Troubles with Targeting » History » Version 2
Amber Herold, 04/27/2010 02:16 PM
| 1 | 1 | Amber Herold | h1. Troubles with Targeting |
|---|---|---|---|
| 2 | |||
| 3 | |||
| 4 | |||
| 5 | |||
| 6 | |||
| 7 | * <link linkend="atlassmall">Grid Atlas is too small to see the details</link> |
||
| 8 | |||
| 9 | |||
| 10 | * <link linkend="atlasbad">Grid Atlas is not well patched</link> |
||
| 11 | |||
| 12 | |||
| 13 | * <link linkend="atlassmall">Grid Atlas displayed in Leginon and the Viewers on the web |
||
| 14 | are not the samei</link> |
||
| 15 | |||
| 16 | |||
| 17 | * <link linkend="atlassamelabel">Grid Atlas displays incorrect current |
||
| 18 | position</link> |
||
| 19 | |||
| 20 | |||
| 21 | * <link linkend="toffinsamedir">Targeting is consistently off in the same direction and |
||
| 22 | distance</link> |
||
| 23 | |||
| 24 | |||
| 25 | * <link linkend="toffindiffdir">Targeting is consistently off in opposite direction from |
||
| 26 | where you pick</link> |
||
| 27 | |||
| 28 | |||
| 29 | * <link linkend="toffinconsist">Targeting is off in an inconsistant way or goes off again |
||
| 30 | in the next run after correction</link> |
||
| 31 | |||
| 32 | |||
| 33 | * <link linkend="toffawayfromc">Targeting is off more for targets away from the center of |
||
| 34 | the image</link> |
||
| 35 | |||
| 36 | |||
| 37 | * <link linkend="hlfindfail">Auto hole finder fails to find holes with good ice after set |
||
| 38 | up</link> |
||
| 39 | |||
| 40 | |||
| 41 | * <link linkend="no_broken">There is no broken square for reference target</link> |
||
| 42 | |||
| 43 | |||
| 44 | |||
| 45 | |||
| 46 | |||
| 47 | |||
| 48 | |||
| 49 | h2. Grid mosaic is too small to see the details |
||
| 50 | |||
| 51 | |||
| 52 | |||
| 53 | |||
| 54 | |||
| 55 | <title>Why:</title> |
||
| 56 | * The image scale for creating mosaic is set too small. |
||
| 57 | |||
| 58 | |||
| 59 | |||
| 60 | |||
| 61 | |||
| 62 | |||
| 63 | |||
| 64 | |||
| 65 | <title>Solution:</title> |
||
| 66 | * Leginon/Square Targeting/Mosaic Settings> Change "Scale Image" to a larger |
||
| 67 | size |
||
| 68 | |||
| 69 | |||
| 70 | * Leginon/Square Targeting/Mosaic Settings> Click "Apply" and then "Create" the |
||
| 71 | mosaic atlas. |
||
| 72 | |||
| 73 | |||
| 74 | |||
| 75 | |||
| 76 | |||
| 77 | |||
| 78 | |||
| 79 | |||
| 80 | |||
| 81 | h2. Grid mosaic is not patched together well |
||
| 82 | |||
| 83 | |||
| 84 | |||
| 85 | |||
| 86 | |||
| 87 | <title>Commonly Why:</title> |
||
| 88 | * LM mode U-center defocus not set to 0. |
||
| 89 | |||
| 90 | |||
| 91 | * LM mode rotation center is off. |
||
| 92 | |||
| 93 | |||
| 94 | * Bad stage position matrix calibration |
||
| 95 | |||
| 96 | |||
| 97 | |||
| 98 | |||
| 99 | |||
| 100 | |||
| 101 | |||
| 102 | |||
| 103 | <title>Solutions:</title> |
||
| 104 | * Correct the alignemnt error or redo <link linkend="mcal_st_pos">stage position |
||
| 105 | matrix calibration</link> |
||
| 106 | |||
| 107 | |||
| 108 | * Acquire a new atlas |
||
| 109 | * Leginon/Grid Targeting/Setting> Enter a new label |
||
| 110 | |||
| 111 | |||
| 112 | * Leginon/Grid Targeting> Click "Calculate Atlas" |
||
| 113 | |||
| 114 | |||
| 115 | * Leginon/Grid Targeting> Click "Submit Targets" (play icon) to begin |
||
| 116 | acquiring the atlas. |
||
| 117 | |||
| 118 | |||
| 119 | * Leginon/Square Targeting> Follow the atlas progress in the |
||
| 120 | display. |
||
| 121 | |||
| 122 | |||
| 123 | |||
| 124 | |||
| 125 | |||
| 126 | |||
| 127 | |||
| 128 | |||
| 129 | |||
| 130 | |||
| 131 | |||
| 132 | |||
| 133 | h2. Grid mosaic display in Leginon is different from that of the |
||
| 134 | Web Viewer |
||
| 135 | |||
| 136 | |||
| 137 | |||
| 138 | |||
| 139 | |||
| 140 | <title>Why:</title> |
||
| 141 | * The goniometer does not reach the intended position due to ,most commonly, a pole |
||
| 142 | touch error that temporaily disables the goniometer. |
||
| 143 | |||
| 144 | |||
| 145 | * The calibration used to move to the target in "Grid" node is not the same as the |
||
| 146 | calibration parameter used to create the mosaic in "Square Targeting" node are not |
||
| 147 | consistent. |
||
| 148 | |||
| 149 | |||
| 150 | |||
| 151 | |||
| 152 | |||
| 153 | |||
| 154 | |||
| 155 | |||
| 156 | <title>Solution:</title> |
||
