Setup df preset » History » Version 1
Anchi Cheng, 04/29/2020 06:11 PM
1 | 1 | Anchi Cheng | h1. Setup df preset |
---|---|---|---|
2 | |||
3 | h2. Microscope setup for df preset |
||
4 | |||
5 | 1. Center the beam and set in nanoprobe parallel illumication and eucentric focus in SA range imaging mode. |
||
6 | 2. Change spot size to larger number such as spot 10 and center this beam is needed through Direct Alignment reset Beam Shift. |
||
7 | 3. Change gun lens to get much weaker beam. For example from 3.3 to 7.1 reduces the beam current to 25%. You probably won't be able to see the beam any more on the flucam. |
||
8 | 4. Push the diffraction button on the control panel to put this in diffraction mode. |
||
9 | 5. Select desired camera length |
||
10 | 6. Push the eucentric button on the control panel to focus the beam on the diffraction plane. |
||
11 | 7. Insert beam stop and center the diffraction beam behind it. |
||
12 | |||
13 | h2. Camera setup for df preset |
||
14 | |||
15 | dimension: 2048 x 2048 |
||
16 | binning: 2 x 2 |
||
17 | exposure time: 1000 ms (not used during movie collection, but used for DPreview node) |
||