Running the application » History » Version 19
Anchi Cheng, 07/15/2010 09:58 PM
1 | 1 | Amber Herold | h1. Running the application |
---|---|---|---|
2 | |||
3 | |||
4 | |||
5 | h2. Import Notes about Image Intensity Recorded through |
||
6 | Tomography Node |
||
7 | |||
8 | |||
9 | |||
10 | 14 | Amber Herold | Tomography node saves the images in a different format from other Acquisition nodes. By |
11 | 1 | Amber Herold | default, the flat-field correct CCD counts are multiplied by 10 and converted to signed |
12 | 16-bit integer before the image is displayed and saved. This makes CCD counts of 3276.8 or |
||
13 | larger overflow to negatives. Other Leginon Acquisition images are saved as float without |
||
14 | manipulation. |
||
15 | |||
16 | |||
17 | To avoid this problem, find out what exposure time corresponds to the fractionated dose |
||
18 | from your tilt angle step and range and total dose and take an image at tomo preset with |
||
19 | such an exposure in Navigation node. You will need to reduce the total dose if a good |
||
20 | fraction of the counts are larger than 3200 even though it would not appear to be saturated |
||
21 | in the float scale without the 10x factor. Alternatively, change the scale factor in |
||
22 | Tomography node. |
||
23 | |||
24 | |||
25 | |||
26 | |||
27 | |||
28 | h2. Multiscale Imaging |
||
29 | |||
30 | |||
31 | |||
32 | * Preset image shift alignment/beam shift alignment are the same as in MSI |
||
33 | application |
||
34 | |||
35 | |||
36 | 5 | Amber Herold | * New dark/bright references should be reacquired for "tomo" preset that acquires the |
37 | 1 | Amber Herold | final data. It is best to do this at the same dose per tomography image calculated from |
38 | the total dose, the tilt parameters, and the dose measurement. |
||
39 | |||
40 | |||
41 | * For best focusing result, perform autofocus at the same magnification as the |
||
42 | tomography data collection, align microscope well at the eucentric focus and the |
||
43 | rotation center and save them before data collection. |
||
44 | |||
45 | |||
46 | |||
47 | |||
48 | |||
49 | |||
50 | |||
51 | h2. Using Tomography Preview |
||
52 | |||
53 | |||
54 | |||
55 | * Preview targets (pink) can be selected when selecting targets in "Tomography |
||
56 | Targeting" |
||
57 | |||
58 | |||
59 | * When the targets are processed, targets that are of the type "preview" are |
||
60 | processed before focus and acquisition targets. |
||
61 | |||
62 | |||
63 | * Tomography Preview node acquires a image at the preview target using "preview" |
||
64 | preset which should be set at minimal dose. |
||
65 | |||
66 | |||
67 | |||
68 | |||
69 | |||
70 | |||
71 | |||
72 | h2. Dose Measurement |
||
73 | |||
74 | |||
75 | |||
76 | If "Measure Dose before collection" is checked in Tomography node, the stage will be |
||
77 | moved to the reference target and a dose image of the "tomo" preset will be acquired (center |
||
78 | 512x512 of whateven binning of the preset) before each tilt series if the interval between |
||
79 | the series is longer than the limit time set in the settings of Dose Measurement node. The |
||
80 | measured value will then be used to recalculate the proper exposure time for tomography |
||
81 | imaging. |
||
82 | |||
83 | |||
84 | For this function to behave properly, the followings should be done during |
||
85 | operation: |
||
86 | |||
87 | |||
88 | * One, and only one, "reference" target should be selected in either "Square |
||
89 | Targeting" or "Hole Targeting" or "Tomography Targeting" node. The reference target |
||
90 | should be of either a broken square or a empty hole if no broken square can be |
||
91 | found. |
||
92 | |||
93 | |||
94 | * "Measure Dose" before collection should be selected in Tomography node. |
||
95 | |||
96 | |||
97 | * "Exposure time max/min" in Tomography node should be in a range that can accommodate |
||
98 | the electron beam fluctuation over time. |
||
99 | |||
100 | |||
101 | |||
102 | |||
103 | |||
104 | |||
105 | |||
106 | h2. Align Zero Loss Peak |
||
107 | |||
108 | |||
109 | |||
110 | This function applies only to Gatan energy filter EFTEM. If "Align ZLP before |
||
111 | collection" is checked in Tomography node, the stage will be moved to the reference target |
||
