Tietz camera trouble shoot » History » Revision 1

Revision 1/2 | Next »
Anchi Cheng, 10/31/2013 08:21 PM


When Leginon initializes the camera, a program called CAMC4 should start automatically in the backgound. This is confirmed by the little camera icon showing up in the system tray in the lower right of the screen. Also, task manager can be started to confirm CAMC4 is running. Sometimes problems occur if more than one CAMC4 is running at the same time, so task manager will help identify this.

CAMC4 has its own configuration file "CAMC.ini" that will normally be set up already, but you can search for it on your computer and confirm that everything looks right. It will have configuration parameters for several cameras and only one of them will be activated. The only parameter that may need some adjustment is the readout geometry, which affects the rotation and mirror of the image.

If Leginon is not initializing the camera or not getting proper images, you should first test image acquisition using TVIPS software to make sure it is not a hardware issue. Use TCL or there is a test program called TestMFCCamera that can acquire an image and test the CAMC4 operation.

Updated by Anchi Cheng about 11 years ago · 1 revisions