h1. Troubles with Calibrations - Troubleshooting
2
3 5 Anchi Cheng
h2. No calibration Error during targeting.
4
5
Targeting in Leginon uses the calibration of the move type but with the calibration of the parent image (the image that you select the target on).  One exception is "Image Shift" Calibration which use the calibration of both the parent and the target acquisition presets.  The calibrations done at a particular preset are displayed in Presets Manager Node.
6
!preset_calibrations.png!
7
8
When Leginon complains that there is no calibration, 
9
10
# close the current MSI application
11
# restart Leginon (and its clients) if needed
12
# run the [[Leginon_"Calibrations"_Application|Calibration Application]]
13
# Find the required calibration instruction and follow it, starting from sending the preset with the missing calibration to the scope.
14
15 1 Amber Herold
h2. Matrix calibration fails
16
17 3 Eric Hou
*Display correlation map in the matrix calibration node during the calibration can help identify the source of the failure.
18 1 Amber Herold
19
Observations and Why:
20
21 3 Eric Hou
* The image appears normal with most pixels in medium gray but the correlation images show peaks moving randomly during the calibration - No image contrast for correlation due to no object in view, at Gaussian focus, or bad image intensity either too low or too high
22 1 Amber Herold
23 3 Eric Hou
* One or both images appear white or black and the correlation images show peaks moving randomly during the calibration - Artifacts in the images such as spikes not corrected by the corrector node that fool the correlation procedure to consider the image is never moved or moved a lot.
24 1 Amber Herold
25 3 Eric Hou
* The correlation image give consistent multiple peaks or bars - Bright/Dark correction images are not created.
26 1 Amber Herold
27 3 Eric Hou
* Both images and correlation peaks behave normally, i.e., there is one correlation peak and its location corresponds well to the movement seen in the images and the set shift fraction - Pixel calibration may be off.
28 1 Amber Herold
29
Solutions:
30 3 Eric Hou
31 1 Amber Herold
* No object in view
32 3 Eric Hou
 
33
** EM> move the specimen holder to an area with an isotropic feature well centered within the CCD.
34 1 Amber Herold
35
* No constrast
36 3 Eric Hou
 
37 4 Eric Hou
** Leginon/Presets Manager> Adjust defocus, and/or exposure time of the preset that the calibraion uses and send the new values to scope.
38 1 Amber Herold
39
* Spikes in the image
40 3 Eric Hou
 
41
** Leginon/Presets Manager> Reduce exposure time? Might reduce number of random event
42
 
43
** Leginon/Correction/Settings> Select the camera configuration of the problem image.
44
 
45
** Leginon/Correction/Settings> Reduce the despike threshold and/or increase the neighborhood size. The aim is to allow spikes with lower values and larger footprint, respectively, to be removed.
46 1 Amber Herold
47
* No bright/dark images
48 3 Eric Hou
 
49
** Leginon/Correction/Settings> Select the camera configuration of the problem image.
50
 
51
** Leginon/Correction> Acquire bright/dark images
52 1 Amber Herold
53
* Pixel Calibration is off
54 3 Eric Hou
 
55
** Leginon/Matrix/Settings> Increase Tolerance to allow the error to pass through.
56
 
57
** Leginon/Matrix> redo the calibration. If it is successful, test the calibration in Navigation
58
 
59
** If the calibration test is satisfactory, the result is acceptable. Pixel calibration can be off since it is only an estimate for lower mags. You may want to modify the pixel calibration entry according to the calculated value in matrix calibration. Leaving the value as is will not cause major problem.
60 1 Amber Herold
61
h2. Beam shift matrix calibration fails
62
63 3 Eric Hou
* Other than the failures common for all matrix calibraion, there are some additional reasons for beam shift calibration to fail.
64 1 Amber Herold
65
Observations and Why:
66 3 Eric Hou
* The images shows not just a circle of the beam but a treak - The beam is too intense that the CCD picks up the movement of the beam when the beam shutter is opened and closed.
67 1 Amber Herold
68 3 Eric Hou
* The second images shows a shadow of the first- The CCD is saturated from the first image acquisition.
69 1 Amber Herold
70 3 Eric Hou
* Not the whole image moves during the beam shift - This calibration have to see only the beam not grid bars or objects on the grid . Neither of the latter will move when the beam is shifted and therefore will show as false peak in the correlation image.
71 1 Amber Herold
72 3 Eric Hou
* Part of the beam moves out of the view during beam shift - The applied shift or the beam too large.
73 1 Amber Herold
74
Solutions:
75 3 Eric Hou
76 1 Amber Herold
* Intense beam, saturated CCD
77 3 Eric Hou
 
78
** EM> reduce the intensity by a change in spot size.
79
 
80
** Leginon> reduce the binning so that the exposure time can be increased without saturating the camera. The longer exposure also reduces the relative contribution of the shutter effect.
81 1 Amber Herold
82
* Unmoved object
83 3 Eric Hou
 
84
** EM> move to an empty square.
85
 
86
** EM> If not possible to exclude the grid bars in the empty square at low magnifications, pull out the holder by about 3 cm (1 inch) and park it with a spacer during the calibration.
87 1 Amber Herold
88
* Beam shift too large
89 3 Eric Hou
 
90
** Leginon/Matrix/Settings> Reduce shift fraction.
91 1 Amber Herold
92
* Beam too large
93 3 Eric Hou
 
94
** EM> Reduce the beam size.
95
 
96
** Leginon/Matrix> Acquire test images and adjust exposure time, beam spot size etc. to match.
97 2 Amber Herold