Setup Diffraction mode TEM instrument » History » Revision 2
Revision 1 (Anchi Cheng, 12/06/2019 02:37 PM) → Revision 2/6 (Anchi Cheng, 12/06/2019 03:46 PM)
h1. Setup Diffraction mode TEM instrument On microscope PC, with myami-ed git branch of installation h2. Edit Instruments.cfg used by leginon. Typically in C:\Program Files\myami\ Add diffraction instrument as another tem. The class name is your normal instrument, for example, Glacios, with Diffr prefix. Therefore, add <pre> [difftem] class: fei.DiffrGlacios cs: 2.7e-3 </pre> h2. Add items required for the instrument and workflow to your fei.cfg. This should have exists with instruments.cfg Decide where you will put all the AutoIt script executable. For example: C:\Program Files\AutoIt3\ Copy the items from your pyscope/fei.cfg.template to make sure you have the right syntax. h3. Under [camera] module, add <pre> AUTOIT_TUI_ACQUIRE_EXE_PATH = C:\Program Files\Autoit3\TuiAcquire.exe AUTOIT_TIA_EXPORT_SERIES_EXE_PATH = C:\Program Files\Autoit3\TiaExportSeries.exe AUTOIT_TIA_EXPORTED_DATA_DIR = C:\Users\supervisor\Documents\FEI\TIA\Exported Data </pre> h3. Add [beamstop] module and items needed at the end of some module. <pre> [beamstop] AUTOIT_IN_EXE_PATH = C:\Program Files\Autoit3\BeamstopIn.exe AUTOIT_OUT_EXE_PATH = C:\Program Files\Autoit3\BeamstopOut.exe </pre>