Actions
Tivips f416 electron diffraction recording » History » Revision 4
« Previous |
Revision 4/6
(diff)
| Next »
Anchi Cheng, 12/12/2019 06:48 PM
TVIPS f416 electron diffraction recording¶
TVIPS F416 camera has rolling shutter movie collection capacity, accessible through EMMENU.
Leginon view port configuration setup¶
This setup will be used by Leginon to control view port 1. The configurations will be changed through EMMNENU scripting
Recorder configuration¶
Make sure recorder tool (red dot) is shown in toolbar¶
Reach toolbar view selection through EMMENU4/View/Toolbar>
Set recorder options¶
Click on the hammer next to the recorder tool to set these.
Basic settings
Advanced settings: save in tvips image set format
Leginon instruments.cfg¶
Replace
class:tietz.TietzF416
with
clas:tietz2.EmMenuF416
Set up tvips.cfg¶
- Copy pyscope/tvips.cfg.template to the same directory where instruments.cfg is read on TVIPS PC and rename it as tvips.cfg
- Match the recorder directory and default filename with what is in EMMENU4 recorder settings.
Updated by Anchi Cheng almost 5 years ago · 4 revisions