Troubles with Targeting » History » Version 1
Amber Herold, 04/23/2010 09:13 AM
1 | 1 | Amber Herold | h1. Troubles with Targeting |
---|---|---|---|
2 | |||
3 | |||
4 | |||
5 | |||
6 | |||
7 | * <link linkend="atlassmall">Grid Atlas is too small to see the details</link> |
||
8 | |||
9 | |||
10 | * <link linkend="atlasbad">Grid Atlas is not well patched</link> |
||
11 | |||
12 | |||
13 | * <link linkend="atlassmall">Grid Atlas displayed in Leginon and the Viewers on the web |
||
14 | are not the samei</link> |
||
15 | |||
16 | |||
17 | * <link linkend="atlassamelabel">Grid Atlas displays incorrect current |
||
18 | position</link> |
||
19 | |||
20 | |||
21 | * <link linkend="toffinsamedir">Targeting is consistently off in the same direction and |
||
22 | distance</link> |
||
23 | |||
24 | |||
25 | * <link linkend="toffindiffdir">Targeting is consistently off in opposite direction from |
||
26 | where you pick</link> |
||
27 | |||
28 | |||
29 | * <link linkend="toffinconsist">Targeting is off in an inconsistant way or goes off again |
||
30 | in the next run after correction</link> |
||
31 | |||
32 | |||
33 | * <link linkend="toffawayfromc">Targeting is off more for targets away from the center of |
||
34 | the image</link> |
||
35 | |||
36 | |||
37 | * <link linkend="hlfindfail">Auto hole finder fails to find holes with good ice after set |
||
38 | up</link> |
||
39 | |||
40 | |||
41 | * <link linkend="no_broken">There is no broken square for reference target</link> |
||
42 | |||
43 | |||
44 | |||
45 | |||
46 | |||
47 | |||
48 | |||
49 | h2. Grid mosaic is too small to see the details |
||
50 | |||
51 | |||
52 | |||
53 | |||
54 | |||
55 | <title>Why:</title> |
||
56 | * The image scale for creating mosaic is set too small. |
||
57 | |||
58 | |||
59 | |||
60 | |||
61 | |||
62 | |||
63 | |||
64 | |||
65 | <title>Solution:</title> |
||
66 | * Leginon/Square Targeting/Mosaic Settings> Change "Scale Image" to a larger |
||
67 | size |
||
68 | |||
69 | |||
70 | * Leginon/Square Targeting/Mosaic Settings> Click "Apply" and then "Create" the |
||
71 | mosaic atlas. |
||
72 | |||
73 | |||
74 | |||
75 | |||
76 | |||
77 | |||
78 | |||
79 | |||
80 | |||
81 | h2. Grid mosaic is not patched together well |
||
82 | |||
83 | |||
84 | |||
85 | |||
86 | |||
87 | <title>Commonly Why:</title> |
||
88 | * LM mode U-center defocus not set to 0. |
||
89 | |||
90 | |||
91 | * LM mode rotation center is off. |
||
92 | |||
93 | |||
94 | * Bad stage position matrix calibration |
||
95 | |||
96 | |||
97 | |||
98 | |||
99 | |||
100 | |||
101 | |||
102 | |||
103 | <title>Solutions:</title> |
||
104 | * Correct the alignemnt error or redo <link linkend="mcal_st_pos">stage position |
||
105 | matrix calibration</link> |
||
106 | |||
107 | |||
108 | * Acquire a new atlas |
||
109 | * Leginon/Grid Targeting/Setting> Enter a new label |
||
110 | |||
111 | |||
112 | * Leginon/Grid Targeting> Click "Calculate Atlas" |
||
113 | |||
114 | |||
115 | * Leginon/Grid Targeting> Click "Submit Targets" (play icon) to begin |
||
116 | acquiring the atlas. |
||
117 | |||
118 | |||
119 | * Leginon/Square Targeting> Follow the atlas progress in the |
||
120 | display. |
||
121 | |||
122 | |||
123 | |||
124 | |||
125 | |||
126 | |||
127 | |||
128 | |||
129 | |||
130 | |||
131 | |||
132 | |||
133 | h2. Grid mosaic display in Leginon is different from that of the |
||
134 | Web Viewer |
||
135 | |||
136 | |||
137 | |||
138 | |||
139 | |||
140 | <title>Why:</title> |
||
141 | * The goniometer does not reach the intended position due to ,most commonly, a pole |
||
142 | touch error that temporaily disables the goniometer. |
||
143 | |||
144 | |||
145 | * The calibration used to move to the target in "Grid" node is not the same as the |
||
146 | calibration parameter used to create the mosaic in "Square Targeting" node are not |
||
147 | consistent. |
||
148 | |||
149 | |||
150 | |||
151 | |||
152 | |||
153 | |||
