Project

General

Profile

Running the application » History » Version 3

Amber Herold, 06/18/2010 03:42 PM

1 1 Amber Herold
h1. Running the application
2
3
4
5
h2. Import Notes about Image Intensity Recorded through
6
Tomography Node
7
8
9
10
Tomography node save the images in a different format from other Acquisition nodes. By
11
default, the flat-field correct CCD counts are multiplied by 10 and converted to signed
12
16-bit integer before the image is displayed and saved. This makes CCD counts of 3276.8 or
13
larger overflow to negatives. Other Leginon Acquisition images are saved as float without
14
manipulation.
15
16
17
To avoid this problem, find out what exposure time corresponds to the fractionated dose
18
from your tilt angle step and range and total dose and take an image at tomo preset with
19
such an exposure in Navigation node. You will need to reduce the total dose if a good
20
fraction of the counts are larger than 3200 even though it would not appear to be saturated
21
in the float scale without the 10x factor. Alternatively, change the scale factor in
22
Tomography node.
23
24
25
26
27
28
h2. Multiscale Imaging
29
30
31
32
*  Preset image shift alignment/beam shift alignment are the same as in MSI
33
application
34
35
36
*  New dark/bright references should be reaquired for "tomo" preset that acquires the
37
final data. It is best to do this at the same dose per tomography image calculated from
38
the total dose, the tilt parameters, and the dose measurement.
39
40
41
*  For best focusing result, perform autofocus at the same magnification as the
42
tomography data collection, align microscope well at the eucentric focus and the
43
rotation center and save them before data collection.
44
45
46
47
48
49
50
51
h2. Using Tomography Preview
52
53
54
55
*  Preview targets (pink) can be selected when selecting targets in "Tomography
56
Targeting"
57
58
59
*  When the targets are processed, targets that are of the type "preview" are
60
processed before focus and acquisition targets.
61
62
63
*  Tomography Preview node acquires a image at the preview target using "preview"
64
preset which should be set at minimal dose.
65
66
67
68
69
70
71
72
h2. Dose Measurement
73
74
75
76
If "Measure Dose before collection" is checked in Tomography node, the stage will be
77
moved to the reference target and a dose image of the "tomo" preset will be acquired (center
78
512x512 of whateven binning of the preset) before each tilt series if the interval between
79
the series is longer than the limit time set in the settings of Dose Measurement node. The
80
measured value will then be used to recalculate the proper exposure time for tomography
81
imaging.
82
83
84
For this function to behave properly, the followings should be done during
85
operation:
86
87
88
* One, and only one, "reference" target should be selected in either "Square
89
Targeting" or "Hole Targeting" or "Tomography Targeting" node. The reference target
90
should be of either a broken square or a empty hole if no broken square can be
91
found.
92
93
94
* "Measure Dose" before collection should be selected in Tomography node.
95
96
97
* "Exposure time max/min" in Tomography node should be in a range that can accommodate
98
the electron beam fluctuation over time.
99
100
101
102
103
104
105
106
h2. Align Zero Loss Peak
107
108
109
110
This function applies only to Gatan energy filter EFTEM. If "Align ZLP before
111
collection" is checked in Tomography node, the stage will be moved to the reference target
112
and starts the procedure to align zero loss peak before each tilt series if the interval
113
between the series is longer than the limit time set in the settings of Dose Measurement
114
node.
115
116
117
For this function to behave properly, the followings should be done during
118
operation:
119
120
121
* One, and only one, "reference" target should be selected in either "Square
122
Targeting" or "Hole Targeting" or "Tomography Targeting" node. The reference target
123
should be of either a broken square or a empty hole if no broken square can be found.
124
This is the same reference target used for dose measurement.
125
126
127
* "Align ZLP" before collection should be selected in Tomography node.
128
129
130
131
132
133
134
135
h2. Low Magnification Model Fitting
136
137
138
139
The fitting of optical axis offset does not always works if the offset is so large that
140
the feature moves out of view with even a small tilt. In such a case, it is worth first
141
collect a tomography series at a lower magnification to define roughly the model.
