Troubles with Calibrations - Troubleshooting » History » Version 1
Amber Herold, 04/23/2010 09:16 AM
1 | 1 | Amber Herold | h1. Troubles with Calibrations - Troubleshooting |
---|---|---|---|
2 | |||
3 | |||
4 | |||
5 | |||
6 | * <link linkend="pixeloff">Available magnification from pixal size calibration is |
||
7 | different from that said on the scope (See Installation Trouble Shoot)</link> |
||
8 | |||
9 | |||
10 | * <link linkend="tr_matrx">Matrix calibration fails due to large pixel error</link> |
||
11 | |||
12 | |||
13 | * <link linkend="tr_matrx_beam">Beam shift calibration fails</link> |
||
14 | |||
15 | |||
16 | |||
17 | |||
18 | |||
19 | |||
20 | |||
21 | h2. Matrix calibration fails |
||
22 | |||
23 | |||
24 | |||
25 | * Display correlation map in the matrix calibration node during the calibration can help |
||
26 | identify the source of the failure. |
||
27 | |||
28 | |||
29 | Observations and Why: |
||
30 | * The image appears normal with most pixels in medium gray but the correlation |
||
31 | images show peaks moving randomly during the calibration - No image contrast for |
||
32 | correlation due to no object in view, at Gaussian focus, or bad image intensity either |
||
33 | too low or too high |
||
34 | |||
35 | |||
36 | * One or both images appear white or black and the correlation images show peaks |
||
37 | moving randomly during the calibration - Artifacts in the images such as spikes not |
||
38 | corrected by the corrector node that fool the correlation procedure to consider the |
||
39 | image is never moved or moved a lot. |
||
40 | |||
41 | |||
42 | * The correlation image give consistent multiple peaks or bars - Bright/Dark |
||
43 | correction images are not created. |
||
44 | |||
45 | |||
46 | * Both images and correlation peaks behave normally, i.e., there is one correlation |
||
47 | peak and its location corresponds well to the movement seen in the images and the set |
||
48 | shift fraction - Pixel calibration may be off. |
||
49 | |||
50 | |||
51 | |||
52 | |||
53 | |||
54 | Solutions: |
||
55 | * No object in view |
||
56 | |||
57 | |||
58 | * EM> move the specimen holder to an area with an isotropic feature well |
||
59 | centered within the CCD. |
||
60 | |||
61 | |||
62 | |||
63 | |||
64 | * No constrast |
||
65 | |||
66 | |||
67 | * Leginon/Presets Manager> Adjust defocus, and/or exposure time of the preset |
||
68 | that the calibraion uses and send the new values to scope. |
||
69 | |||
70 | |||
71 | |||
72 | |||
73 | * Spikes in the image |
||
74 | |||
75 | |||
76 | * Leginon/Presets Manager> Reduce exposure time? Might reduce number of |
||
77 | random event |
||
78 | |||
79 | |||
80 | * Leginon/Correction/Settings> Select the camera configuration of the problem |
||
81 | image. |
||
82 | |||
83 | |||
84 | * Leginon/Correction/Settings> Reduce the despike threshold and/or increase |
||
85 | the neighborhood size. The aim is to allow spikes with lower values and larger |
||
86 | footprint, respectively, to be removed. |
||
87 | |||
88 | |||
89 | |||
90 | |||
91 | * No bright/dark images |
||
92 | |||
93 | |||
94 | * Leginon/Correction/Settings> Select the camera configuration of the problem |
||
95 | image. |
||
96 | |||
97 | |||
98 | * Leginon/Correction> Acquire bright/dark images |
||
99 | |||
100 | |||
101 | |||
102 | |||
103 | * Pixel Calibration is off |
||
104 | |||
105 | |||
106 | * Leginon/Matrix/Settings> Increase Tolerance to allow the error to pass |
||
107 | through. |
||
108 | |||
109 | |||
110 | * Leginon/Matrix> redo the calibration. If it is successful, test the |
||
111 | calibration in Navigation |
||
112 | |||
113 | |||
114 | * If the calibration test is satisfactory, the result is acceptable. Pixel |
||
115 | calibration can be off since it is only an estimate for lower mags. You may want |
||
116 | to modify the pixel calibration entry according to the calculated value in matrix |
||
117 | calibration. Leaving the value as is will not cause major problem. |
||
118 | |||
119 | |||
120 | |||
121 | |||
122 | |||
123 | |||
124 | |||
125 | |||
126 | |||
127 | |||
128 | h2. Beam shift matrix calibration fails |
||
129 | |||
130 | |||
131 | |||
132 | * Other than the <link linkend="tr_matrx">failures common for all matrix |
||
133 | calibraion</link>, there are some additional reasons for beam shift calibration to |
||
134 | fail. |
||
135 | |||
136 | |||
137 | Observations and Why: |
||
138 | * The images shows not just a circle of the beam but a treak - The beam is too |
||
139 | intense that the CCD picks up the movement of the beam when the beam shutter is opened |
||
140 | and closed. |
||
141 | |||
142 | |||
143 | * The second images shows a shadow of the first- The CCD is saturated from the first |
||
144 | image acquisition. |
||
145 | |||
146 | |||
147 | * Not the whole image moves during the beam shift - This calibration have to see |
||
148 | only the beam not grid bars or objects on the grid . Neither of the latter will move |
||
149 | when the beam is shifted and therefore will show as false peak in the correlation |
||
150 | image. |
||
151 | |||
152 | |||
153 | * Part of the beam moves out of the view during beam shift - The applied shift or |
||
154 | the beam too large. |
||
155 | |||
156 | |||
157 | |||
158 | |||
159 | |||
160 | Solutions: |
||
161 | * Intense beam, saturated CCD |
||
162 | |||
163 | |||
164 | * EM> reduce the intensity by a change in spot size. |
||
165 | |||
166 | |||
167 | * Leginon> reduce the binning so that the exposure time can be increased |
||
168 | without saturating the camera. The longer exposure also reduces the relative |
||
169 | contribution of the shutter effect. |
||
170 | |||
171 | |||
172 | |||
173 | |||
174 | * Unmoved object |
||
175 | |||
176 | |||
177 | * EM> move to an empty square. |
||
178 | |||
179 | |||
180 | * EM> If not possible to exclude the grid bars in the empty square at low |
||
181 | magnifications, pull out the holder by about 3 cm (1 inch) and park it with a |
||
182 | spacer during the calibration. |
||
183 | |||
184 | |||
185 | |||
186 | |||
187 | * Beam shift too large |
||
188 | |||
189 | |||
190 | * Leginon/Matrix/Settings> Reduce shift fraction. |
||
191 | |||
192 | |||
193 | |||
194 | |||
195 | * Beam too large |
||
196 | |||
197 | |||
198 | * EM> Reduce the beam size. |
||
199 | |||
200 | |||
201 | * Leginon/Matrix> Acquire test images and adjust exposure time, beam spot |
||
202 | size etc. to match. |