154 | |||
155 | |||
156 | <title>Solution:</title> |
||
157 | * Other than fix the goniometer, Targeting on these mosaic can still work if the |
||
158 | error does not occur frequently enough to interfere drift management. |
||
159 | |||
160 | |||
161 | * Check "Leginon/Grid/Settings/Use ?? to move to target" against "Leginon/Square |
||
162 | Targeting/Mosaic Settings/CalibrationParameters". The latter should be changed if the |
||
163 | mosaic is already acquire and what seen on the Web is better than what is in Leginon. |
||
164 | The mosaic atlas can be recreated by clicking the "Create" button. |
||
165 | |||
166 | |||
167 | |||
168 | |||
169 | |||
170 | |||
171 | |||
172 | |||
173 | <title>More Explanation:</title> |
||
174 | * The web tools and Leginon main program use different mechanism to paste together |
||
175 | the atlas. |
||
176 | |||
177 | |||
178 | The targets Leginon "Grid Targeting" node send out is an even spaced raster. |
||
179 | "Grid" node uses that information and the move type set for the node to calculate the |
||
180 | goniometer position it needs to aim for in order to get to where it really wants to go |
||
181 | to and acquires the image. It is done this way so that it can handle any complex move |
||
182 | type. For example, lets say that the model is p=x^2 so that to get to position p' from |
||
183 | 0, the goniometer x needs to be sqrt(p'). "Grid Targeting" send the target as p', |
||
184 | "Grid" figure out that x'=sqrt(p') and send the x value to the scope to move. |
||
185 | |||
186 | |||
187 | When it needs to display the atlas in "Square targeting", Leginon uses the actual |
||
188 | goniometer position of each tile image back calculates the targeted position according |
||
189 | to the model. This has the advantage that if there is a problem with the goniometer |
||
190 | and the stage does not move to where it needs to (such as after a pole touch error), |
||
191 | the tile image will appear at its actual position. In the example above, the image |
||
192 | will appear at p' from the image because from the goniometer position x'=sqrt(p'), |
||
193 | p=x^2=(sqrt(p'))^2. |
||
194 | |||
195 | |||
196 | The webtool, however, pastes the atlas together using the targets saved from "Grid |
||
197 | Targeting". It does not care what you have set in "Square Targeting" node. Therefore, |
||
198 | it will always look right, unless of course the stage did not move to the target when |
||
199 | it acquired the image. |
||
200 | |||
201 | |||
202 | |||
203 | |||
204 | |||
205 | |||
206 | |||
207 | |||
208 | |||
209 | h2. Commonly Why: |
||
210 | |||
211 | |||
212 | * You did not initiate refresh by clicking the "current position" tool |
||
213 | |||
214 | |||
215 | * More than one atlas is of the same label. |
||
216 | |||
217 | |||
218 | |||
219 | |||
220 | |||
221 | |||
222 | |||
223 | |||
224 | <title>Solutions:</title> |
||
225 | * No user correction can be made with atlas of the same label. It requires |
||
226 | modification of the database. |
||
227 | |||
228 | |||
229 | * Acquire a new atlas with a different label |
||
230 | |||
231 | |||
232 | |||
233 | |||
234 | |||
235 | |||
236 | |||
237 | |||
238 | |||
239 | h2. Targeting is consistently off in the same direction and |
||
240 | distance |
||
241 | |||
242 | |||
243 | |||
244 | Commonly Why: Image shifts in "hl", "sq", or "gr"presets are not aligned with the |
||
245 | presets at higher mags. The low mag "sq" preset is especially prone to this problem. |
||
246 | |||
247 | |||
248 | ***Important: When the targeted image is off, it normally means the image shift of the |
||
249 | preset where the target comes from is off. That is, if hole images appear off, it is not the |
||
250 | image shift of "hl" preset that needs correcting but that of "sq" preset. |
||
251 | |||
252 | |||
253 | Solution: |
||
254 | |||
255 | |||
256 | |||
257 | |||
