h1. Sub-3Å proteasome image-shift data collection
2
3
h2. Instruments
4
5
SEMC@NYSBC Krios
6
# Scope: FEI Titan-Krios, 300 kV, micro probe at low mag, nano probe at high mags
7
# Camera: Gatan K2. mount below Falcon/Ceta. 1.21 Å at 29kx scope nominal mag
8
9
h2. Grid
10
11
* 1.2/1.3 C-flat carbon support; 400 mesh Copper grid.
12
13
[[Krios1-K2-NYSBC]]
14
15
h2. Application:
16
17
MSI-T2 Advanced (= MSI-T + "Beam Tilt Image" for coma-free alignment + "N2 Filling" for Krios auto nitrogen filler)
18
19
h2. Presets Manager:
20
21
No preset cycling (Krios has its own normalization).
22
23
h2. Queuing usage:
24
25
Hole Targeting: Queuing, target at the carbon area to cover four holes.
26
Exposure Targeting: *No Queuing*. automated targeting. See [[Exposure Targeting Set-up for MSi-T for Four Holes]]
27
28
h2. Move method to reach the targets:
29
30
Hole: [[Iterative_Stage_Movement|Navigator Iterative movement]]: Target Tolerance: 1e-7 m; Acceptable Tolerance: 3e-7 m
31
Focus:  Presets Manager image shift
32
Exposure: Presets Manager image shift
33
34
h2. Wait time before final exposure:
35
3 seconds
36
10 seconds addition for the first target