Project

General

Profile

Special Operation Preference setup » History » Version 1

Amber Herold, 04/22/2010 05:02 PM

1 1 Amber Herold
h1. Special Operation Preference setup
2
3
4
5
h2.  Taking final exposures at very high magnification
6
7
8
9
Hysteresis of image shift at very high mags such as 200 kx can be reduced by setting the
10
following configuration
11
12
13
*  Presets Manager/Settings>
14
15
16
Cycle Magnification Only = no
17
18
19
*  If <link linkend="QExpTarget">queuing is used at exposure targeting</link>,
20
21
22
Exposure Targeting/Settings>
23
24
25
Declare drift when queue submitted = yes
26
27
28
29
30
31
32
33
h2.  Avoiding drift correction that uses LM sq
34
preset
35
36
37
38
If the grid is flat and the offset of targets is small enough that it can be corrected
39
in the hole image, we recommand using the <link linkend="QExpTarget">Queuing option at
40
Exposure Targeting</link>. The sq preset is still used after Stage Z adjustment to correct
41
the hole targets but will not be used when queued exposure targets are processed.
42
43
44
If the offset of targets is minimal even after stage Z adjustment, set the following to
45
avoid using ancestor images obtained in LM for drift management purpose, queuing or
46
not.
47
48
49
*  Target Adjustment/Settings>
50
51
52
Minimum Magnification =Lowest Magnification you want to allow the node to use for
53
determining the new target.
54
55
56
57
58
59
60
61
h2.  Minimize the use of image shift for the final exposure (Approach 1)
62
63
64
65
Large image shift can cause sufficient beam tilt that creates problems such as bad
66
autofocus, beam shift, and loss of resolution (but probably only if better than 4 angstrum
67
is needed). Image shift is used in Leginon for targeting exposure because it is much more
68
accurate than a single movement by the specimen goniometer, even after it is modeled for
69
mechanical periodicity.
70
71
72
The accuracy of the stage movement is lower when it moves a long distance. Therefore,
73
using iterative stage movement can improve the targeting, and therefore reduce the amount of
74
image shift required for the final exposure. To use this feature, change the following
75
preference in Hole node (or Subsquare node in MSI-Raster).
76
77
78
It is important to know that in order to check whether the precision is reached, image
79
is taken at the preset at which the target is selected on (in this case, sq preset), so the
80
dose should be kept at minimal. On our microscope 0.2 micron precision can be obtained
81
within 2 to 3 moves.
82
83
84
Hole/Settings/Image Acquisition>
85
86
87
* Mover=navigator
88
89
90
* Navigator Target Tolerance = 2 e-7 m (or whatever tolerance you like): This sets the
91
goal for multiple movement
92
93
94
* Navigator Acceptable Tolerance = 1 e-6 m (or whatever tolerance you like): If
95
further movement causes an increase rather than a decrease of targeting accuracy, this
96
value determines whether the target is aborted or not.
97
98
99
* Final Image Shift = No
100
101
102
103
104
Navigation/Settings/Error Checking and Correction>
105
106
107
*  Preset cycle after each move = yes if your sq preset is in LM mode and you don't
108
mind wait a little longer; = no if you don't have hysteresis problem staying in sq
109
preset for the unknown amount of time during the iterative move.
110
111
112
113
114
115
116
117
h2.  Minimize the use of image shift for the final exposure (Approach 2)
118
119
120
121
A different approach is being developed to minimize image shift while maintaining final
122
targeting accuracy-A combination of stage movement and image shift. This approach can only
123
be used in limited cases such as targets selected for the tomography node or exposure node,
124
and is only experimental in the latter case.
125
126
127
Tomography or Exposure/Settings/Image Acquisition>
128
129
130
* Move Type = modeled stage position
131
132
133
* Mover=navigator
134
135
136
* Navigator Target Tolerance = 1 e-7 m (or whatever tolerance you like)
137
138
139
* Navigator Acceptable Tolerance = 1 e-6 m (or whatever tolerance you like)
140
141
142
* Final Image Shift = Yes
143
144
145
146
147
148
149
150
h2. Optimize autofocusing sequence
151
152
153
154
The focus sequence in Focus and Z Focus nodes can and should be optimize for specific
155
cases. The activated sequence in the default setting is good for a flat holey grid and is
156
meant for high resolution imaging at 50k x or higher. If accurate defocus is not important,
157
focusing once per grid square may be sufficient, for which all focus sequence in Focus node
158
can be deactivated and the same autofocus step can be added to Z Focus node instead.
159
160
161
There are two methods to determine the defocus automatically : Stage tilt (Equivalent to
162
stage wobbling) and Beam tilt. There are also two possible ways for correcting the defocus
163
measured: Defocus (Equivalent to turning the focusing knob on the scope and reset the
164
defocus) and Stage Z (Moving the stage to the zero defocus height).
165
166
167
The following observations at NRAMM may help you determine what is the best to
168
use
169
170
171
*  Both defocus determination method requires calculation correlation between two
172
images. Therefore, the magnification and the location of the autofocusing should include
173
area with contrast.
174
175
176
*  The higher the magnification, the more accurate the defocus determination but the
177
smaller the range of defocus the defocus determination can handle.
178
179
180
*  Beam Tilt is more reproducible and preferred method to determing defocus in HM
181
mode.
182
183
184
*  Stage Tilt (Wobbling) is the preferred method to determine defocus in LM
185
mode.
186
187
188
189
190
MSI-Raster Chapter contains an example of an <link linkend="MSI-R_auto_focus"
191
>alternative Z Focus node focus sequence</link> for grid that show little contrast at
192
medium magnifications.