h1. Setup Diffraction mode TEM instrument 

 On microscope PC, with myami-ed git branch of installation 

 h2. Edit Instruments.cfg used by leginon.    Typically in C:\Program Files\myami\ 

 Add diffraction instrument as another tem.    The class name is your normal instrument, for example, Glacios, with Diffr prefix.    Therefore, add 
 <pre> 
 [difftem] 
 class: fei.DiffrGlacios 
 cs: 2.7e-3 
 </pre> 

 h2. Add items required for the instrument and workflow to your fei.cfg.    This should have exists with instruments.cfg 

 Decide where you will put all the AutoIt script executable.    For example: C:\Program Files\AutoIt3\ 
 Copy the items from your pyscope/fei.cfg.template to make sure you have the right syntax. 

 h3. Under [camera] module, add 
 <pre> 
 AUTOIT_TUI_ACQUIRE_EXE_PATH = C:\Program Files\Autoit3\TuiAcquire.exe 
 AUTOIT_TIA_EXPORT_SERIES_EXE_PATH = C:\Program Files\Autoit3\TiaExportSeries.exe 
 AUTOIT_TIA_EXPORTED_DATA_DIR = C:\Users\supervisor\Documents\FEI\TIA\Exported Data 
 </pre> 

 h3. Add [beamstop] module and items needed at the end of some module. 
 <pre> 
 [beamstop] 
 AUTOIT_IN_EXE_PATH =    C:\Program Files\Autoit3\BeamstopIn.exe 
 AUTOIT_OUT_EXE_PATH = C:\Program Files\Autoit3\BeamstopOut.exe 
 </pre>