142
143
144
At beginning of each session, or forced by the user, the model is initialized. By
145
default, at the initialization, Tomography node uses past fitting results that show good
146
agreement with the experimental data at the magnification of the preset used. If a good
147
model is not found, that from lower magnifications will be used. It is possible to force the
148
node to use a model fitted at a particular magnification by selecting it in
149
Tomography/Settings/Model.
150
151
152
Therefore, we recommend that, in case of fitting failure on good contrast images, the
153
followings should be done:
154
155
156 3 Amber Herold
#  Tomography/Settings/Image Acquisition> change the preset to "hl".
157
#  Tomography/Settings> adjust Tilt and Exposure parameters to match.
158
#  Acquire the tiltseries images.
159
#  If the tracking is good, change the preset back.
160
#  Tomography/Settings/Model> Initialize with the model of (the mag of "hl"
161 1 Amber Herold
preset).
162 3 Amber Herold
#  Acquire the tomo-series.
163
#  If tracking is good, change back to Initialize with the model of "this preset and
164 1 Amber Herold
lower" mag.
165
166
167
168
169
170
171
172
h2. What is a Good Tilt-Axis Model?
173
174
175
176
The goniometer-tilt-axis-based tracking model developped by Zheng et. al. corrects the
177
speciman height (z-axis) by a change of defocus using measured shift of feature shifts in
178
the images (x and y-axes). The tracking in the x and y directions does not involve the use
179
of such model, but is done by smooth curve fitting or preceeding tilts. Therefore, to judge
180
the adequacy of the model, one should check the resulting defocii of the images in the
181
series remain unchanged.
182
183
184
On the other hand, the feature tracking in x and y is likely to fail only if the tilting
185
does not induce a smooth shift of the imaging feature a sudden drop of specimen position at
186
a particular tilt angle often throws off the smooth curve fitting. It is possible to reduce
187
such effect by increasing the number of data points included in the smoothing as set in the
188
model section of the tomography node settings window. Otherwise, the goniometer need to be
189
serviced.
190
191
192
193
194
195
h2. Checking and Improving Tilt-Axis Model
196
197
198
199
Test runs should be done at the magnification, defocus, and tilts up to the highest
200
values planned for the real ones with a specimen with isotropic features in view and
201
adjusted to eucentric height. Tilt steps should be made to allow at least 15 images taken
202
per tilt direction.
203
204
205
206
h2. First Test Run: Fixed Model
207
208
209
210
The first time the tomography application is used on a microscope, all model
211
parameters are default to zero. A test run should be done with the specimen at eucentric
212
height and all default settings including "keep the tilt axis parameters fixed" activated.
213
If the feature tracking is good, and the defocii of the images in the tilt series do not
214
significantly changed, the goniometer behavior and camera alignment are close to ideal,
215
and no improvement of the model is required.
216
217
218
If the xy tracking deteriorates quickly in the first few tilts, the magnification
219
should be dropped so that tracking is possible through out the tilt series. Once a rough
220
model is established, refinement can then be done at higher magnifications.
221
222
223
224
225
226
h2. Runs to Improve the Model-Fitting
227
228
229
230
A complete tilt model fitting is difficult with small number of data points or at only
231
low tilts. Therefore, two tilt series need to be taken in order to get a good averaged
232
model that is later fixed.
233
234
235
* tomography/settings/model>deactivate "keep the tilt axis parameters
236
fixed".
237
238
239
* tomography/settings/model>initialize the model with "this preset only" unless
240
this is a refinement of <link linkend="Tomo_low_mag_fitting">a rough model obtained at
241
a lower magnification</link>, in which case, initialize the model with "this preset
242
and lower mags".
243
244
245
* tomography/toolbar>click on "reset learning" !http://emg.nysbc.org/software/leginon/images/icons/refresh.png! to have
246
a fresh start of tilt series data included in the fitting.
247
248
249
* Presets Manager>send the "tomo" preset to scope.
250
251
252
* tomography>Start the tilt series image collection by clicking on "Simulated
253
Target" tool.
