Troubles with Focusing and Drift Check » History » Version 2
Amber Herold, 04/27/2010 02:18 PM
1 | 1 | Amber Herold | h1. Troubles with Focusing and Drift Check |
---|---|---|---|
2 | |||
3 | |||
4 | |||
5 | |||
6 | * <link linkend="driftforever">Focuser and drift manager toggle back and forth |
||
7 | forever</link> |
||
8 | |||
9 | |||
10 | * <link linkend="focfail">Autofocus failure-high mags</link> |
||
11 | |||
12 | |||
13 | * <link linkend="focfailz">Autofocus to eucentric (Z Focus node) failure</link> |
||
14 | |||
15 | |||
16 | * <link linkend="focdark">Autofocus image of holes become dark at one or more beam |
||
17 | tilts</link> |
||
18 | |||
19 | |||
20 | * <link linkend="focfail">The specimen appears over-dosed in the high magnification images |
||
21 | (while using a Gatan CCD)</link> |
||
22 | |||
23 | |||
24 | |||
25 | |||
26 | |||
27 | |||
28 | |||
29 | h2. Focuser and drift manager toggle back and forth |
||
30 | forever |
||
31 | |||
32 | |||
33 | |||
34 | Commonly Why: Threshold set in focuser/autofocus is significantly smaller than that in |
||
35 | drift manager in absolute scale. This is a bug in Leginon to some extend caused by a small |
||
36 | movement of the image when the a new state of the beam is set so that the focuser records |
||
37 | bigger drift than the drift manager. |
||
38 | |||
39 | |||
40 | Solution: Increase Focuser/Autofocus Drift Threshold and/or increase the wait time |
||
41 | between checking drift in Drift Manager. |
||
42 | |||
43 | |||
44 | Still not working: Despike might not work in the corrector |
||
45 | |||
46 | |||
47 | Solution: Consider increasing Despike neighbor size (odd number only) or decreasing |
||
48 | despike threshold. |
||
49 | |||
50 | |||
51 | Still not working: You are on a day with a lot of cosmic events that cause the cross |
||
52 | correlation fails. |
||
53 | |||
54 | |||
55 | Solution: Finding a way to reduce exposure time might help as it reduces the possibility |
||
56 | of cosmic ray hitting the CCD. |
||
57 | |||
58 | |||
59 | |||
60 | |||
61 | |||
62 | h2. Failed autofocus in Focus node |
||
63 | |||
64 | |||
65 | |||
66 | Commonly Why: |
||
67 | * The actual defocus is outside of measurable range. |
||
68 | |||
69 | |||
70 | * No image contrast for correlation due to thick ice, no ice and support film, or at |
||
71 | Gaussian focus |
||
72 | |||
73 | |||
74 | * Stage drift, charging, or ice melts during the focus measurement. All three cause |
||
75 | change in images not accountable by beam tilt induced image shift. |
||
76 | |||
77 | |||
78 | * Artifacts in the images such as spikes not corrected by the corrector node that |
||
79 | fool the correlation procedure to consider the image is never moved or moved a |
||
80 | lot. |
||
81 | |||
82 | |||
83 | * Bad rotaion center and beam tilt pivot point alignment |
||
84 | |||
85 | |||
86 | * <link linkend="focdark">Objective aperture is off alignment</link>. |
||
87 | |||
88 | |||
89 | * Bad calibration |
||
90 | |||
91 | |||
92 | |||
93 | |||
94 | |||
95 | Solutions: |
||
96 | * Out of range: |
||
97 | |||
98 | |||
99 | * Make sure that the lower mag autofocus, i.e., Z focus node, is preformed |
||
100 | successfully by observing its progress, and, if necessary, adjust the stage height |
||
101 | in its Manual focus tool. |
||
102 | |||
103 | |||
104 | * Make sure that Z Focus node autofocus accuracy is compatible with the Focus |
||
105 | node. For example, Z Focus at 5000x has an accuracy of +/- 5 um and Focus at |
||
106 | 50,000x can handle range up to about +/- 10 um. |
||
107 | |||
108 | |||
109 | * Choose a flater grid square. |
||
110 | |||
111 | |||
112 | |||
113 | |||
114 | * For thick ice |
||
115 | |||
116 | |||
117 | * Leginon/Focus/Settings/Autofocus> Increase the ice melt time in autofocus |
