Project

General

Profile

Sub-3Å proteasome image-shift data collection » History » Version 7

Anchi Cheng, 08/18/2016 05:03 PM

1 1 Anchi Cheng
h1. Sub-3Å proteasome image-shift data collection
2
3 4 Anchi Cheng
This data collection obtained same quality data set as stage shift at 1.4x speed with the same presets. More speed gain relative to super-resolution mode data collection. Final reconstruction from the better of the two 3D subclasses (48,000 or ~57,000 cleaned-up particles) gave FSC 0.143 at 2.89 A.
4 2 Anchi Cheng
5
The speed gain relative to previous stage shift data collection is from
6
7
# removing iterative movement at Exposure node
8 4 Anchi Cheng
# removing queue processing at Exposure node since both need target adjustment and long wait time before stable imaging.
9 2 Anchi Cheng
# using counting mode. not super-resolution mode on K2.
10
11
*VERY IMPORTANT*: Measure the amount of beam tilt associated with the image-shift at your scope so that it is not compromising the data. Our scope is at 0.3 mrad at the 1.7 um image shift we applied in this experiment.
12
13
The Transfer Function will reduce all structure factor to 78% at these beamtilts assuming 0.167 mrad per 1 um image shift:
14
15
|image shift (um) |45 degree phase modification resolution(0.78 transfer function amplitude) (Angstroms)|
16
|1|2.4|
17
|2|3.0|
18
|3|3.5|
19
|4|3.8|
20
21 1 Anchi Cheng
h2. Instruments
22
23
SEMC@NYSBC Krios
24
# Scope: FEI Titan-Krios, 300 kV, micro probe at low mag, nano probe at high mags
25
# Camera: Gatan K2. mount below Falcon/Ceta. 1.21 Å at 29kx scope nominal mag
26
27
h2. Grid
28
29
* 1.2/1.3 C-flat carbon support; 400 mesh Copper grid.
30
31 3 Anchi Cheng
{{include(Krios1-K2-NYSBC)}}
32 1 Anchi Cheng
33
h2. Application:
34
35
MSI-T2 Advanced (= MSI-T + "Beam Tilt Image" for coma-free alignment + "N2 Filling" for Krios auto nitrogen filler)
36
37
h2. Presets Manager:
38
39
No preset cycling (Krios has its own normalization).
40
41
h2. Queuing usage:
42
43
Hole Targeting: Queuing, target at the carbon area to cover four holes.
44
Exposure Targeting: *No Queuing*. automated targeting. See [[Exposure Targeting Set-up for MSi-T for Four Holes]]
45
46
h2. Move method to reach the targets:
47
48
Hole: [[Iterative_Stage_Movement|Navigator Iterative movement]]: Target Tolerance: 1e-7 m; Acceptable Tolerance: 3e-7 m
49
Focus:  Presets Manager image shift
50
Exposure: Presets Manager image shift
51
52 4 Anchi Cheng
h2. Wait time before final exposure chosen to reduce maximal frame drift to average to 2 Angstroms :
53 1 Anchi Cheng
3 seconds
54
10 seconds addition for the first target
55 4 Anchi Cheng
56
h2. A trick to get Image shift targeting accuracy better:
57
58 7 Anchi Cheng
{{include(Scaling Stage Position Matrix Calibration)}}