254
255
256
* repeat 4 and 5 to dynamically fit the model the second time if the first one goes
257
to completion.
258
259
260
If the target shifts away during the run, reduce the magnification, repeat 1
261
through 9 in this procedure and then return to the higher, intended mag and refined
262
the model using the same procedure and proper setting for refinement as mentioned in
263
2.
264
265
266
* Check on web image viewer to see if the result of the second run in step 6
267
produces images of constant defocus through out each half of the tilt seriese.
268
269
270
* If the model is good, activate "keep the tilt axis parameters fixed" in the
271
tomography settings.
272
273
274
* Repeat step 2. Other than reset the learning, the tool produces an output of image
275
shift you should copy down and apply to the "tomo" preset in the future runs.
276
277
278
279
280
281
282
283
284
285
h2. Understand the results of the model fitting
286
287
288
289
In the above procedure does not yield a good model, it is necessary to study the graphs
290
output by the run to determine possible causes and derive the best fixed model that at least
291
work for most angles.
292
293
294
Figure 1 shows a typical result from the first dynamically fitted tilt series run. x and
295
y axes contains the tilt axis and are referenced to columns and rows on the images acquired.
296
z-axis is parallel to the electron beam. The "Prediction" data in the x and y-axis plot is
297
the extrapoloted result of the smooth curve fitting from preceeding (usually 4) tilts to the
298
current tilt angle. The "Position" is where features on the specimen end up to be relative
299
to the origin of the image shift at the scope as determined by correlation of images. Both
300
"Prediction" and "Position" are expressed in microns. The "Feature" is the shift of the
301
feature on the image. It corresponds to the difference between "Prediction" and "Position"
302
and is expressed in pixels of the image acquired. The z-axis "Prediction" is the correction
303
of defocus, i.e., specimen z-height change, the current model suggests and applied for each
304
tilt angle to keep the specimen focused. Originally, it is possible to measure defocus at
305
each tilt and obtain "Measured" values in this plot. However, the function is currently
306
disabled because the procedure does give accurate results.
307
308
309
According to the Zheng et. al. model, the tilt-axis can be characterized by three
310
parameters: phi, the angle between the tilt axis and the detector x axis (column), offset,
311
the distance between the center of the detector and the tilt axis on the xy plane, and z0,
312
the distance between the specimen from the xy plane that contains the tilt axis. If the tilt
313
axis is parallel to x-axis of the detector, and that the model does not change through the
314
tilt series, y-axis(row) "Prediction" and "Position" should remain constant. While the same
315
values for x-axis follows a cosine curve as it tilts if the only deviation from the ideal
316
model is an offset of the tilt axis from detector center in x-direction (non-zero offset).
317
On the other hand, if the only deviation the ideal model is an offset in z direction
318
(non-zero z0)
319
320
321
As a first tilt series used in the model fitting, the data is not fitted until it
322
accumulate enough data points and spead over more tilts. We arbitrarily select 30 degrees as
323
the starting angle where the fit is started. As can be seen in Figure 1, the model is
324
significantly different from the initial as z-axis "Prediction" jumps once the fitting is
325
started. In particular, the model z0 which corresponds to the offset of the specimen from
326
the tilt axis in z direction dominates deviation. Such problem is evident by the
327
near-straight line feature shift in x-axis ove more than 1 um. There is also some offset in
328
x direction since the "Prediction" and "Position" flattens somewhat as it approaches zero
329
tilt. The tilt axis has minimal tilt from x-axis since in the y-direction, despite the bump
330
at about 40 degrees, the range of variation in the whole tilt series is about 0.06
331
um.
332
333
334
Figure 1!http://emg.nysbc.org/software/leginon/images/images/tomoxyz1.png!
335
336
337
By activate "show model parameters" check box in the viewer, more plots are shown
338
(Figure 2). phi and "optical axis" offset remain fixed below 30 deg while z0 is recalculated
339
at each tilt. As you can see, the fitting of the former two parameters are not very stable
340
in this first trial. Therefore, we need to do the second run.