112 | and starts the procedure to align zero loss peak before each tilt series if the interval |
||
113 | between the series is longer than the limit time set in the settings of Dose Measurement |
||
114 | node. |
||
115 | |||
116 | |||
117 | For this function to behave properly, the followings should be done during |
||
118 | operation: |
||
119 | |||
120 | |||
121 | * One, and only one, "reference" target should be selected in either "Square |
||
122 | Targeting" or "Hole Targeting" or "Tomography Targeting" node. The reference target |
||
123 | should be of either a broken square or a empty hole if no broken square can be found. |
||
124 | This is the same reference target used for dose measurement. |
||
125 | |||
126 | |||
127 | * "Align ZLP" before collection should be selected in Tomography node. |
||
128 | |||
129 | |||
130 | |||
131 | |||
132 | |||
133 | |||
134 | |||
135 | h2. Low Magnification Model Fitting |
||
136 | |||
137 | |||
138 | |||
139 | The fitting of optical axis offset does not always works if the offset is so large that |
||
140 | the feature moves out of view with even a small tilt. In such a case, it is worth first |
||
141 | collect a tomography series at a lower magnification to define roughly the model. |
||
142 | |||
143 | |||
144 | At beginning of each session, or forced by the user, the model is initialized. By |
||
145 | default, at the initialization, Tomography node uses past fitting results that show good |
||
146 | agreement with the experimental data at the magnification of the preset used. If a good |
||
147 | model is not found, that from lower magnifications will be used. It is possible to force the |
||
148 | node to use a model fitted at a particular magnification by selecting it in |
||
149 | Tomography/Settings/Model. |
||
150 | |||
151 | |||
152 | Therefore, we recommend that, in case of fitting failure on good contrast images, the |
||
153 | followings should be done: |
||
154 | |||
155 | |||
156 | 3 | Amber Herold | # Tomography/Settings/Image Acquisition> change the preset to "hl". |
157 | # Tomography/Settings> adjust Tilt and Exposure parameters to match. |
||
158 | # Acquire the tiltseries images. |
||
159 | # If the tracking is good, change the preset back. |
||
160 | # Tomography/Settings/Model> Initialize with the model of (the mag of "hl" |
||
161 | 1 | Amber Herold | preset). |
162 | 3 | Amber Herold | # Acquire the tomo-series. |
163 | # If tracking is good, change back to Initialize with the model of "this preset and |
||
164 | 1 | Amber Herold | lower" mag. |
165 | |||
166 | |||
167 | |||
168 | |||
169 | |||
170 | |||
171 | |||
172 | h2. What is a Good Tilt-Axis Model? |
||
173 | |||
174 | |||
175 | |||
176 | 5 | Amber Herold | The goniometer-tilt-axis-based tracking model developed by Zheng et. al. corrects the |
177 | specimen height (z-axis) by a change of defocus using measured shift of feature shifts in |
||
178 | 1 | Amber Herold | the images (x and y-axes). The tracking in the x and y directions does not involve the use |
179 | 5 | Amber Herold | of such model, but is done by smooth curve fitting or preceding tilts. Therefore, to judge |
180 | 1 | Amber Herold | the adequacy of the model, one should check the resulting defocii of the images in the |
181 | series remain unchanged. |
||
182 | |||
183 | |||
184 | On the other hand, the feature tracking in x and y is likely to fail only if the tilting |
||
185 | does not induce a smooth shift of the imaging feature a sudden drop of specimen position at |
||
186 | a particular tilt angle often throws off the smooth curve fitting. It is possible to reduce |
||
187 | such effect by increasing the number of data points included in the smoothing as set in the |
||
188 | model section of the tomography node settings window. Otherwise, the goniometer need to be |
||
189 | serviced. |
||
190 | |||
191 | 19 | Anchi Cheng | [[Troubles with Tomography|Tomography Trouble Shooting]] section covers many of the problems we have encountered. |
192 | 1 | Amber Herold | |
193 | |||
194 | |||
195 | 13 | Amber Herold | |
196 | |||
197 | |||
198 | |||
199 | ______ |
||
200 | |||
201 | [[Set-up Before Running|< Set-up Before Running]] | [[Full Protocol on a F30 with an energy filter| Full Protocol on a F30 with an energy filter >]] |