1 | 1 | Amber Herold | h1. Test Leginon on the Computer Controlling the Microscope |
---|---|---|---|
2 | |||
3 | h2. Using a Gatan Camera |
||
4 | |||
5 | 15 | Amber Herold | # Open Digital Micrograph (or check that the Tietz software is closed). |
6 | |
||
7 | # insert the Gatan CCD in Digital Micrograph. |
||
8 | 1 | Amber Herold | |
9 | h2. Using a Tietz Camera |
||
10 | |||
11 | 15 | Amber Herold | # Close Tietz TCL program |
12 | |
||
13 | # Put camera switch box to select CCD. |
||
14 | 1 | Amber Herold | |
15 | h2. Make sure that you have [[Installation on the microscope computer|Leginon installed and configured ]] |
||
16 | |||
17 | h2. Start Leginon on the Microscope computer and Import Magnifications |
||
18 | |||
19 | 15 | Amber Herold | # scope> Start Leginon (Either with the shortcut on the Desktop or follow the Start Menu-\Start\Programs\Leginon>Leginon) |
20 | |
||
21 | # Leginon Setup Wizard> select "Create session" in session type. Even though you have created a session for simulation test, it is better to create a new session so that the path mapping difference betwee Windows and Linux does not cause confusion. |
||
22 | |
||
23 | 1 | Amber Herold | ** Fill in session name and comments. Comments will appear in Viewers |
24 | 15 | Amber Herold | |
25 | 10 | Amber Herold | ** Select instrument and project. |
26 | 15 | Amber Herold | |
27 | 1 | Amber Herold | ** Image path is automatically selected based on your [[Configure leginon.cfg|leginon.cfg]] file. |
28 | 15 | Amber Herold | |
29 | # Choose "Finish" in the set up wizard to start the session. |
||
30 | |
||
31 | # Leginon/Main menu/Application> select "Run...." and choose the application named "Manual", both "main" and "scope" launchers will be set to the scope computer, and then launch the application. |
||
32 | |
||
33 | # Leginon/Node Selector> select Instrument node. |
||
34 | |
||
35 | # Leginon/Instrument> check that TEM is selected to display the parameters. |
||
36 | |
||
37 | # Left-click on the tool "Get Magnification". !http://emg.nysbc.org/software/leginon/images/icons/calculate.png! |
||
38 | |
||
39 | 10 | Amber Herold | This will import the valid TEM magnification from the microscope into the database. |
40 | 15 | Amber Herold | |
41 | 5 | Anchi Cheng | *Note:* During Leginon run an Application Error Message may pop up saying |
42 | 15 | Amber Herold | |
43 | 1 | Amber Herold | <pre> |
44 | 5 | Anchi Cheng | |
45 | Exception EAccessViolation in module adaExp.exe at .... |
||
46 | 1 | Amber Herold | Access violation at address xxxx in module 'adaExp.exe'. Read of address xxxx |
47 | |||
48 | </pre> |
||
49 | 15 | Amber Herold | |
50 | 1 | Amber Herold | Do NOT touch this message box, or leginon client will need restarting. |
51 | |||
52 | 10 | Amber Herold | h2. Acquiring the first real image |
53 | 1 | Amber Herold | |
54 | 15 | Amber Herold | # scope> make sure you have beam and something other than a blank to image on the CCD. |
55 | |
||
56 | # Leginon/Application/Run> select Manual Application. |
||
57 | |
||
58 | # Leginon> From the list of nodes on the left panel, select Manual node. |
||
59 | |
||
60 | # Leginon/Manual> Open the "Setting" tool and configure the settings: |
||
61 | |
||
62 | 17 | Amber Herold | ** Your TEM and CCDCamera should be available for selection. |
63 | 15 | Amber Herold | |
64 | 17 | Amber Herold | ** Acquisition/Camera Configuration- use one of the preconfigured selection for binning and size such 512x8. |
65 | 15 | Amber Herold | |
66 | 17 | Amber Herold | ** Save image to the database=yes. |
67 | 15 | Amber Herold | |
68 | 17 | Amber Herold | ** Leave the rest of the configuration to default. |
69 | 15 | Amber Herold | |
70 | # You should see your first real image acquired through Leginon in the image |
||
71 | 13 | Amber Herold | display panel. |
72 | 1 | Amber Herold | |
73 | h2. Check on the [[Using_the_Web_viewer|Web viewer]] |
||
74 | |||
75 | 5 | Anchi Cheng | h2. Closing Leginon program on Windows |
76 | 1 | Amber Herold | |
77 | 5 | Anchi Cheng | The best way to close Leginon program is to close the python command window (normally black) that shows up when the Leginon GUI appears since Leginon GUI window on Windows is controlled through the python command window |
78 | 1 | Amber Herold | |
79 | If you first close the GUI window, the python process is not ended. Therefore, the latter become frozen and you will have to wait for TaskManager to ask you to close the broken process. |
||
80 | |||
81 | 14 | Amber Herold | h2. Problem solving for the first mircoscope test |
82 | 1 | Amber Herold | |
83 | 18 | Amber Herold | Most common problem at this point is that your TEM or CCDCamera does not show up in Leginon for use. It is caused by lack of proper functioning or non-communicating microscope scripting or CCD drivers to Leginon system's python wrapper in pyscope package. See [[Installation Troubleshooting]] and "pyScope Bulletin Board regarding testing if pyscope is working":http://emg.nysbc.org/boards/7/topics/607 for more. |
84 | 5 | Anchi Cheng | |
85 | 2 | Amber Herold | |
86 | If everything goes smoothly, test the network connection as in the next section and then do the same above image acquisition test through from the remote computer. |
||
87 | |||
88 | ______ |
||
89 | |||
90 | [[Test Leginon with Simulator|< Test Leginon with Simulator]] | [[Test Network Connection Between Remote and Microscope Computers|Test Network Connection Between Remote and Microscope Computers >]] |
||