Setup Ceta to use Advanced TEM Scripting¶
At this point, only one camera can be controlled through Advanced TEM Scripting. If the system has both Falcon and Ceta camera, Falcon camera has to be removed from instruments.cfg and Ceta camera should use feicam rather than tia python module in instruments.cfg.
Example of proper instruments.cfg for diffraction recording¶
The exact module name (camera1,camera2,cetacam in this example) is not important. They just need to be unique. Class name needs to be exact match to the actual class.
#[camera1] #class: feicam.Falcon3 #zplane: 50 #height: 4096 #width: 4096 #[camera2] #class: feicam.Falcon3EC #zplane: 50 #height: 4096 #width: 4096 [cetacam] #class: tia.TIA_Ceta class: feicam.Ceta zplane: 50 height: 4096 width: 4096