Project

General

Profile

Running the application » History » Version 2

Amber Herold, 06/18/2010 03:41 PM

1 1 Amber Herold
h1. Running the application
2
3
4
5
h2. Import Notes about Image Intensity Recorded through
6
Tomography Node
7
8
9
10
Tomography node save the images in a different format from other Acquisition nodes. By
11
default, the flat-field correct CCD counts are multiplied by 10 and converted to signed
12
16-bit integer before the image is displayed and saved. This makes CCD counts of 3276.8 or
13
larger overflow to negatives. Other Leginon Acquisition images are saved as float without
14
manipulation.
15
16
17
To avoid this problem, find out what exposure time corresponds to the fractionated dose
18
from your tilt angle step and range and total dose and take an image at tomo preset with
19
such an exposure in Navigation node. You will need to reduce the total dose if a good
20
fraction of the counts are larger than 3200 even though it would not appear to be saturated
21
in the float scale without the 10x factor. Alternatively, change the scale factor in
22
Tomography node.
23
24
25
26
27
28
h2. Multiscale Imaging
29
30
31
32
*  Preset image shift alignment/beam shift alignment are the same as in MSI
33
application
34
35
36
*  New dark/bright references should be reaquired for "tomo" preset that acquires the
37
final data. It is best to do this at the same dose per tomography image calculated from
38
the total dose, the tilt parameters, and the dose measurement.
39
40
41
*  For best focusing result, perform autofocus at the same magnification as the
42
tomography data collection, align microscope well at the eucentric focus and the
43
rotation center and save them before data collection.
44
45
46
47
48
49
50
51
h2. Using Tomography Preview
52
53
54
55
*  Preview targets (pink) can be selected when selecting targets in "Tomography
56
Targeting"
57
58
59
*  When the targets are processed, targets that are of the type "preview" are
60
processed before focus and acquisition targets.
61
62
63
*  Tomography Preview node acquires a image at the preview target using "preview"
64
preset which should be set at minimal dose.
65
66
67
68
69
70
71
72
h2. Dose Measurement
73
74
75
76
If "Measure Dose before collection" is checked in Tomography node, the stage will be
77
moved to the reference target and a dose image of the "tomo" preset will be acquired (center
78
512x512 of whateven binning of the preset) before each tilt series if the interval between
79
the series is longer than the limit time set in the settings of Dose Measurement node. The
80
measured value will then be used to recalculate the proper exposure time for tomography
81
imaging.
82
83
84
For this function to behave properly, the followings should be done during
85
operation:
86
87
88
* One, and only one, "reference" target should be selected in either "Square
89
Targeting" or "Hole Targeting" or "Tomography Targeting" node. The reference target
90
should be of either a broken square or a empty hole if no broken square can be
91
found.
92
93
94
* "Measure Dose" before collection should be selected in Tomography node.
95
96
97
* "Exposure time max/min" in Tomography node should be in a range that can accommodate
98
the electron beam fluctuation over time.
99
100
101
102
103
104
105
106
h2. Align Zero Loss Peak
107
108
109
110
This function applies only to Gatan energy filter EFTEM. If "Align ZLP before
111
collection" is checked in Tomography node, the stage will be moved to the reference target
112
and starts the procedure to align zero loss peak before each tilt series if the interval
113
between the series is longer than the limit time set in the settings of Dose Measurement
114
node.
115
116
117
For this function to behave properly, the followings should be done during
118
operation:
119
120
121
* One, and only one, "reference" target should be selected in either "Square
122
Targeting" or "Hole Targeting" or "Tomography Targeting" node. The reference target
123
should be of either a broken square or a empty hole if no broken square can be found.
124
This is the same reference target used for dose measurement.
125
126
127
* "Align ZLP" before collection should be selected in Tomography node.
128
129
130
131
132
133
134
135
h2. Low Magnification Model Fitting
136
137
138
139
The fitting of optical axis offset does not always works if the offset is so large that
140
the feature moves out of view with even a small tilt. In such a case, it is worth first
141
collect a tomography series at a lower magnification to define roughly the model.