341
342
343
Figure 2!http://emg.nysbc.org/software/leginon/images/images/tomomodel1.png!
344
345
346
Figure 3 shows the z-axis plot of the second run since the x and y does not change
347
significantly. Figure 4 shows its corresponding model parameters. The zero and first tilt
348
uses the initial model and the rest are fitted dynamically. In this run, because more than
349
one tilt series is found in the memory since last "reset learning", the model fitting starts
350
after the first tilt image is acquired. model parameters are rather stable over the whole
351
tilt series and suggests that the tilt axis is tilted from x-axis by about -1.3 degree, and
352
offset from the center of the detector by -0.6 um, and the specimen is off from eucentric
353
height by about -0.7 um
354
355
356
Figure 3!http://emg.nysbc.org/software/leginon/images/images/tomoz2.png!
357
358
359
Figure 4!http://emg.nysbc.org/software/leginon/images/images/tomomodel2.png!
360
361
362
363
364
Although it is not possible to maintain perfect defocus prediction in the full range of
365
the tilt by using a fixed model, we found that the overall performance is better if an
366
average tilt axis model that works well in the mid-range tilt is used as a fixed model. For
367
example, for the behavior in the above figure, we choose:
368
369
370
371
372
Use these custom values as initial model and turn on "Keep the tilt axis parameters
373
fixed". If the result is good, judging by consistent defocus and target tracking through out
374
the tilt series, this model will be saved in the database as best model automatically, and
375
you can revert back to initialize with the model of "only this preset".
376
377
378
379
380
381
h2. Failure of xy feature tracking
382
383
384
385
feature tracking in x and y axes is a 2nd order polynomial fit of preceeding data
386
points. The default uses 5 data points. When a sudden jump occurs in the tracking error, it
387
tend to follow the trend of the last point. If the jump is a temporary clich in the
388
goniometer, this tend to over correct the tracking error and eventually loose track as shown
389
in Figure 5. A possible fix is to increase the number of data points in the fitting. This
390
can be set in the tomography setting "Smooth n tilts for defocus prediction". 4 in defocus
391
prediction is equivalent to 5 points (n+1) for xy tracking.
392
393
394
Figure 5!http://emg.nysbc.org/software/leginon/images/images/tomoxbad.png!
395
396
397
398
399
400
h2. Large tracking error between the first and second tilt
401
images
402
403
404
405
The first image in each tilt group of the tilt series at the "start" angle (normally 0
406
deg) and the second image at tilt of "step" angle from the "start" angle do not use the
407
fitted model.  It is assumed that the eucentric height judged by stage alpha wobbling in the
408
"Tomo Focus" node gives a stage height that the tracking of feature by such a small tilt
409
would be good enough.  In most cases this is a reasonable assumption.  However, we have had
410
experience of goniometer alignment problem where the assumption fails.  The symptom is
411
illustrated in Figure 6 below.  Note that the Feature tracking error is displayed as
412
percentage of the image length.
413
414
415
Figure 6!http://emg.nysbc.org/software/leginon/images/images/tomoz0bad.png!
416
417
418
This tilt series was taken with a starting angle of zero and at an image size of < 1
419
um.  As can be seen here, apart from the +2 and -2 degree tilts, the tracking error was less
420
than 2 % of the image.  Only the tracking of the feature between 0 and +/- 2 degrees are
421
large.  At close to 20 % error, this made the overlap between +/- 2 degrees unacceptable and
422
often cause popular alignment programs to misalign the two half of the series.
423
424
425
The first solution is of course to report it to your microscope service engineer.  When
426
we had this problem, many users noticed that it was difficult to adjust stage to eucentric
427
height manually with alpha wobbler.  Features jumped while the goniometer changed rotation
428
direction.  In addition, different magnitude of tilt range suggests different eucentric
429
heights.  It is not easy to fix this, so it might take a while.
