Project

General

Profile

Troubles with Calibrations - Troubleshooting » History » Version 3

Eric Hou, 06/24/2010 02:52 PM

1 1 Amber Herold
h1. Troubles with Calibrations - Troubleshooting
2
3
h2. Matrix calibration fails
4
5 3 Eric Hou
*Display correlation map in the matrix calibration node during the calibration can help identify the source of the failure.
6 1 Amber Herold
7
Observations and Why:
8
9 3 Eric Hou
* The image appears normal with most pixels in medium gray but the correlation images show peaks moving randomly during the calibration - No image contrast for correlation due to no object in view, at Gaussian focus, or bad image intensity either too low or too high
10 1 Amber Herold
11 3 Eric Hou
* One or both images appear white or black and the correlation images show peaks moving randomly during the calibration - Artifacts in the images such as spikes not corrected by the corrector node that fool the correlation procedure to consider the image is never moved or moved a lot.
12 1 Amber Herold
13 3 Eric Hou
* The correlation image give consistent multiple peaks or bars - Bright/Dark correction images are not created.
14 1 Amber Herold
15 3 Eric Hou
* Both images and correlation peaks behave normally, i.e., there is one correlation peak and its location corresponds well to the movement seen in the images and the set shift fraction - Pixel calibration may be off.
16 1 Amber Herold
17
Solutions:
18 3 Eric Hou
19 1 Amber Herold
* No object in view
20 3 Eric Hou
 
21
** EM> move the specimen holder to an area with an isotropic feature well centered within the CCD.
22 1 Amber Herold
23
* No constrast
24 3 Eric Hou
 
25
* Leginon/Presets Manager> Adjust defocus, and/or exposure time of the preset that the calibraion uses and send the new values to scope.
26 1 Amber Herold
27
* Spikes in the image
28 3 Eric Hou
 
29
** Leginon/Presets Manager> Reduce exposure time? Might reduce number of random event
30
 
31
** Leginon/Correction/Settings> Select the camera configuration of the problem image.
32
 
33
** Leginon/Correction/Settings> Reduce the despike threshold and/or increase the neighborhood size. The aim is to allow spikes with lower values and larger footprint, respectively, to be removed.
34 1 Amber Herold
35
* No bright/dark images
36 3 Eric Hou
 
37
** Leginon/Correction/Settings> Select the camera configuration of the problem image.
38
 
39
** Leginon/Correction> Acquire bright/dark images
40 1 Amber Herold
41
* Pixel Calibration is off
42 3 Eric Hou
 
43
** Leginon/Matrix/Settings> Increase Tolerance to allow the error to pass through.
44
 
45
** Leginon/Matrix> redo the calibration. If it is successful, test the calibration in Navigation
46
 
47
** If the calibration test is satisfactory, the result is acceptable. Pixel calibration can be off since it is only an estimate for lower mags. You may want to modify the pixel calibration entry according to the calculated value in matrix calibration. Leaving the value as is will not cause major problem.
48 1 Amber Herold
49
h2. Beam shift matrix calibration fails
50
51 3 Eric Hou
* Other than the failures common for all matrix calibraion, there are some additional reasons for beam shift calibration to fail.
52 1 Amber Herold
53
Observations and Why:
54 3 Eric Hou
* The images shows not just a circle of the beam but a treak - The beam is too intense that the CCD picks up the movement of the beam when the beam shutter is opened and closed.
55 1 Amber Herold
56 3 Eric Hou
* The second images shows a shadow of the first- The CCD is saturated from the first image acquisition.
57 1 Amber Herold
58 3 Eric Hou
* Not the whole image moves during the beam shift - This calibration have to see only the beam not grid bars or objects on the grid . Neither of the latter will move when the beam is shifted and therefore will show as false peak in the correlation image.
59 1 Amber Herold
60 3 Eric Hou
* Part of the beam moves out of the view during beam shift - The applied shift or the beam too large.
61 1 Amber Herold
62
Solutions:
63 3 Eric Hou
64 1 Amber Herold
* Intense beam, saturated CCD
65 3 Eric Hou
 
66
** EM> reduce the intensity by a change in spot size.
67
 
68
** Leginon> reduce the binning so that the exposure time can be increased without saturating the camera. The longer exposure also reduces the relative contribution of the shutter effect.
69 1 Amber Herold
70
* Unmoved object
71 3 Eric Hou
 
72
** EM> move to an empty square.
73
 
74
** EM> If not possible to exclude the grid bars in the empty square at low magnifications, pull out the holder by about 3 cm (1 inch) and park it with a spacer during the calibration.
75 1 Amber Herold
76
* Beam shift too large
77 3 Eric Hou
 
78
** Leginon/Matrix/Settings> Reduce shift fraction.
79 1 Amber Herold
80
* Beam too large
81 3 Eric Hou
 
82
** EM> Reduce the beam size.
83
 
84
** Leginon/Matrix> Acquire test images and adjust exposure time, beam spot size etc. to match.
85 2 Amber Herold
86
______
87
88
[[Troubles with Focusing and Drift Check|< Troubles with Focusing and Drift Check]] | [[Troubles with Tomography|Troubles with Tomography >]]
89
90
______