h1. Setup Ceta to use Advanced TEM Scripting Diffraction mode image acquisition is not available through TIA Scripting. Advanced TEM Scripting must be used. *At At this point, only one camera can be controlled through Advanced TEM Scripting. If the system has both Falcon and Ceta camera, Falcon camera has to be removed from instruments.cfg and Ceta camera should use feicam rather than tia python module in instruments.cfg.* instruments.cfg. We recommend that you estabilish protocol to switch instruments.cfg used for MSI imaging that requires Falcon camera to the diffraction recording that uses Ceta camera only. h2. Example of proper instruments.cfg for diffraction recording The exact module name (cetacam in this example) is not important. They just need to be unique. Class name needs to be exact match to the actual class. <pre> [cetacam] class: feicam.Ceta zplane: 50 height: 4096 width: 4096 </pre> h2. Programs to open before Leginon Client: TIA. h2. Testing with pyscope In python command <pre> from pyscope import feicam g = feicam.Ceta() g.setExposureTime(200) g.getImage() </pre> {{include(pyscope_test_result array)}}