Project

General

Profile

Troubles with Imaging - Troubleshooting » History » Version 7

Eric Hou, 06/24/2010 02:11 PM

1 4 Eric Hou
h1. Troubles with Imaging
2 1 Amber Herold
3
h2. Image is acquired but contains no real information
4
5
Commonly Why:
6 3 Eric Hou
7 1 Amber Herold
* Main screen at the scope is down
8
9
* Camera is not inserted (Gatan camera)
10
11
* Digital Micrograph program is not running (Gatan camera)
12
13 3 Eric Hou
* Shutter switch box is set to CM, i.e., to collect film data only (Tietz camera)
14 1 Amber Herold
15 3 Eric Hou
h2. Image is acquired but contains artifacts either in image or in Fourier Transform
16 1 Amber Herold
17
Commonly Why:
18 3 Eric Hou
19 1 Amber Herold
* Beam is partially covering the CCD.
20
21
* Scope viewing window is not covered and room light is on
22
23
* Dark/Bright images are not available or need to be reaquire.
24
25 3 Eric Hou
* Bias level and quadrant correction may need to be reset (Digital Micrograph and Gatan camera)
26 1 Amber Herold
27 3 Eric Hou
h2. Can not read reference images that were acquired by other users
28 1 Amber Herold
29
Commonly Why: You don't have permission to read their image files
30
31
Solution:
32
33 3 Eric Hou
*  Have the person change the group or even world permission for reading the data directory and the image files in it.
34 1 Amber Herold
35
The person is not around when this happens:
36
37 3 Eric Hou
Acquire your own reference images at the camera setting that needs the dark and bright images. However, Leginon always look for the most recent calibration in the database by anyone. Every user will have to do this everytime if the permission problem is not solved.
38 1 Amber Herold
39
h2. Beam shifts away from imaging area over time
40
41
Commonly Why:
42
43 3 Eric Hou
* Residual lens hysteresis if you haven't allowed the scope to warm up for HM mode after acquiring LM atlas.
44 1 Amber Herold
45 3 Eric Hou
* Focus correction has taken the absolute focus point far away from U-centric focus.
46 1 Amber Herold
47 3 Eric Hou
h3. Solutions:
48 1 Amber Herold
49
* Wait until the optics is stabilized
50
51 3 Eric Hou
* Activate "Publish and wait for the reference target" in Exposure Node. This will enable "Fix Beam" node to search for the beam in the desired range and shift the beam to the best result. A reference target, normally an empty square, need to be selected on the grid to use this function. See description in the node description chapter for usage.
52 1 Amber Herold
* Pause MSI and correct for residual beam shift
53 5 Eric Hou
 
54
## Pause before next hole or square by clicking Hole/Toolbar>Pause button (Pause icon) or Square/Toolbar>Pause button (Pause icon), respectively. Which one to pause depends on what is the next target.
55
 
56
## Leginon/Presets Manager> When the acquisition sequence is stopped, send the preset that has beam shift problem "To Scope" (ToScope icon).
57
 
58
## Leginon/Navigation/Toolbar> Choose move type to be "Beam Shift" and make sure "Use Camera Configuration" is unchecked in settings.
59
 
60
## Leginon/Navigation> "Navigate"*(navigate icon) to center the beam. If "Navigate" does not bring the beam in, manually shift the beam at the microscope. Do not forget to lift the screen and cover the viewing area after done.
61
 
62
## Leginon/Presets Manager> save the new beam shift by retrieving it "From Scope" (From Scope icon).
63
 
64
## "Continue" data collection on next hole or square target by clicking Hole/Toolbar> "Continue" button or Square/Toolbar> "Continue" button, respectively.
65
 
66
"Navigate" means selecting the navigate tool on the top right of image display and then single-left-clicking the location of the new center that is translated by the given TEM Parameter.
67 1 Amber Herold
68
* Reset the defocus=0 point back to U-centric focus.
69 6 Eric Hou
 
70
Leginon reset defocus after each defocus correction. Therefore, the point where the user defines as defocus=0 may drift away from U-centric focus over time or after a very large correction before it is reset to U-centric focus by z focus correction.
71
 
72
##  Pause before next hole or square by clicking Hole/Toolbar>Pause button (Pause icon) or Square/Toolbar>Pause button (Pause icon), respectively. Which one to pause depends on what is the next target.
73
 
74
##  Leginon/Presets Manager> Send "hl" preset "To Scope" (ToScope icon).
75
 
76
##  Leginon/Z Focus/Toolbar> Send stored U-centric focus to scope.
77
 
78
##  Leginon/Z Focus/Toolbar> Reset defocus to 0
79
 
80
##  Scope> U-center the stage if it is very out of focus.
81
 
82
##  Leginon/Presets Manager> Send the preset that showed bad beam shift "To Scope"
83
 
84
##  Scope and Leginon/Presets Manager> Check the beam shift, make correction, and then retrieve the correct value "From Scope"
85
 
86
##  Leginon/Drift Manager> Declare Drift (declare drift icon) to force correction at next possible point in the process.
87
 
88
##  "Continue" data collection on next hole or square target by clicking Hole/Toolbar> "Continue" button or Square/Toolbar> "Continue" button, respectively.
89 1 Amber Herold
90 3 Eric Hou
h2. Beam shifts away from imaging area only when targeting move type is image shift
91 1 Amber Herold
92
Commonly Why:
93 3 Eric Hou
94 1 Amber Herold
* Image/Beam Shift calibration on the microscope is not optimal
95
96
* Defocus is very far from eucentric focus
97
98 3 Eric Hou
h3. Solutions:
99 1 Amber Herold
100
* Bad Image/Beam calibration
101 6 Eric Hou
 
