h1. Test Leginon on the Computer Controlling the Microscope
2
3
h2. Using a Gatan Camera
4
5
*  Open Digital Micrograph (or check that the Tietz software is closed).
6
7
*  insert the Gatan CCD in Digital Micrograph.
8
9
h2. Using a Tietz Camera
10
11
*  Close Tietz TCL program
12
13
*  Put camera switch box to select CCD.
14
15
h2. Make sure that you have [[Installation on the microscope computer|Leginon installed and configured ]]
16
17 10 Amber Herold
h2. Start Leginon on the Microscope computer and Import Magnifications
18 1 Amber Herold
19 5 Anchi Cheng
*  scope> Start Leginon (Either with the shortcut on the Desktop or follow the Start Menu-\Start\Programs\Leginon>Leginon)
20 1 Amber Herold
21
*  Leginon Setup Wizard> select "Create session" in session type. Even though you have created a session for simulation test, it is better to create a new session so that the path mapping difference betwee Windows and Linux does not cause confusion.
22 10 Amber Herold
**  Fill in session name and comments. Comments will appear in Viewers
23
**  Select instrument and project.
24
**  Image path is automatically selected based on your [[Configure leginon.cfg|leginon.cfg]] file.
25 1 Amber Herold
26
*  Choose "Finish" in the set up wizard to start the session.
27
28 5 Anchi Cheng
*  Leginon/Main menu/Application> select "Run...." and choose the application named "Manual", both "main" and "scope" launchers will be set to the scope computer, and then launch the application.
29 1 Amber Herold
30
*  Leginon/Node Selector> select Instrument node.
31
32 5 Anchi Cheng
*  Leginon/Instrument> check that TEM is selected to display the parameters.
33
34 1 Amber Herold
*  Left-click on the tool "Get Magnification". !http://emg.nysbc.org/software/leginon/images/icons/calculate.png!
35 5 Anchi Cheng
This will import the valid TEM magnification from the microscope into the database.
36 1 Amber Herold
37 10 Amber Herold
*Note:* During Leginon run an Application Error Message may pop up saying
38 1 Amber Herold
39 5 Anchi Cheng
<pre>
40 1 Amber Herold
41
Exception EAccessViolation in module adaExp.exe at ....
42 5 Anchi Cheng
Access violation at address xxxx in module 'adaExp.exe'. Read of address xxxx
43
44 1 Amber Herold
</pre>
45
46
Do NOT touch this message box, or leginon client will need restarting.
47
48
h2. Acquiring the first real image
49
50 10 Amber Herold
*  scope> make sure you have beam and something other than a blank to image on the CCD.
51 1 Amber Herold
52
*  Leginon/Application/Run> select Manual Application.
53
54
*  Leginon> From the list of nodes on the left panel, select Manual node.
55
56
*  Leginon/Manual> Open the "Setting" tool and configure the settings:
57
58 7 Anchi Cheng
> > Your TEM and CCDCamera should be available for selection.
59 1 Amber Herold
60 7 Anchi Cheng
> > Acquisition/Camera Configuration- use one of the preconfigured selection for binning and size such 512x8.
61 1 Amber Herold
62 7 Anchi Cheng
> > Save image to the database=yes.
63 1 Amber Herold
64 7 Anchi Cheng
> > Leave the rest of the configuration to default.
65 1 Amber Herold
66
*  You should see your first real image acquired through Leginon in the image
67
display panel.
68
69 11 Amber Herold
h2. Check on the [[Image_Viewers|Web viewer]]
70 1 Amber Herold
71
h2. Closing Leginon program on Windows
72
73 5 Anchi Cheng
The best way to close Leginon program is to close the python command window (normally black) that shows up when the Leginon GUI appears since Leginon GUI window on Windows is controlled through the python command window
74 1 Amber Herold
75 5 Anchi Cheng
If you first close the GUI window, the python process is not ended. Therefore, the latter become frozen and you will have to wait for TaskManager to ask you to close the broken process.
76 1 Amber Herold
77
h2. Problem solving for the first mircoscope test
78
79 8 Anchi Cheng
Most common problem at this point is that your TEM or CCDCamera does not show up in Leginon for use. It is caused by lack of proper functioning or non-communicating microscope scripting or CCD drivers to Leginon system's python wrapper in pyscope package. See [[Installation Troubleshooting]] and  "pyScope Bulletin Board regarding
80
testing if pyscope is working":http://emg.nysbc.org/bb/viewforum.php?t=246  for more.
81 1 Amber Herold
82
83 5 Anchi Cheng
If everything goes smoothly, test the network connection as in the next section and then do the same above image acquisition test through from the remote computer.
84 2 Amber Herold
85
______
86
87
[[Test Leginon with Simulator|< Test Leginon with Simulator]] | [[Test Network Connection Between Remote and Microscope Computers|Test Network Connection Between Remote and Microscope Computers >]]