Project

General

Profile

Troubles with Focusing and Drift Check » History » Version 2

Amber Herold, 04/27/2010 02:18 PM

1 1 Amber Herold
h1. Troubles with Focusing and Drift Check
2
3
4
5
6
* <link linkend="driftforever">Focuser and drift manager toggle back and forth
7
forever</link>
8
9
10
* <link linkend="focfail">Autofocus failure-high mags</link>
11
12
13
* <link linkend="focfailz">Autofocus to eucentric (Z Focus node) failure</link>
14
15
16
* <link linkend="focdark">Autofocus image of holes become dark at one or more beam
17
tilts</link>
18
19
20
* <link linkend="focfail">The specimen appears over-dosed in the high magnification images
21
(while using a Gatan CCD)</link>
22
23
24
25
26
27
28
29
h2. Focuser and drift manager toggle back and forth
30
forever
31
32
33
34
Commonly Why: Threshold set in focuser/autofocus is significantly smaller than that in
35
drift manager in absolute scale. This is a bug in Leginon to some extend caused by a small
36
movement of the image when the a new state of the beam is set so that the focuser records
37
bigger drift than the drift manager.
38
39
40
Solution: Increase Focuser/Autofocus Drift Threshold and/or increase the wait time
41
between checking drift in Drift Manager.
42
43
44
Still not working: Despike might not work in the corrector
45
46
47
Solution: Consider increasing Despike neighbor size (odd number only) or decreasing
48
despike threshold.
49
50
51
Still not working: You are on a day with a lot of cosmic events that cause the cross
52
correlation fails.
53
54
55
Solution: Finding a way to reduce exposure time might help as it reduces the possibility
56
of cosmic ray hitting the CCD.
57
58
59
60
61
62
h2. Failed autofocus in Focus node
63
64
65
66
Commonly Why:
67
* The actual defocus is outside of measurable range.
68
69
70
* No image contrast for correlation due to thick ice, no ice and support film, or at
71
Gaussian focus
72
73
74
* Stage drift, charging, or ice melts during the focus measurement. All three cause
75
change in images not accountable by beam tilt induced image shift.
76
77
78
* Artifacts in the images such as spikes not corrected by the corrector node that
79
fool the correlation procedure to consider the image is never moved or moved a
80
lot.
81
82
83
* Bad rotaion center and beam tilt pivot point alignment
84
85
86
* <link linkend="focdark">Objective aperture is off alignment</link>.
87
88
89
* Bad calibration
90
91
92
93
94
95
Solutions:
96
* Out of range:
97
98
99
* Make sure that the lower mag autofocus, i.e., Z focus node, is preformed
100
successfully by observing its progress, and, if necessary, adjust the stage height
101
in its Manual focus tool.
102
103
104
* Make sure that Z Focus node autofocus accuracy is compatible with the Focus
105
node. For example, Z Focus at 5000x has an accuracy of +/- 5 um and Focus at
106
50,000x can handle range up to about +/- 10 um.
107
108
109
* Choose a flater grid square.
110
111
112
113
114
* For thick ice
115
116
117
* Leginon/Focus/Settings/Autofocus> Increase the ice melt time in autofocus
118
setting
119
120
121
* Leginon/Exposure Targeting/Hole Targeting Setting/Target Template/Focus
122
Template Thickness> Adjust allowed ice thickness values to pick focus target in
123
area with thinner ice
124
125
126
* Leginon/Presets Manager> reduce "fa" preset CCD binning and/or increase
127
exposure time.
128
129
130
* Bad rotaion center and beam tilt pivot point alignment
131
132
133
* Bad calibration
134
135
136
137
138
* For stage drift during the procedure
139
140
141
* Leginon/Focus/Settings/Autofocus> Reduce drift check threshold.
142
143
144
145
146
* For charging during the procedure
147
148
149
* Leginon/Exposure Targeting> Pick targets away from ice chunk, grid bar, and
150
large ice gradient.
151
152
153
154
155
* For ice melting during the procedure
156
157
158
* Leginon/Focus/Settings/Autofocus> Increase ice melt time
159
160
161
162
163
* For image spikes
164
165
166
* Leginon/Correction> Adjust despike parameters
167
168
169
* Leginon/Correction> Reduce exposure time? Might reduce number of random
170
event
171
172
173
174
175
176
177
178
179
180
181
h2. Failed autofocus in Z Focus node
182
183
184
185
Commonly Why:
186
* The stage height relative to Eucentric point is outside of measurable range due to
187
very uneven grid or stage far away from U-center.
188
189
190
* Empty area or area with very thick ice, or at Gaussian focus
191
192
193