430
431
432
Before the hardware is fixed physically, it is still possible collect tomograms.  The
433
model fitting of the overall curve in the above case gave z0 of +5 um through the whole tilt
434
series (Figure not shown).  Therefore, by moving the stage up by such an amount after the
435
stage-tilt-based autofocusing can bring us to the correct height for tomography.  This can
436
be acheived by saving the "tomo eucentric" focus current to the database, align rotation
437
center for this stage height and focus.  Then change the correction type of the
438
"Beam_Tilt_Fine" focusing step to "Stage Z".
439
440
441
442
443
444
h2. Failure of model-based correction
445
446
447
448
The model used in the defocus correction in Leginon tomography node is a very simplified
449
one. There are a few cases when the approach fails. Here are ones that we have
450
encountered:
451
452
453
454
h2. Y-axis looping
455
456
457
458
The microscope goniometer does not move on only the tilt axis. With its complex
459
structure, a common problem is that when the stage is highly tilt, the position slips in
460
the y-direction. This is known as looping. Figure 7 shows an example of this problem.
461
462
463
Figure 7!http://emg.nysbc.org/software/leginon/images/images/tomoxyloop.png!
464
465
466
While the x-axis position shifts monotonically as a stable model should be, the y-axis
467
in the positive tilt direction changes little from 0-30 degrees before it increases
468
rapidly after 30 degrees. Even though the tracking in xy plane is still good, the defocii
469
correction at these higher tilts may no longer be correct if the tilt axis parameters are
470
fitted dynamically. Figure 8 shows the model parameters of the same tilt series where the
471
fitted phi and offset starts to change above 30 degrees even though the tilt axis has not
472
moved according to the shrinking behavior of the images during the tilts. Note that in
473
this particular case the looping problem is still mild so that the over-correction is not
474
very strong. only a small slope change is resulted in z0 prediction. In worst cases, the
475
defocus over-correction is so large that the adjacent images can not correlate properly
476
and even the xy tracking would fail. The spikes around zero tilt is a display data sorting
477
error of the identical starting tilt of the two tilt groups.
478
479
480
Figure 8!http://emg.nysbc.org/software/leginon/images/images/tomomodelloop.png!
481
482
483
Other than asking microscope service engineer to fix the looping, one can find the
484
best fixed model in the series to apply to future tilt data collection. To make the fixed
485
model permantly saved to the database, follow these steps:
486
487
488
* tomography/settings/model>activate "keep the tilt axis parameters
489
fixed".
490
491
492
* tomography/settings/model>initialize the model with "custom values". Enter best
493
estimate of the fixed model. For example, in the positive direction, enter phi as
494
-2.17 degrees and axis offset as -1.52 um. since these are the stable values up to the
495
point the y-looping starts.
496
497
498
* tomography>collect a full tilt series. If the run is successful with good
499
tracking in all three axis, the model will be saved in the database for this
500
magnification.
501
502
503
* tomography/settings/model>From now on, you can initialize the model with "only
504
this preset" or "this preset and lower mags"
505
506
507
508
509
510
511
512
h2. Grid slips between the first and second tilt directions
513
514
515
516
When the holder does not hold the grid tightly, the grid slips to a different position
517
when the first tilt direction ends and the goniometer quickly returns to zero tilt.
518
Leginon is designed to adjust the target before the second tilt group starts. The default
519
setting for this function is to use only the parent image (i.e. one ancestor) where the
520
target comes from as reference. If the slip is larger than the size of the parent image,
521
the adjustment may fail, and a random target would be acquired in the second tilt
522
group.
523
524
525
Starting from Leginon 1.6, the target adjustment can be done with all ancestor images
526
of the target by choosing "all" in the acquisition part of the tomography node setting to
527
adjust target with all ancestors. The node "Taget Adjustment" limits the lowest
528
magnification that this target adjustment would go up in ancestry. The default is at 300x
529
so that the presence of the objective aperture does not create difference in the reaquired
530
ancestor image from its original.
531
532
533
534
535
536
h2. Strong and continuous specimen drift
537
538
539
540
The model used in Leginon considers any shift of feature in the image a result of tilt
541
axis not aligning to the center of the detector. With the phi and offset fixed, all errors
542
are accumulated in z0 and results in bad defocus correction. There is no solution to this
543
at the moment.