Test Leginon on the Computer Controlling the Microscope » History » Version 4

Anchi Cheng, 05/22/2010 10:42 PM

1 1 Amber Herold
h1. Test Leginon on the Computer Controlling the Microscope
2
3
4
5
h2. Using a Gatan Camera
6
7
8
9
10
11
*  Open Digital Micrograph (or check that the Tietz software is closed).
12
13
14
*  insert the Gatan CCD in Digital Micrograph.
15
16
17
18
19
20
21
22
23
24
h2. Using a Tietz Camera
25
26
27
28
29
30
*  Close Tietz TCL program
31
32
33
*  Put camera switch box to select CCD.
34
35 4 Anchi Cheng
h2. Make sure that you have [[Installation on the microscope computer|Leginon installed and configured ]]
36 1 Amber Herold
37
h2. Start Leginon on the Microscope computer & Import Magnifications
38
39
40
41
42
43
*  scope> Start Leginon (Either with the shortcut on the Desktop or follow the
44
Start Menu-\Start\Programs\Leginon>Leginon)
45
46
47
*  Leginon Setup Wizard> select "Create session" in session type. Even though you
48
have created a session for simulation test, if the other machine uses linux os, it is
49
better to create a new session so that the path mapping difference betwee Windows and
50
Linux does not cause confusion.
51
52
53
54
55
*  Fill in session name and comments. Comments will appear in Viewers
56
57
58
*  Select instrument and project, if project database is used. Image path is
59
automatically selected.
60
61
62
*  Image path is automatically selected based on your leginon.cfg file.
63
64
65
66
67
68
69
*  Choose "Finish" in the set up wizard to start the session.
70
71
72
*  Leginon/Main menu/Application> select "Run...." and choose the application
73
named "Manual", both "main" and "scope" launchers will be set to the scope computer,
74
and then launch the application.
75
76
77
*  Leginon/Node Selector> select Instrument node.
78
79
80
*  Leginon/Instrument> check that TEM is selected to display the
81
parameters.
82
83
84
*  Left-click on the tool "Get Magnification".
85
86
87
88
!http://emg.nysbc.org/software/leginon/images/icons/calculate.png!
89
90
91
92
This will import the valid TEM magnification from the microscope into the
93
database.
94
95
96
97
98
99
100
* During Leginon run an Application Error Message may pop up saying
101
102
<blockquote>
103
104
_"Exception EAccessViolation in module adaExp.exe at ...._
105
106
107
_Access violation at address xxxx in module 'adaExp.exe'. Read of address
108
xxxx"_
109
110
</blockquote>
111
112
Do NOT touch this message box, or leginon client will need restarting.
113
114
115
116
117
118
h2. Acquiring the first real image
119
120
121
122
123
124
*  scope> make sure you have beam and something other than a blank to image on
125
the CCD.
126
127
128
*  Leginon/Application/Run> select Manual Application.
129
130
131
*  Leginon> From the list of nodes on the left panel, select Manual node.
132
133
134
*  Leginon/Manual> Open the "Setting" tool and configure the settings:
135
136
137
Your TEM and CCDCamera should be available for selection.
138
139
140
Acquisition/Camera Configuration- use one of the preconfigured selection for
141
binning and size such 512x8.
142
143
144
Save image to the database=yes.
145
146
147
Leave the rest of the configuration to default.
148
149
150
*  You should see your first real image acquired through Leginon in the image
151
display panel.
152
153
154
155
156
157
158
159
160
161
h2. Check on the Web viewer
162
163
164
165
Follow the <link linkend="AppManual_webviewer">instruction in Manual Application
166
Chapter</link>.
167
168
169
170
171
172
h2. Closing Leginon program on Windows
173
174
175
176
The best way to close Leginon program is to close the python command window (normally
177
black) that shows up when the Leginon GUI appears since Leginon GUI window on Windows is
178
controlled through the python command window
179
180
181
If you first close the GUI window, the python process is not ended. Therefore, the
182
latter become frozen and you will have to wait for TaskManager to ask you to close the
183
broken process.
184
185
186
187
188
189
h2. Problem solving for the first mircoscope test
190
191
192
193
Most common problem at this point is that your TEM or CCDCamera does not show up in
194
Leginon for use. It is caused by lack of proper functioning or non-communicating microscope
195
scripting or CCD drivers to Leginon system's python wrapper in pyScope package. See <link
196
linkend="InstT_testrun">Installation Troubleshooting</link> and  "pyScope Bulletin Board regarding
197
"testing if pyScope is working"":http://emg.nysbc.org/bb/viewforum.php?t=246  for more.
198
199
200
If everything goes smoothly, test the network connection as in the next section and then
201
do the same above image acquisition test through from the remote computer.
202 2 Amber Herold
203
______
204
205
[[Test Leginon with Simulator|< Test Leginon with Simulator]] | [[Test Network Connection Between Remote and Microscope Computers|Test Network Connection Between Remote and Microscope Computers >]]
206
207
______