142
143
144
At beginning of each session, or forced by the user, the model is initialized. By
145
default, at the initialization, Tomography node uses past fitting results that show good
146
agreement with the experimental data at the magnification of the preset used. If a good
147
model is not found, that from lower magnifications will be used. It is possible to force the
148
node to use a model fitted at a particular magnification by selecting it in
149
Tomography/Settings/Model.
150
151
152
Therefore, we recommend that, in case of fitting failure on good contrast images, the
153
followings should be done:
154
155
156
*  Tomography/Settings/Image Acquisition> change the preset to "hl".
157
158
159
*  Tomography/Settings> adjust Tilt and Exposure parameters to match.
160
161
162
*  Acquire the tiltseries images.
163
164
165
*  If the tracking is good, change the preset back.
166
167
168
*  Tomography/Settings/Model> Initialize with the model of (the mag of "hl"
169
preset).
170
171
172
*  Acquire the tomo-series.
173
174
175
*  If tracking is good, change back to Initialize with the model of "this preset and
176
lower" mag.
177
178
179
180
181
182
183
184
h2. What is a Good Tilt-Axis Model?
185
186
187
188
The goniometer-tilt-axis-based tracking model developped by Zheng et. al. corrects the
189
speciman height (z-axis) by a change of defocus using measured shift of feature shifts in
190
the images (x and y-axes). The tracking in the x and y directions does not involve the use
191
of such model, but is done by smooth curve fitting or preceeding tilts. Therefore, to judge
192
the adequacy of the model, one should check the resulting defocii of the images in the
193
series remain unchanged.
194
195
196
On the other hand, the feature tracking in x and y is likely to fail only if the tilting
197
does not induce a smooth shift of the imaging feature a sudden drop of specimen position at
198
a particular tilt angle often throws off the smooth curve fitting. It is possible to reduce
199
such effect by increasing the number of data points included in the smoothing as set in the
200
model section of the tomography node settings window. Otherwise, the goniometer need to be
201
serviced.
202
203
204
205
206
207
h2. Checking and Improving Tilt-Axis Model
208
209
210
211
Test runs should be done at the magnification, defocus, and tilts up to the highest
212
values planned for the real ones with a specimen with isotropic features in view and
213
adjusted to eucentric height. Tilt steps should be made to allow at least 15 images taken
214
per tilt direction.
215
216
217
218
h2. First Test Run: Fixed Model
219
220
221
222
The first time the tomography application is used on a microscope, all model
223
parameters are default to zero. A test run should be done with the specimen at eucentric
224
height and all default settings including "keep the tilt axis parameters fixed" activated.
225
If the feature tracking is good, and the defocii of the images in the tilt series do not
226
significantly changed, the goniometer behavior and camera alignment are close to ideal,
227
and no improvement of the model is required.
228
229
230
If the xy tracking deteriorates quickly in the first few tilts, the magnification
231
should be dropped so that tracking is possible through out the tilt series. Once a rough
232
model is established, refinement can then be done at higher magnifications.
233
234
235
236
237
238
h2. Runs to Improve the Model-Fitting
239
240
241
242
A complete tilt model fitting is difficult with small number of data points or at only
243
low tilts. Therefore, two tilt series need to be taken in order to get a good averaged
244
model that is later fixed.
245
246
247
* tomography/settings/model>deactivate "keep the tilt axis parameters
248
fixed".
249
250
251
* tomography/settings/model>initialize the model with "this preset only" unless
252
this is a refinement of <link linkend="Tomo_low_mag_fitting">a rough model obtained at
253
a lower magnification</link>, in which case, initialize the model with "this preset
254
and lower mags".
255
256
257
* tomography/toolbar>click on "reset learning" !http://emg.nysbc.org/software/leginon/images/icons/refresh.png! to have
258
a fresh start of tilt series data included in the fitting.
259
260
261
* Presets Manager>send the "tomo" preset to scope.
262
263
264
* tomography>Start the tilt series image collection by clicking on "Simulated
265
Target" tool.