* Objective aperture is off alignment.
194
195
196
* Bad recorded eucentric focus.
197
198
199
* Bad rotaion center and beam tilt pivot point alignment
200
201
202
* Bad calibration
203
204
205
To identify which is the source, observe the image, correlation, and peak
206
during beam tilt autofocusing. Also check in the log the residual value (so called min) of
207
the fitting.
208
209
210
* Image has very good contrast but features are completely different from one tilt
211
image to the next-Out of range.
212
213
214
* No image contrast or correlation peaks move randomly-empty or thick ice, or at
215
Gaussian focus
216
217
218
* Contrast reverse in the holes relative to the support matrix in one or more beam
219
tilt images-Objective aperture is off alignment.
220
221
222
* If you know that the stage is at U-center, but the images appear to be strongly
223
defocused-Recorded eucentric focus is incorrect.
224
225
226
* Correlation peaks are well defined but fitting fails-Bad rotaion center and beam
227
tilt pivot point alignment, or bad calibration.
228
229
230
231
232
Solutions:
233
* Out of range:
234
235
236
* U-center the stage before the run and when Z focus fails.
237
238
239
* Activate Manual Check after autofocus to make adjustment in stage Z
240
height.
241
242
243
* For expert user: Add another focus step before the current autofocus step at
244
even lower magnification for a larger range usability. For example, use sq preset.
245
You will need to perform a calibration with Beam Tilt node in the Calibration
246
application.
247
248
249
250
251
* Empty area or thick ice:
252
253
254
* Pick better focus target.
255
256
257
* Leginon/Z Focus/Hole Targeting Settings/Ice Thickness Threshold/Focus Hole
258
Selection> Consider using "Good hole" rather than "Any hole".
259
260
261
262
263
* At Guassian focus:
264
265
266
* There is no problem, don't worry.
267
268
269
270
271
* Bad recorded eucentric focus
272
273
274
*  Leginon> send to scope the preset used to autofocus in Z Focus node (i.e.,
275
hl in MSI).
276
277
278
*  TEM> Adjust the stage to U-center.
279
280
281
*  TEM> Focus the image and reset defocus.
282
283
284
*  Leginon/Z Focus> start Manual focus tool.
285
286
287
*  Leginon/Z Focus/Manual Focus> send 0 defocus to scope.
288
289
290
*  Leginon/Z Focus/Manual Focus> save "Eucentric focus from
291
Instrument".
292
293
294
295
296
* Bad rotation center and beam tilt pivot point alignment:
297
298
299
* Make sure you are at the eucentric focus during the alignment.
300
301
302
303
304
* Bad alignment on objective aperture:
305
306
307
* <link linkend="focdark">Center the objective aperture in diffraction
308
mode</link>.
309
310
311
312
313
314
315
316
317
318
319
h2. Objects reverse contrast during autofocus
320
321
322
323
In this case of autofocus failure, objects such as holes change from lighter than matrix
324
to darker as if a dark-field image is collected during the autofocus beam tilts.
325
326
327
Commonly Why:
328
* The objective aperture is off.
329
330
331
* Bad rotation center and beam tilt pivot point alignment.
332
333
334
335
336
337
Solutions:
338
* Align objective aperture at mags where autofocus is performed in HM mode. Here is
339
how:
340
341
342
*  TEM>Make sure the rotation center is correct.
343
344
345
*  TEM>Change the mag to that of the autofocus preset that shows the
346
problem.
347
348
349
or
350
351
352
Leginon/Presets Manager>Send the autofocus preset.
353
354
355
*  Imaging an area that have scattering matter such as with ice or
356
carbon.
357
358
359
*  TEM>Press the diffraction button to go into diffraction mode.
360
361
362
*  TEM>Use the Mag adjustment knob to change diffraction distance to ~1
363
m.
364
365
366
*  TEM>You should see a bright and condensed image of your imaging area which
367
would condenses to a point if you are right at focus (adjust with focus knob). The
368
faint disk outside of the bright area defines the opening of the objective
369
aperture. Center the image using the MULTIFUNCTION KNOBS if you can not see the
370
edge. Move the STAGE to area with more scattering matter if the faint disk is not
371
visible.
372
373
374
*  TEM>Physically move the aperture so the faint disk is centered on the
375
bright image.
376 2 Amber Herold
377
378
379
______
380
381
[[Troubles with Targeting|< Troubles with Targeting]] | [[Troubles with Calibrations - Troubleshooting|Troubles with Calibrations >]]
382
383
______