||
118 | setting |
||
119 | |||
120 | |||
121 | * Leginon/Exposure Targeting/Hole Targeting Setting/Target Template/Focus |
||
122 | Template Thickness> Adjust allowed ice thickness values to pick focus target in |
||
123 | area with thinner ice |
||
124 | |||
125 | |||
126 | * Leginon/Presets Manager> reduce "fa" preset CCD binning and/or increase |
||
127 | exposure time. |
||
128 | |||
129 | |||
130 | * Bad rotaion center and beam tilt pivot point alignment |
||
131 | |||
132 | |||
133 | * Bad calibration |
||
134 | |||
135 | |||
136 | |||
137 | |||
138 | * For stage drift during the procedure |
||
139 | |||
140 | |||
141 | * Leginon/Focus/Settings/Autofocus> Reduce drift check threshold. |
||
142 | |||
143 | |||
144 | |||
145 | |||
146 | * For charging during the procedure |
||
147 | |||
148 | |||
149 | * Leginon/Exposure Targeting> Pick targets away from ice chunk, grid bar, and |
||
150 | large ice gradient. |
||
151 | |||
152 | |||
153 | |||
154 | |||
155 | * For ice melting during the procedure |
||
156 | |||
157 | |||
158 | * Leginon/Focus/Settings/Autofocus> Increase ice melt time |
||
159 | |||
160 | |||
161 | |||
162 | |||
163 | * For image spikes |
||
164 | |||
165 | |||
166 | * Leginon/Correction> Adjust despike parameters |
||
167 | |||
168 | |||
169 | * Leginon/Correction> Reduce exposure time? Might reduce number of random |
||
170 | event |
||
171 | |||
172 | |||
173 | |||
174 | |||
175 | |||
176 | |||
177 | |||
178 | |||
179 | |||
180 | |||
181 | h2. Failed autofocus in Z Focus node |
||
182 | |||
183 | |||
184 | |||
185 | Commonly Why: |
||
186 | * The stage height relative to Eucentric point is outside of measurable range due to |
||
187 | very uneven grid or stage far away from U-center. |
||
188 | |||
189 | |||
190 | * Empty area or area with very thick ice, or at Gaussian focus |
||
191 | |||
192 | |||
193 | * Objective aperture is off alignment. |
||
194 | |||
195 | |||
196 | * Bad recorded eucentric focus. |
||
197 | |||
198 | |||
199 | * Bad rotaion center and beam tilt pivot point alignment |
||
200 | |||
201 | |||
202 | * Bad calibration |
||
203 | |||
204 | |||
205 | To identify which is the source, observe the image, correlation, and peak |
||
206 | during beam tilt autofocusing. Also check in the log the residual value (so called min) of |
||
207 | the fitting. |
||
208 | |||
209 | |||
210 | * Image has very good contrast but features are completely different from one tilt |
||
211 | image to the next-Out of range. |
||
212 | |||
213 | |||
214 | * No image contrast or correlation peaks move randomly-empty or thick ice, or at |
||
215 | Gaussian focus |
||
216 | |||
217 | |||
218 | * Contrast reverse in the holes relative to the support matrix in one or more beam |
||
219 | tilt images-Objective aperture is off alignment. |
||
220 | |||
221 | |||
222 | * If you know that the stage is at U-center, but the images appear to be strongly |
||
223 | defocused-Recorded eucentric focus is incorrect. |
||
224 | |||
225 | |||
226 | * Correlation peaks are well defined but fitting fails-Bad rotaion center and beam |
||
227 | tilt pivot point alignment, or bad calibration. |
||
228 | |||
229 | |||
230 | |||
231 | |||
232 | Solutions: |
||
233 | * Out of range: |
||
234 | |||
235 | |||
236 | * U-center the stage before the run and when Z focus fails. |
||
237 | |||
238 | |||
239 | * Activate Manual Check after autofocus to make adjustment in stage Z |
||
240 | height. |
||
241 | |||
242 | |||
243 | * For expert user: Add another focus step before the current autofocus step at |
||
244 | even lower magnification for a larger range usability. For example, use sq preset. |
||
245 | You will need to perform a calibration with Beam Tilt node in the Calibration |