266
267
268
* repeat 4 and 5 to dynamically fit the model the second time if the first one goes
269
to completion.
270
271
272
If the target shifts away during the run, reduce the magnification, repeat 1
273
through 9 in this procedure and then return to the higher, intended mag and refined
274
the model using the same procedure and proper setting for refinement as mentioned in
275
2.
276
277
278
* Check on web image viewer to see if the result of the second run in step 6
279
produces images of constant defocus through out each half of the tilt seriese.
280
281
282
* If the model is good, activate "keep the tilt axis parameters fixed" in the
283
tomography settings.
284
285
286
* Repeat step 2. Other than reset the learning, the tool produces an output of image
287
shift you should copy down and apply to the "tomo" preset in the future runs.
288
289
290
291
292
293
294
295
296
297
h2. Understand the results of the model fitting
298
299
300
301
In the above procedure does not yield a good model, it is necessary to study the graphs
302
output by the run to determine possible causes and derive the best fixed model that at least
303
work for most angles.
304
305
306
Figure 1 shows a typical result from the first dynamically fitted tilt series run. x and
307
y axes contains the tilt axis and are referenced to columns and rows on the images acquired.
308
z-axis is parallel to the electron beam. The "Prediction" data in the x and y-axis plot is
309
the extrapoloted result of the smooth curve fitting from preceeding (usually 4) tilts to the
310
current tilt angle. The "Position" is where features on the specimen end up to be relative
311
to the origin of the image shift at the scope as determined by correlation of images. Both
312
"Prediction" and "Position" are expressed in microns. The "Feature" is the shift of the
313
feature on the image. It corresponds to the difference between "Prediction" and "Position"
314
and is expressed in pixels of the image acquired. The z-axis "Prediction" is the correction
315
of defocus, i.e., specimen z-height change, the current model suggests and applied for each
316
tilt angle to keep the specimen focused. Originally, it is possible to measure defocus at
317
each tilt and obtain "Measured" values in this plot. However, the function is currently
318
disabled because the procedure does give accurate results.
319
320
321
According to the Zheng et. al. model, the tilt-axis can be characterized by three
322
parameters: phi, the angle between the tilt axis and the detector x axis (column), offset,
323
the distance between the center of the detector and the tilt axis on the xy plane, and z0,
324
the distance between the specimen from the xy plane that contains the tilt axis. If the tilt
325
axis is parallel to x-axis of the detector, and that the model does not change through the
326
tilt series, y-axis(row) "Prediction" and "Position" should remain constant. While the same
327
values for x-axis follows a cosine curve as it tilts if the only deviation from the ideal
328
model is an offset of the tilt axis from detector center in x-direction (non-zero offset).
329
On the other hand, if the only deviation the ideal model is an offset in z direction
330
(non-zero z0)
331
332
333
As a first tilt series used in the model fitting, the data is not fitted until it
334
accumulate enough data points and spead over more tilts. We arbitrarily select 30 degrees as
335
the starting angle where the fit is started. As can be seen in Figure 1, the model is
336
significantly different from the initial as z-axis "Prediction" jumps once the fitting is
337
started. In particular, the model z0 which corresponds to the offset of the specimen from
338
the tilt axis in z direction dominates deviation. Such problem is evident by the
339
near-straight line feature shift in x-axis ove more than 1 um. There is also some offset in
340
x direction since the "Prediction" and "Position" flattens somewhat as it approaches zero
341
tilt. The tilt axis has minimal tilt from x-axis since in the y-direction, despite the bump
342
at about 40 degrees, the range of variation in the whole tilt series is about 0.06
343
um.
344
345
346
Figure 1!http://emg.nysbc.org/software/leginon/images/images/tomoxyz1.png!
347
348
349
By activate "show model parameters" check box in the viewer, more plots are shown
350
(Figure 2). phi and "optical axis" offset remain fixed below 30 deg while z0 is recalculated
351
at each tilt. As you can see, the fitting of the former two parameters are not very stable
352
in this first trial. Therefore, we need to do the second run.