258 | * Pause before next hole or square by clicking Hole/Toolbar>Pause button or |
||
259 | Square/Toolbar>Pause button, respectively. Which one to pause depends on what is |
||
260 | the next target. |
||
261 | |||
262 | |||
263 | * Center the grid at a location that can be recognized by either: |
||
264 | |||
265 | |||
266 | |||
267 | |||
268 | * Move to the alignment position: |
||
269 | |||
270 | |||
271 | (It may be necessary to adjust the position at the scope after this since |
||
272 | the goniometer can not move back to a position as accurately as we need |
||
273 | it). |
||
274 | |||
275 | |||
276 | |||
277 | |||
278 | * Leginon/Navigation/Stage Location> Choose the name for the stored |
||
279 | alignment position ("align" if you have been following the setup check |
||
280 | list). and send it "To Scope" |
||
281 | |||
282 | |||
283 | |||
284 | |||
285 | |||
286 | |||
287 | * scope> move the specimen to a position on the same square, either on a |
||
288 | recognizable object or burn a good hole through ice as preparation for lower mag |
||
289 | preset image shift correction |
||
290 | |||
291 | |||
292 | |||
293 | |||
294 | |||
295 | |||
296 | * Leginon/Presets Manager> send "fa" or other high mag preset with (0.0,0.0) |
||
297 | image shift "To Scope" |
||
298 | |||
299 | |||
300 | * Leginon/Navigation> "Acquire" an image. The recognizable object or the burn |
||
301 | hole should be at the center of the image, if not move the grid at the |
||
302 | microscope. |
||
303 | |||
304 | |||
305 | * Check, adjust, and save the lower mag preset image shift in hole and sq |
||
306 | |||
307 | |||
308 | |||
309 | |||
310 | * Leginon/Presets Manager> send the preset "To Scope". |
||
311 | |||
312 | |||
313 | * Leginon/Navigation> "Acquire" an image in the Camera |
||
314 | Configuration. |
||
315 | |||
316 | |||
317 | * Leginon/Navigation> "Navigate" by "Image Shift" to center on the |
||
318 | reference object or the burn mark made in the high mag presets. |
||
319 | |||
320 | |||
321 | * Leginon/Presets Manager> save the new image shift by retrieving them |
||
322 | "From Scope" |
||
323 | |||
324 | |||
325 | *"Navigate" means selecting the navigate tool on the top right of image |
||
326 | display and then left-clicking the location of the new center that is translated |
||
327 | by the given TEM Parameter. |
||
328 | |||
329 | |||
330 | |||
331 | |||
332 | |||
333 | |||
334 | * Leginon/Drift Manager/Toolbar>"Declare Drift"(declare drift icon) to force |
||
335 | target shift check at the next possible time point. This will correct further |
||
336 | targeting. |
||
337 | |||
338 | |||
339 | * "Continue" data collection on next hole or square target by clicking |
||
340 | Hole/Toolbar> "Continue" button or Square/> "Continue" button, |
||
341 | respectively. |
||
342 | |||
343 | |||
344 | |||
345 | |||
346 | |||
347 | |||
348 | Still not working: Your high mag "fc", "en", or "ef" has an image shift other than that |
||
349 | of "fa". The latter is commonly used in image shift alignment of presets. |
||
350 | |||
351 | |||
352 | Solution: |
||
353 | |||
354 | |||
355 | |||
356 | |||
357 | * Leginon/Presets Manager> select individual suspect to check its image shift x |
||
358 | and y, change them to the correct values but not send the preset to Scope. |
||
359 | |||
360 | |||
361 | |||
362 | |||
363 | |||
364 | |||
365 | |||
366 | |||
367 | |||
368 | h2. Targeting is consistently off in opposite direction from |
||
369 | where you pick |
||
370 | |||
371 | |||
372 | |||
373 | Commonly Why: Camera configuration is different in calibration and now. |
||
374 | |||
375 | |||
376 | |||
377 | |||
378 | * All your targeting would be wrong. "Abort" target processing in the Target |
||
379 | Watcher of "squares" and "holes". |
||
380 | |||
381 | |||
382 | * If you use the Tietz camera: scope/Leginon/Instrument/Camera> check your |
||
383 | camera configuration. mirror should be ['vertical']. Change it if not, and "Set" the |