||
246 | application. |
||
247 | |||
248 | |||
249 | |||
250 | |||
251 | * Empty area or thick ice: |
||
252 | |||
253 | |||
254 | * Pick better focus target. |
||
255 | |||
256 | |||
257 | * Leginon/Z Focus/Hole Targeting Settings/Ice Thickness Threshold/Focus Hole |
||
258 | Selection> Consider using "Good hole" rather than "Any hole". |
||
259 | |||
260 | |||
261 | |||
262 | |||
263 | * At Guassian focus: |
||
264 | |||
265 | |||
266 | * There is no problem, don't worry. |
||
267 | |||
268 | |||
269 | |||
270 | |||
271 | * Bad recorded eucentric focus |
||
272 | |||
273 | |||
274 | * Leginon> send to scope the preset used to autofocus in Z Focus node (i.e., |
||
275 | hl in MSI). |
||
276 | |||
277 | |||
278 | * TEM> Adjust the stage to U-center. |
||
279 | |||
280 | |||
281 | * TEM> Focus the image and reset defocus. |
||
282 | |||
283 | |||
284 | * Leginon/Z Focus> start Manual focus tool. |
||
285 | |||
286 | |||
287 | * Leginon/Z Focus/Manual Focus> send 0 defocus to scope. |
||
288 | |||
289 | |||
290 | * Leginon/Z Focus/Manual Focus> save "Eucentric focus from |
||
291 | Instrument". |
||
292 | |||
293 | |||
294 | |||
295 | |||
296 | * Bad rotation center and beam tilt pivot point alignment: |
||
297 | |||
298 | |||
299 | * Make sure you are at the eucentric focus during the alignment. |
||
300 | |||
301 | |||
302 | |||
303 | |||
304 | * Bad alignment on objective aperture: |
||
305 | |||
306 | |||
307 | * <link linkend="focdark">Center the objective aperture in diffraction |
||
308 | mode</link>. |
||
309 | |||
310 | |||
311 | |||
312 | |||
313 | |||
314 | |||
315 | |||
316 | |||
317 | |||
318 | |||
319 | h2. Objects reverse contrast during autofocus |
||
320 | |||
321 | |||
322 | |||
323 | In this case of autofocus failure, objects such as holes change from lighter than matrix |
||
324 | to darker as if a dark-field image is collected during the autofocus beam tilts. |
||
325 | |||
326 | |||
327 | Commonly Why: |
||
328 | * The objective aperture is off. |
||
329 | |||
330 | |||
331 | * Bad rotation center and beam tilt pivot point alignment. |
||
332 | |||
333 | |||
334 | |||
335 | |||
336 | |||
337 | Solutions: |
||
338 | * Align objective aperture at mags where autofocus is performed in HM mode. Here is |
||
339 | how: |
||
340 | |||
341 | |||
342 | * TEM>Make sure the rotation center is correct. |
||
343 | |||
344 | |||
345 | * TEM>Change the mag to that of the autofocus preset that shows the |
||
346 | problem. |
||
347 | |||
348 | |||
349 | or |
||
350 | |||
351 | |||
352 | Leginon/Presets Manager>Send the autofocus preset. |
||
353 | |||
354 | |||
355 | * Imaging an area that have scattering matter such as with ice or |
||
356 | carbon. |
||
357 | |||
358 | |||
359 | * TEM>Press the diffraction button to go into diffraction mode. |
||
360 | |||
361 | |||
362 | * TEM>Use the Mag adjustment knob to change diffraction distance to ~1 |
||
363 | m. |
||
364 | |||
365 | |||
366 | * TEM>You should see a bright and condensed image of your imaging area which |
||
367 | would condenses to a point if you are right at focus (adjust with focus knob). The |
||
368 | faint disk outside of the bright area defines the opening of the objective |
||
369 | aperture. Center the image using the MULTIFUNCTION KNOBS if you can not see the |
||
370 | edge. Move the STAGE to area with more scattering matter if the faint disk is not |
||
371 | visible. |
||
372 | |||
373 | |||
374 | * TEM>Physically move the aperture so the faint disk is centered on the |
||
375 | bright image. |
||
376 | 2 | Amber Herold | |
377 | |||
378 | |||
379 | ______ |
||
380 | |||
381 | [[Troubles with Targeting|< Troubles with Targeting]] | [[Troubles with Calibrations - Troubleshooting|Troubles with Calibrations >]] |
||