Tips on Operation » History » Version 2
Amber Herold, 06/21/2010 01:45 PM
1 | 1 | Amber Herold | h1. Tips on Operation |
---|---|---|---|
2 | |||
3 | |||
4 | |||
5 | h2. Presets |
||
6 | |||
7 | |||
8 | |||
9 | * This application does not need sq, en, and ef preset. |
||
10 | |||
11 | |||
12 | * Keep fa and fc preset if accurate defocus is necessary. The default values assume fa and fc do not exist. |
||
13 | |||
14 | |||
15 | * You may also use lower magnification for gr preset to reduce the time required for acquire the grid atlas. |
||
16 | |||
17 | |||
18 | * It is more efficient to use lower mag and lower camera binning on the "hl" preset used for final acquisition |
||
19 | |||
20 | |||
21 | |||
22 | |||
23 | |||
24 | |||
25 | |||
26 | h2. Warm up objective lens after acquiring grid atlas |
||
27 | |||
28 | |||
29 | |||
30 | * With enough over raster overlap and large enough raster, there is no need to do this. |
||
31 | |||
32 | |||
33 | |||
34 | |||
35 | |||
36 | |||
37 | |||
38 | h2. Grid Targeting Robot Node |
||
39 | |||
40 | |||
41 | |||
42 | * Enter a unique name for the grid as Mosaic Label so that you can collect images on multiple grids in the same session and have the label displayed on the web viewer. |
||
43 | |||
44 | |||
45 | |||
46 | |||
47 | |||
48 | |||
49 | |||
50 | h2. Move Type |
||
51 | |||
52 | |||
53 | |||
54 | * Use modeled-stage position move type for all Acquisition and Focuser nodes because the movement is always large. |
||
55 | |||
56 | |||
57 | |||
58 | |||
59 | |||
60 | |||
61 | |||
62 | h2. Section Z Focus Node |
||
63 | |||
64 | |||
65 | |||
66 | * The focus target is selected on the grid atlas in "Raster Center Targeting" node. |
||
67 | |||
68 | |||
69 | 2 | Amber Herold | * It may help to follow the [[Improving Autofocusing|focus sequence suggested in MSI-Raster Application]], but use "gr" preset for Stage_Wobble step. If this is difficult due to the inserted objective aperture, you will still need to use a "sq" preset that is about 500-800x. |
70 | 1 | Amber Herold | |
71 | |||
72 | |||
73 | |||
74 | |||
75 | |||
76 | |||
77 | h2. Pre-exposure |
||
78 | |||
79 | |||
80 | |||
81 | * In certain cases, a pre-exposure is required for image stability. Adding "gr" preset to "Final Section" so the region is first exposed over a large area has been shown to help more than a local pre-exposure defined by "hl" preset. |
||
82 | |||
83 | |||
84 | |||
85 | |||
86 | |||
87 | |||
88 | |||
89 | h2. Queuing |
||
90 | |||
91 | |||
92 | |||
93 | * "Raster Center Targeting" is always in queuing mode. |
||
94 | |||
95 | |||
96 | |||
97 | |||
98 | ______ |
||
99 | |||
100 | [[Other special preference set-up|< Other special preference set-up]] | [[Leginon "RobotMSI-Section" Application|Leginon "RobotMSI-Section" Application ^]] |
||