Setup Diffraction mode TEM instrument

On microscope PC, with myami-ed git branch of installation

Edit Instruments.cfg used by leginon. Typically in C:\Program Files\myami\

Add diffraction instrument as another tem. The class name is your normal instrument, for example, Glacios, with Diffr prefix. Therefore, add

[difftem]
class: fei.DiffrGlacios
cs: 2.7e-3

Add items required for the instrument and workflow to your fei.cfg. This should have exists with instruments.cfg

Decide where you will put all the AutoIt script executable. For example: C:\Program Files\AutoIt3\
Copy the items from your pyscope/fei.cfg.template to make sure you have the right syntax.

Under [camera] module, add
AUTOIT_TUI_ACQUIRE_EXE_PATH = C:\Program Files\Autoit3\TuiAcquire.exe
AUTOIT_TIA_EXPORT_SERIES_EXE_PATH = C:\Program Files\Autoit3\TiaExportSeries.exe
AUTOIT_TIA_EXPORTED_DATA_DIR = C:\Users\supervisor\Documents\FEI\TIA\Exported Data

Add [beamstop] module and items needed at the end of some module.
[beamstop]
AUTOIT_IN_EXE_PATH =  C:\Program Files\Autoit3\BeamstopIn.exe
AUTOIT_OUT_EXE_PATH = C:\Program Files\Autoit3\BeamstopOut.exe