||
384 | configuration. |
||
385 | |||
386 | |||
387 | * If you use the Gatan camera: scope/Digital Micrograph> check your camera |
||
388 | configuration. |
||
389 | |||
390 | |||
391 | * Leginon/Presets Manager> send "sq" or "grid" "To Scope" |
||
392 | |||
393 | |||
394 | * Leginon/Navigation> "Acquire" an image with "Use this configuration" Unchecked |
||
395 | in the Camera Configuration. |
||
396 | |||
397 | |||
398 | * Leginon/Navigation> "Navigate" by "Image Shift", "Beam Shift", or "Stage |
||
399 | Position" to center on a reference object or burn mark made in the high mag presets. |
||
400 | This is to check if it behaves correctly now. |
||
401 | |||
402 | |||
403 | * Select targets again, or, if atlas is wrong, republish the target list to make |
||
404 | the atlas. |
||
405 | |||
406 | |||
407 | |||
408 | |||
409 | |||
410 | |||
411 | |||
412 | |||
413 | |||
414 | h2. Hole targeting is inconsistently off or goes bad in the |
||
415 | next run after correction |
||
416 | |||
417 | |||
418 | |||
419 | Commonly Why: Not well understood but may be related to microscopy alignment and |
||
420 | settings such as normalization of lenses. up to 1.5 um error has been observed even with a |
||
421 | reasonable calibration. The following solutions only works some times. |
||
422 | |||
423 | |||
424 | |||
425 | |||
426 | * Leginon/Drift Manager>"Declare Drift" to force correction for possible |
||
427 | drift. |
||
428 | |||
429 | |||
430 | * Repeat image shift correction alignment <link linkend="toffinsamedir">as in the |
||
431 | case when targeting is consistently off</link>. |
||
432 | |||
433 | |||
434 | * Ignore the first targetted '"hole" in the "squares" image. The second and on are |
||
435 | more consistent when there is a hysteresis problem. |
||
436 | |||
437 | |||
438 | * Repeat <link linkend="gonm_m_only">"Mag Only" Modeled Stage Calibration</link> at |
||
439 | the "square" preset. |
||
440 | |||
441 | |||
442 | |||
443 | |||
444 | |||
445 | |||
446 | |||
447 | |||
448 | |||
449 | h2. Hole targeting is off more when the image shift target is |
||
450 | away from the center of the image |
||
451 | |||
452 | |||
453 | |||
454 | Commonly Why: Not well understood but may be related to microscopy image shift causes |
||
455 | different physical movement at different magnifications since Leginon uses the image shift |
||
456 | calibration at the lower mag where the target comes from to determine the required image |
||
457 | shift at the higher mag. |
||
458 | |||
459 | |||
460 | |||
461 | |||
462 | No current solution |
||
463 | |||
464 | |||
465 | |||
466 | |||
467 | |||
468 | h2. Automatic hole finding fails to pick up good ice thickness |
||
469 | after set up |
||
470 | |||
471 | |||
472 | |||
473 | Commonly Why: |
||
474 | * Electron beam intensity has changed |
||
475 | |||
476 | |||
477 | * Blob threshold is not inclusive enough to tolerate variations |
||
478 | |||
479 | |||
480 | * Ice contaimination |
||
481 | |||
482 | |||
483 | |||
484 | |||
485 | |||
486 | Solutions |
||
487 | * Monitor electron beam with a saved location with an empty hole in navigator and |
||
488 | check the recorded intensity periodically and adjust zero ice thickness |
||
489 | accordingly. |
||
490 | |||
491 | |||
492 | * Reduce Blob threshold in Alias nodes of the Hole Finder. |
||
493 | |||
494 | |||
495 | * Make sure the anticontamination cryo box is properly inserted on the scope. |
||
496 | |||
497 | |||
498 | * Remove the film camera from the scope if you don't need film. |
||
499 | |||
500 | |||
501 | |||
502 | |||
503 | |||
504 | |||
505 | |||
506 | |||
507 | h2. No broken grid square for dose measurement and ZLP alignment |
||
508 | (MSI-Tomography) |
||
509 | |||
510 | |||
511 | |||
512 | Solution |
||
513 | * Burn a hole clear of ice and pick it as the reference target in "Hole Targeting" |
||
514 | or "Tomography Targeting" |