Sub-3Å proteasome image-shift data collection » History » Version 7
Anchi Cheng, 08/18/2016 05:03 PM
1 | 1 | Anchi Cheng | h1. Sub-3Å proteasome image-shift data collection |
---|---|---|---|
2 | |||
3 | 4 | Anchi Cheng | This data collection obtained same quality data set as stage shift at 1.4x speed with the same presets. More speed gain relative to super-resolution mode data collection. Final reconstruction from the better of the two 3D subclasses (48,000 or ~57,000 cleaned-up particles) gave FSC 0.143 at 2.89 A. |
4 | 2 | Anchi Cheng | |
5 | The speed gain relative to previous stage shift data collection is from |
||
6 | |||
7 | # removing iterative movement at Exposure node |
||
8 | 4 | Anchi Cheng | # removing queue processing at Exposure node since both need target adjustment and long wait time before stable imaging. |
9 | 2 | Anchi Cheng | # using counting mode. not super-resolution mode on K2. |
10 | |||
11 | *VERY IMPORTANT*: Measure the amount of beam tilt associated with the image-shift at your scope so that it is not compromising the data. Our scope is at 0.3 mrad at the 1.7 um image shift we applied in this experiment. |
||
12 | |||
13 | The Transfer Function will reduce all structure factor to 78% at these beamtilts assuming 0.167 mrad per 1 um image shift: |
||
14 | |||
15 | |image shift (um) |45 degree phase modification resolution(0.78 transfer function amplitude) (Angstroms)| |
||
16 | |1|2.4| |
||
17 | |2|3.0| |
||
18 | |3|3.5| |
||
19 | |4|3.8| |
||
20 | |||
21 | 1 | Anchi Cheng | h2. Instruments |
22 | |||
23 | SEMC@NYSBC Krios |
||
24 | # Scope: FEI Titan-Krios, 300 kV, micro probe at low mag, nano probe at high mags |
||
25 | # Camera: Gatan K2. mount below Falcon/Ceta. 1.21 Å at 29kx scope nominal mag |
||
26 | |||
27 | h2. Grid |
||
28 | |||
29 | * 1.2/1.3 C-flat carbon support; 400 mesh Copper grid. |
||
30 | |||
31 | 3 | Anchi Cheng | {{include(Krios1-K2-NYSBC)}} |
32 | 1 | Anchi Cheng | |
33 | h2. Application: |
||
34 | |||
35 | MSI-T2 Advanced (= MSI-T + "Beam Tilt Image" for coma-free alignment + "N2 Filling" for Krios auto nitrogen filler) |
||
36 | |||
37 | h2. Presets Manager: |
||
38 | |||
39 | No preset cycling (Krios has its own normalization). |
||
40 | |||
41 | h2. Queuing usage: |
||
42 | |||
43 | Hole Targeting: Queuing, target at the carbon area to cover four holes. |
||
44 | Exposure Targeting: *No Queuing*. automated targeting. See [[Exposure Targeting Set-up for MSi-T for Four Holes]] |
||
45 | |||
46 | h2. Move method to reach the targets: |
||
47 | |||
48 | Hole: [[Iterative_Stage_Movement|Navigator Iterative movement]]: Target Tolerance: 1e-7 m; Acceptable Tolerance: 3e-7 m |
||
49 | Focus: Presets Manager image shift |
||
50 | Exposure: Presets Manager image shift |
||
51 | |||
52 | 4 | Anchi Cheng | h2. Wait time before final exposure chosen to reduce maximal frame drift to average to 2 Angstroms : |
53 | 1 | Anchi Cheng | 3 seconds |
54 | 10 seconds addition for the first target |
||
55 | 4 | Anchi Cheng | |
56 | h2. A trick to get Image shift targeting accuracy better: |
||