353
354
355
Figure 2!http://emg.nysbc.org/software/leginon/images/images/tomomodel1.png!
356
357
358
Figure 3 shows the z-axis plot of the second run since the x and y does not change
359
significantly. Figure 4 shows its corresponding model parameters. The zero and first tilt
360
uses the initial model and the rest are fitted dynamically. In this run, because more than
361
one tilt series is found in the memory since last "reset learning", the model fitting starts
362
after the first tilt image is acquired. model parameters are rather stable over the whole
363
tilt series and suggests that the tilt axis is tilted from x-axis by about -1.3 degree, and
364
offset from the center of the detector by -0.6 um, and the specimen is off from eucentric
365
height by about -0.7 um
366
367
368
Figure 3!http://emg.nysbc.org/software/leginon/images/images/tomoz2.png!
369
370
371
Figure 4!http://emg.nysbc.org/software/leginon/images/images/tomomodel2.png!
372
373
374
375
376
Although it is not possible to maintain perfect defocus prediction in the full range of
377
the tilt by using a fixed model, we found that the overall performance is better if an
378
average tilt axis model that works well in the mid-range tilt is used as a fixed model. For
379
example, for the behavior in the above figure, we choose:
380
381
382
383
384
Use these custom values as initial model and turn on "Keep the tilt axis parameters
385
fixed". If the result is good, judging by consistent defocus and target tracking through out
386
the tilt series, this model will be saved in the database as best model automatically, and
387
you can revert back to initialize with the model of "only this preset".
388
389
390
391
392
393
h2. Failure of xy feature tracking
394
395
396
397
feature tracking in x and y axes is a 2nd order polynomial fit of preceeding data
398
points. The default uses 5 data points. When a sudden jump occurs in the tracking error, it
399
tend to follow the trend of the last point. If the jump is a temporary clich in the
400
goniometer, this tend to over correct the tracking error and eventually loose track as shown
401
in Figure 5. A possible fix is to increase the number of data points in the fitting. This
402
can be set in the tomography setting "Smooth n tilts for defocus prediction". 4 in defocus
403
prediction is equivalent to 5 points (n+1) for xy tracking.
404
405
406
Figure 5!http://emg.nysbc.org/software/leginon/images/images/tomoxbad.png!
407
408
409
410
411
412
h2. Large tracking error between the first and second tilt
413
images
414
415
416
417
The first image in each tilt group of the tilt series at the "start" angle (normally 0
418
deg) and the second image at tilt of "step" angle from the "start" angle do not use the
419
fitted model.  It is assumed that the eucentric height judged by stage alpha wobbling in the
420
"Tomo Focus" node gives a stage height that the tracking of feature by such a small tilt
421
would be good enough.  In most cases this is a reasonable assumption.  However, we have had
422
experience of goniometer alignment problem where the assumption fails.  The symptom is
423
illustrated in Figure 6 below.  Note that the Feature tracking error is displayed as
424
percentage of the image length.
425
426
427
Figure 6!http://emg.nysbc.org/software/leginon/images/images/tomoz0bad.png!
428
429
430
This tilt series was taken with a starting angle of zero and at an image size of < 1
431
um.  As can be seen here, apart from the +2 and -2 degree tilts, the tracking error was less
432
than 2 % of the image.  Only the tracking of the feature between 0 and +/- 2 degrees are
433
large.  At close to 20 % error, this made the overlap between +/- 2 degrees unacceptable and
434
often cause popular alignment programs to misalign the two half of the series.
435
436
437
The first solution is of course to report it to your microscope service engineer.  When
438
we had this problem, many users noticed that it was difficult to adjust stage to eucentric
439
height manually with alpha wobbler.  Features jumped while the goniometer changed rotation
440
direction.  In addition, different magnitude of tilt range suggests different eucentric
441
heights.  It is not easy to fix this, so it might take a while.