102 3 Eric Hou
Follow instruction for performing the calibration for FEI Tecnai machines under Alignments/Image HM-TEM(or LM)/Image-Beam calibration
103 6 Eric Hou
 
104
##  Pause before next hole or square by clicking Hole/Toolbar>Pause button (Pause icon) or Square/Toolbar>Pause button (Pause icon), respectively. Which one to pause depends on what is the next target. If in queuing mode, pause at the node that process the queue.
105
 
106
##  Scope> move to an unimportant area or pull the holder out enough to allow the beam to go through.
107
 
108
##  Scope> reset defocus to eucentric height.
109
 
110
##  Scope> Follow instruction for performing the calibration for FEI Tecnai machines under Alignments/Image HM-TEM(or LM)/Image-Beam calibration.
111
 
112
##  Leginon/Presets Manager> Cycle the presets a few times. The calibration takes the scope to conditions outside the presets, and can cause strong hysteresis if this step is not done.
113
 
114
##  Leginon> Follow the procedures for confirming and saving good image shift at various presets.
115
 
116
##  "Continue" data collection on next hole or square target by clicking Hole/Toolbar> "Continue" button or squares/Target Toolbar> "Continue" button, respectively.
117
 
118 7 Eric Hou
"Navigate" means selecting the navigate tool on the top right of image display and then single-left-clicking the location of the new center that is translated by the given TEM Parameter.
119 1 Amber Herold
120
* Defocus far from eucentric focus
121 7 Eric Hou
 
122 3 Eric Hou
Image/Beam shift coupling worsens when the defocus is away from where it was calibrated. If, after the above calibration, the problem remains during MSI acquisition, additional focusing sequence should be added at lower magnification that move the stage to close to eucentric height so that the image shift target is selected
123 1 Amber Herold
on an image of a close-to-eucentric location.
124 7 Eric Hou
 
125
##  Leginon/Z Focus/Focus Sequence> activate both Stage_Wobble and Z_to_Encentric steps.
126
 
127
##  Leginon/Hole (or Subsquare) Targeting> selection a focus target if not already done so when a newly acquired sq image comes in.
128
 
129 3 Eric Hou
If the accuracy of moving stage to eucentric height by the focus sequences in "Z Focus" is still not sufficient. Add another focus step performing the same task as Z_to_Encentric step. Repeating beam-tilt based autofocusing often improve the accuracy unless the calibration is off.
130 1 Amber Herold
131 3 Eric Hou
Still not working: You have chosen a target that requires too much image shift for an independent image shift from beam shift. This is necessary when the lower mag targetting is not good either because the preset image shifts are not aligned or the stage position movement is not properly modeled.
132 1 Amber Herold
133 3 Eric Hou
Solution: For image shift problem see solution for "Target is consistently off in the same direction and distance". For stage model problem, see "Modeled stage calibration" in setup notes.
134 1 Amber Herold
135 3 Eric Hou
h2. Beam shifts away from imaging area when targeting move type is stage movement
136 1 Amber Herold
137
Commonly Why:
138 3 Eric Hou
139 1 Amber Herold
* U-center defocus not set to 0 in LM mode.
140
141
* LM alignment such as rotation center is way off.
142
143 3 Eric Hou
Solution: Check and correct microscopic alignment error. Preset parameters checking and reset may be necessary after the correction.
144 1 Amber Herold
145 3 Eric Hou
Comments: If the problem persist, the original stage matrix calibration may have been performed with a bad alignment. The calibration should be redone with a well-aligned scope in LM mode.
146 1 Amber Herold
147 3 Eric Hou
h2. Objective aperture appears off in "sq" preset when it is centered in HM mode
148 1 Amber Herold
149
Commonly Why:
150 3 Eric Hou
151 1 Amber Herold
* U-center defocus not set to 0 in LM mode.
152
153
* LM alignment such as image shift is way off.
154
155 3 Eric Hou
Solution: Check and correct microscopic alignment error. Preset parameters checking and reset may be necessary after the correction.
156 1 Amber Herold
157 3 Eric Hou
h2. The specimen appears over-dosed in the high magnification images or the dark correction image has a high value (while using a Gatan CCD)
158 1 Amber Herold
159 3 Eric Hou
This issue may occur when the shutter is incorrectly configured through Digital Micrograph (the imaging software for the Gatan CCD that resides on the TEM computer) or simply a bad dose calibration or too long an exposure time. To issue these corrections:
160 1 Amber Herold
161 3 Eric Hou
*  Set the "shutter" to "Auto" on the external control box attached to the TEM.
162 1 Amber Herold
163 3 Eric Hou
*  Open Digital Micrograph on the Microscope computer and go to the Camera Set-up page.
164 1 Amber Herold
165
*  Go to Configuration
166
167
*  Set the Primary Shutter to Pre-specimen
168
169
*  Set the Alternate Shutter to Post-specimen
170
171
*  Set the Default Primary Shutter to the Normally Closed position.
172
173 3 Eric Hou
h2. The exposure image has a circular imprint of the beam at the size of the preset used for melting the ice
174 2 Amber Herold
175 3 Eric Hou
This issue may occur when the main screen is left up during very long (such as over 30 sec) ice melting and at high HT. Leginon has a mechanism that put the screen down during ice melting. It is caused by over-saturating the Phosphur layer on the CCD. If you see this effect, the screen may have failed to lower. Please test it by observing the screen movement with simulating target in any focuser node that has ice melt time set not to zero. Report the problem back to the leginon team if you are sure it is not your scope's problem.
176 2 Amber Herold
177
______
178
179
[[Hardware Troubles|< Hardware Troubles]] | [[Troubles with Targeting|Troubles with Targeting >]]
180
181
______