. There are a few cases when the approach fails. Here are ones that we have
267
encountered:
268
269
270
271
h2. Y-axis looping
272
273
274
275
The microscope goniometer does not move on only the tilt axis. With its complex
276
structure, a common problem is that when the stage is highly tilt, the position slips in
277
the y-direction. This is known as looping. Figure 7 shows an example of this problem.
278
279
280 6 Amber Herold
*Figure 7*
281
!http://emg.nysbc.org/software/leginon/images/images/tomoxyloop.png!
282 1 Amber Herold
283
284
While the x-axis position shifts monotonically as a stable model should be, the y-axis
285
in the positive tilt direction changes little from 0-30 degrees before it increases
286
rapidly after 30 degrees. Even though the tracking in xy plane is still good, the defocii
287
correction at these higher tilts may no longer be correct if the tilt axis parameters are
288
fitted dynamically. Figure 8 shows the model parameters of the same tilt series where the
289
fitted phi and offset starts to change above 30 degrees even though the tilt axis has not
290
moved according to the shrinking behavior of the images during the tilts. Note that in
291
this particular case the looping problem is still mild so that the over-correction is not
292
very strong. only a small slope change is resulted in z0 prediction. In worst cases, the
293
defocus over-correction is so large that the adjacent images can not correlate properly
294
and even the xy tracking would fail. The spikes around zero tilt is a display data sorting
295
error of the identical starting tilt of the two tilt groups.
296
297
298 6 Amber Herold
*Figure 8*
299
!http://emg.nysbc.org/software/leginon/images/images/tomomodelloop.png!
300 1 Amber Herold
301
302
Other than asking microscope service engineer to fix the looping, one can find the
303
best fixed model in the series to apply to future tilt data collection. To make the fixed
304
model permantly saved to the database, follow these steps:
305
306
307 12 Amber Herold
# tomography/settings/model>activate "keep the tilt axis parameters
308 1 Amber Herold
fixed".
309 12 Amber Herold
# tomography/settings/model>initialize the model with "custom values". Enter best
310 1 Amber Herold
estimate of the fixed model. For example, in the positive direction, enter phi as
311
-2.17 degrees and axis offset as -1.52 um. since these are the stable values up to the
312
point the y-looping starts.
313 12 Amber Herold
# tomography>collect a full tilt series. If the run is successful with good
314 1 Amber Herold
tracking in all three axis, the model will be saved in the database for this
315
magnification.
316 12 Amber Herold
# tomography/settings/model>From now on, you can initialize the model with "only
317 1 Amber Herold
this preset" or "this preset and lower mags"
318
319
320
321
322
323
324
325
h2. Grid slips between the first and second tilt directions
326
327
328
329
When the holder does not hold the grid tightly, the grid slips to a different position
330
when the first tilt direction ends and the goniometer quickly returns to zero tilt.
331
Leginon is designed to adjust the target before the second tilt group starts. The default
332
setting for this function is to use only the parent image (i.e. one ancestor) where the
333
target comes from as reference. If the slip is larger than the size of the parent image,
334
the adjustment may fail, and a random target would be acquired in the second tilt
335
group.
336
337
338
Starting from Leginon 1.6, the target adjustment can be done with all ancestor images
339
of the target by choosing "all" in the acquisition part of the tomography node setting to
340 12 Amber Herold
adjust target with all ancestors. The node "Target Adjustment" limits the lowest
341 1 Amber Herold
magnification that this target adjustment would go up in ancestry. The default is at 300x
342 12 Amber Herold
so that the presence of the objective aperture does not create difference in the reacquired
343 1 Amber Herold
ancestor image from its original.
344
345
346
347
348
349
h2. Strong and continuous specimen drift
350
351
352
353
The model used in Leginon considers any shift of feature in the image a result of tilt
354
axis not aligning to the center of the detector. With the phi and offset fixed, all errors
355
are accumulated in z0 and results in bad defocus correction. There is no solution to this
356
at the moment.
357 13 Amber Herold
358
359
360
361
______
362
363
[[Set-up Before Running|< Set-up Before Running]] | [[Full Protocol on a F30 with an energy filter| Full Protocol on a F30 with an energy filter >]]