h2. Import Notes about Image Intensity Recorded through
6
Tomography Node
7
8
9
10 14 Amber Herold
Tomography node saves the images in a different format from other Acquisition nodes. By
11 1 Amber Herold
default, the flat-field correct CCD counts are multiplied by 10 and converted to signed
12
16-bit integer before the image is displayed and saved. This makes CCD counts of 3276.8 or
13
larger overflow to negatives. Other Leginon Acquisition images are saved as float without
14
manipulation.
15
16
17
To avoid this problem, find out what exposure time corresponds to the fractionated dose
18
from your tilt angle step and range and total dose and take an image at tomo preset with
19
such an exposure in Navigation node. You will need to reduce the total dose if a good
20
fraction of the counts are larger than 3200 even though it would not appear to be saturated
21
in the float scale without the 10x factor. Alternatively, change the scale factor in
22
Tomography node.
23
24
25
26
27
28
h2. Multiscale Imaging
29
30
31
32
*  Preset image shift alignment/beam shift alignment are the same as in MSI
33
application
34
35
36 5 Amber Herold
*  New dark/bright references should be reacquired for "tomo" preset that acquires the
37 1 Amber Herold
final data. It is best to do this at the same dose per tomography image calculated from
38
the total dose, the tilt parameters, and the dose measurement.
39
40
41
*  For best focusing result, perform autofocus at the same magnification as the
42
tomography data collection, align microscope well at the eucentric focus and the
43
rotation center and save them before data collection.
44
45
46
47
48
49
50
51
h2. Using Tomography Preview
52
53
54
55
*  Preview targets (pink) can be selected when selecting targets in "Tomography
56
Targeting"
57
58
59
*  When the targets are processed, targets that are of the type "preview" are
60
processed before focus and acquisition targets.
61
62
63
*  Tomography Preview node acquires a image at the preview target using "preview"
64
preset which should be set at minimal dose.
65
66
67
68
69
70
71
72
h2. Dose Measurement
73
74
75
76
If "Measure Dose before collection" is checked in Tomography node, the stage will be
77
moved to the reference target and a dose image of the "tomo" preset will be acquired (center
78
512x512 of whateven binning of the preset) before each tilt series if the interval between
79
the series is longer than the limit time set in the settings of Dose Measurement node. The
80
measured value will then be used to recalculate the proper exposure time for tomography
81
imaging.
82
83
84
For this function to behave properly, the followings should be done during
85
operation:
86
87
88
* One, and only one, "reference" target should be selected in either "Square
89
Targeting" or "Hole Targeting" or "Tomography Targeting" node. The reference target
90
should be of either a broken square or a empty hole if no broken square can be
91
found.
92
93
94
* "Measure Dose" before collection should be selected in Tomography node.
95
96
97
* "Exposure time max/min" in Tomography node should be in a range that can accommodate
98
the electron beam fluctuation over time.
99
100
101
102
103
104
105
106
h2. Align Zero Loss Peak
107
108
109
110
This function applies only to Gatan energy filter EFTEM. If "Align ZLP before
111
collection" is checked in Tomography node, the stage will be moved to the reference target
112
and starts the procedure to align zero loss peak before each tilt series if the interval
113
between the series is longer than the limit time set in the settings of Dose Measurement
114
node.
115
116
117
For this function to behave properly, the followings should be done during
118
operation:
119
120
121
* One, and only one, "reference" target should be selected in either "Square
122
Targeting" or "Hole Targeting" or "Tomography Targeting" node. The reference target
123
should be of either a broken square or a empty hole if no broken square can be found.
124
This is the same reference target used for dose measurement.
125
126
127
* "Align ZLP" before collection should be selected in Tomography node.
128
129
h2. What is a Good Tilt-Axis Model?
130
131
132
133
The goniometer-tilt-axis-based tracking model developed by Zheng et. al. corrects the
134 5 Amber Herold
specimen height (z-axis) by a change of defocus using measured shift of feature shifts in
135
the images (x and y-axes). The tracking in the x and y directions does not involve the use
136 1 Amber Herold
of such model, but is done by smooth curve fitting or preceding tilts. Therefore, to judge
137 5 Amber Herold
the adequacy of the model, one should check the resulting defocii of the images in the
138 1 Amber Herold
series remain unchanged.
139
140 22 Anchi Cheng
!tilt.png!
141 1 Amber Herold
142
On the other hand, the feature tracking in x and y is likely to fail only if the tilting
143
does not induce a smooth shift of the imaging feature a sudden drop of specimen position at
144
a particular tilt angle often throws off the smooth curve fitting. It is possible to reduce
145
such effect by increasing the number of data points included in the smoothing as set in the
146
model section of the tomography node settings window. Otherwise, the goniometer need to be
147
serviced.
148
149 19 Anchi Cheng
[[Troubles with Tomography|Tomography Trouble Shooting]] section covers many of the problems we have encountered.
150 1 Amber Herold
151
152
153 13 Amber Herold
154
155
156
157
______
158
159 21 Anchi Cheng
[[Set-up Before Running|< Set-up Before Running]] | [[Reading the tomography graphs| Reading the tomography graphs >]]