442
443
444
Before the hardware is fixed physically, it is still possible collect tomograms.  The
445
model fitting of the overall curve in the above case gave z0 of +5 um through the whole tilt
446
series (Figure not shown).  Therefore, by moving the stage up by such an amount after the
447
stage-tilt-based autofocusing can bring us to the correct height for tomography.  This can
448
be acheived by saving the "tomo eucentric" focus current to the database, align rotation
449
center for this stage height and focus.  Then change the correction type of the
450
"Beam_Tilt_Fine" focusing step to "Stage Z".
451
452
453
454
455
456
h2. Failure of model-based correction
457
458
459
460
The model used in the defocus correction in Leginon tomography node is a very simplified
461
one. There are a few cases when the approach fails. Here are ones that we have
462
encountered:
463
464
465
466
h2. Y-axis looping
467
468
469
470
The microscope goniometer does not move on only the tilt axis. With its complex
471
structure, a common problem is that when the stage is highly tilt, the position slips in
472
the y-direction. This is known as looping. Figure 7 shows an example of this problem.
473
474
475
Figure 7!http://emg.nysbc.org/software/leginon/images/images/tomoxyloop.png!
476
477
478
While the x-axis position shifts monotonically as a stable model should be, the y-axis
479
in the positive tilt direction changes little from 0-30 degrees before it increases
480
rapidly after 30 degrees. Even though the tracking in xy plane is still good, the defocii
481
correction at these higher tilts may no longer be correct if the tilt axis parameters are
482
fitted dynamically. Figure 8 shows the model parameters of the same tilt series where the
483
fitted phi and offset starts to change above 30 degrees even though the tilt axis has not
484
moved according to the shrinking behavior of the images during the tilts. Note that in
485
this particular case the looping problem is still mild so that the over-correction is not
486
very strong. only a small slope change is resulted in z0 prediction. In worst cases, the
487
defocus over-correction is so large that the adjacent images can not correlate properly
488
and even the xy tracking would fail. The spikes around zero tilt is a display data sorting
489
error of the identical starting tilt of the two tilt groups.
490
491
492
Figure 8!http://emg.nysbc.org/software/leginon/images/images/tomomodelloop.png!
493
494
495
Other than asking microscope service engineer to fix the looping, one can find the
496
best fixed model in the series to apply to future tilt data collection. To make the fixed
497
model permantly saved to the database, follow these steps:
498
499
500
* tomography/settings/model>activate "keep the tilt axis parameters
501
fixed".
502
503
504
* tomography/settings/model>initialize the model with "custom values". Enter best
505
estimate of the fixed model. For example, in the positive direction, enter phi as
506
-2.17 degrees and axis offset as -1.52 um. since these are the stable values up to the
507
point the y-looping starts.
508
509
510
* tomography>collect a full tilt series. If the run is successful with good
511
tracking in all three axis, the model will be saved in the database for this
512
magnification.
513
514
515
* tomography/settings/model>From now on, you can initialize the model with "only
516
this preset" or "this preset and lower mags"
517
518
519
520
521
522
523
524
h2. Grid slips between the first and second tilt directions
525
526
527
528
When the holder does not hold the grid tightly, the grid slips to a different position
529
when the first tilt direction ends and the goniometer quickly returns to zero tilt.
530
Leginon is designed to adjust the target before the second tilt group starts. The default
531
setting for this function is to use only the parent image (i.e. one ancestor) where the
532
target comes from as reference. If the slip is larger than the size of the parent image,
533
the adjustment may fail, and a random target would be acquired in the second tilt
534
group.
535
536
537
Starting from Leginon 1.6, the target adjustment can be done with all ancestor images
538
of the target by choosing "all" in the acquisition part of the tomography node setting to
539
adjust target with all ancestors. The node "Taget Adjustment" limits the lowest
540
magnification that this target adjustment would go up in ancestry. The default is at 300x
541
so that the presence of the objective aperture does not create difference in the reaquired
542
ancestor image from its original.
543
544
545
546
547
548
h2. Strong and continuous specimen drift
549
550
551
552
The model used in Leginon considers any shift of feature in the image a result of tilt
553
axis not aligning to the center of the detector. With the phi and offset fixed, all errors
554
are accumulated in z0 and results in bad defocus correction. There is no solution to this
555
at the moment.