Project

General

Profile

Test Leginon on the Computer Controlling the Microscope » History » Version 19

Anchi Cheng, 06/06/2012 06:31 PM

1 1 Amber Herold
h1. Test Leginon on the Computer Controlling the Microscope
2
3
h2. Using a Gatan Camera
4
5 19 Anchi Cheng
#  Open Digital Micrograph
6 15 Amber Herold
 
7
#  insert the Gatan CCD in Digital Micrograph.
8 1 Amber Herold
9
h2. Using a Tietz Camera
10
11
#  Close Tietz TCL program
12 19 Anchi Cheng
 
13
#  Put camera switch box to select CCD.
14
15
h2. Using a Direct Electron DE-12 Camera
16
17
#  Start DirectElectronAPI.exe and MicroManager
18
#  Set the desired *Frames Per Second* in MicroManager
19
#  Close MicroManager but leave DirectElectronAPI.exe running
20
21 15 Amber Herold
 
22
#  Put camera switch box to select CCD.
23 1 Amber Herold
24
h2. Make sure that you have [[Installation on the microscope computer|Leginon installed and configured ]]
25
26
h2. Start Leginon on the Microscope computer and Import Magnifications
27
28 15 Amber Herold
#  scope> Start Leginon (Either with the shortcut on the Desktop or follow the Start Menu-\Start\Programs\Leginon>Leginon)
29
 
30
#  Leginon Setup Wizard> select "Create session" in session type. Even though you have created a session for simulation test, it is better to create a new session so that the path mapping difference betwee Windows and Linux does not cause confusion.
31
 
32 1 Amber Herold
**  Fill in session name and comments. Comments will appear in Viewers
33 15 Amber Herold
 
34 10 Amber Herold
**  Select instrument and project.
35 15 Amber Herold
 
36 1 Amber Herold
**  Image path is automatically selected based on your [[Configure leginon.cfg|leginon.cfg]] file.
37 15 Amber Herold
 
38
#  Choose "Finish" in the set up wizard to start the session.
39
 
40
#  Leginon/Main menu/Application> select "Run...." and choose the application named "Manual", both "main" and "scope" launchers will be set to the scope computer, and then launch the application.
41
 
42
#  Leginon/Node Selector> select Instrument node.
43
 
44
#  Leginon/Instrument> check that TEM is selected to display the parameters.
45
 
46
#  Left-click on the tool "Get Magnification". !http://emg.nysbc.org/software/leginon/images/icons/calculate.png!
47
 
48 10 Amber Herold
This will import the valid TEM magnification from the microscope into the database.
49 15 Amber Herold
 
50 5 Anchi Cheng
*Note:* During Leginon run an Application Error Message may pop up saying
51 15 Amber Herold
 
52 1 Amber Herold
<pre>
53 5 Anchi Cheng
54
Exception EAccessViolation in module adaExp.exe at ....
55 1 Amber Herold
Access violation at address xxxx in module 'adaExp.exe'. Read of address xxxx
56
57
</pre>
58 15 Amber Herold
&nbsp;
59 1 Amber Herold
Do NOT touch this message box, or leginon client will need restarting.
60
61 10 Amber Herold
h2. Acquiring the first real image
62 1 Amber Herold
63 15 Amber Herold
#  scope> make sure you have beam and something other than a blank to image on the CCD.
64
&nbsp;
65
#  Leginon/Application/Run> select Manual Application.
66
&nbsp;
67
#  Leginon> From the list of nodes on the left panel, select Manual node.
68
&nbsp;
69
#  Leginon/Manual> Open the "Setting" tool and configure the settings:
70
&nbsp;
71 17 Amber Herold
** Your TEM and CCDCamera should be available for selection.
72 15 Amber Herold
&nbsp;
73 17 Amber Herold
** Acquisition/Camera Configuration- use one of the preconfigured selection for binning and size such 512x8.
74 15 Amber Herold
&nbsp;
75 17 Amber Herold
** Save image to the database=yes.
76 15 Amber Herold
&nbsp;
77 17 Amber Herold
** Leave the rest of the configuration to default.
78 15 Amber Herold
&nbsp;
79
#  You should see your first real image acquired through Leginon in the image
80 13 Amber Herold
display panel.
81 1 Amber Herold
82
h2. Check on the [[Using_the_Web_viewer|Web viewer]]
83
84 5 Anchi Cheng
h2. Closing Leginon program on Windows
85 1 Amber Herold
86 5 Anchi Cheng
The best way to close Leginon program is to close the python command window (normally black) that shows up when the Leginon GUI appears since Leginon GUI window on Windows is controlled through the python command window
87 1 Amber Herold
88
If you first close the GUI window, the python process is not ended. Therefore, the latter become frozen and you will have to wait for TaskManager to ask you to close the broken process.
89
90 14 Amber Herold
h2. Problem solving for the first mircoscope test
91 1 Amber Herold
92 18 Amber Herold
Most common problem at this point is that your TEM or CCDCamera does not show up in Leginon for use. It is caused by lack of proper functioning or non-communicating microscope scripting or CCD drivers to Leginon system's python wrapper in pyscope package. See [[Installation Troubleshooting]] and  "pyScope Bulletin Board regarding testing if pyscope is working":http://emg.nysbc.org/boards/7/topics/607  for more.
93 5 Anchi Cheng
94 2 Amber Herold
95
If everything goes smoothly, test the network connection as in the next section and then do the same above image acquisition test through from the remote computer.
96
97
______
98
99
[[Test Leginon with Simulator|< Test Leginon with Simulator]] | [[Test Network Connection Between Remote and Microscope Computers|Test Network Connection Between Remote and Microscope Computers >]]
100 1 Amber Herold
101
______