| 157 | * Other than fix the goniometer, Targeting on these mosaic can still work if the |
||
| 158 | error does not occur frequently enough to interfere drift management. |
||
| 159 | |||
| 160 | |||
| 161 | * Check "Leginon/Grid/Settings/Use ?? to move to target" against "Leginon/Square |
||
| 162 | Targeting/Mosaic Settings/CalibrationParameters". The latter should be changed if the |
||
| 163 | mosaic is already acquire and what seen on the Web is better than what is in Leginon. |
||
| 164 | The mosaic atlas can be recreated by clicking the "Create" button. |
||
| 165 | |||
| 166 | |||
| 167 | |||
| 168 | |||
| 169 | |||
| 170 | |||
| 171 | |||
| 172 | |||
| 173 | <title>More Explanation:</title> |
||
| 174 | * The web tools and Leginon main program use different mechanism to paste together |
||
| 175 | the atlas. |
||
| 176 | |||
| 177 | |||
| 178 | The targets Leginon "Grid Targeting" node send out is an even spaced raster. |
||
| 179 | "Grid" node uses that information and the move type set for the node to calculate the |
||
| 180 | goniometer position it needs to aim for in order to get to where it really wants to go |
||
| 181 | to and acquires the image. It is done this way so that it can handle any complex move |
||
| 182 | type. For example, lets say that the model is p=x^2 so that to get to position p' from |
||
| 183 | 0, the goniometer x needs to be sqrt(p'). "Grid Targeting" send the target as p', |
||
| 184 | "Grid" figure out that x'=sqrt(p') and send the x value to the scope to move. |
||
| 185 | |||
| 186 | |||
| 187 | When it needs to display the atlas in "Square targeting", Leginon uses the actual |
||
| 188 | goniometer position of each tile image back calculates the targeted position according |
||
| 189 | to the model. This has the advantage that if there is a problem with the goniometer |
||
| 190 | and the stage does not move to where it needs to (such as after a pole touch error), |
||
| 191 | the tile image will appear at its actual position. In the example above, the image |
||
| 192 | will appear at p' from the image because from the goniometer position x'=sqrt(p'), |
||
| 193 | p=x^2=(sqrt(p'))^2. |
||
| 194 | |||
| 195 | |||
| 196 | The webtool, however, pastes the atlas together using the targets saved from "Grid |
||
| 197 | Targeting". It does not care what you have set in "Square Targeting" node. Therefore, |
||
| 198 | it will always look right, unless of course the stage did not move to the target when |
||
| 199 | it acquired the image. |
||
| 200 | |||
| 201 | |||
| 202 | |||
| 203 | |||
| 204 | |||
| 205 | |||
| 206 | |||
| 207 | |||
| 208 | |||
| 209 | h2. Commonly Why: |
||
| 210 | |||
| 211 | |||
| 212 | * You did not initiate refresh by clicking the "current position" tool |
||
| 213 | |||
| 214 | |||
| 215 | * More than one atlas is of the same label. |
||
| 216 | |||
| 217 | |||
| 218 | |||
| 219 | |||
| 220 | |||
| 221 | |||
| 222 | |||
| 223 | |||
| 224 | <title>Solutions:</title> |
||
| 225 | * No user correction can be made with atlas of the same label. It requires |
||
| 226 | modification of the database. |
||
| 227 | |||
| 228 | |||
| 229 | * Acquire a new atlas with a different label |
||
| 230 | |||
| 231 | |||
| 232 | |||
| 233 | |||
| 234 | |||
| 235 | |||
| 236 | |||
| 237 | |||
| 238 | |||
| 239 | h2. Targeting is consistently off in the same direction and |
||
| 240 | distance |
||
| 241 | |||
| 242 | |||
| 243 | |||
| 244 | Commonly Why: Image shifts in "hl", "sq", or "gr"presets are not aligned with the |
||
| 245 | presets at higher mags. The low mag "sq" preset is especially prone to this problem. |
||
| 246 | |||
| 247 | |||
| 248 | ***Important: When the targeted image is off, it normally means the image shift of the |
||
| 249 | preset where the target comes from is off. That is, if hole images appear off, it is not the |
||
| 250 | image shift of "hl" preset that needs correcting but that of "sq" preset. |
||
| 251 | |||
| 252 | |||
| 253 | Solution: |
||
| 254 | |||
| 255 | |||
| 256 | |||
| 257 | |||
| 258 | * Pause before next hole or square by clicking Hole/Toolbar>Pause button or |
||
| 259 | Square/Toolbar>Pause button, respectively. Which one to pause depends on what is |
||
| 260 | the next target. |
||
| 261 | |||
| 262 | |||
| 263 | * Center the grid at a location that can be recognized by either: |
||
| 264 | |||
| 265 | |||
| 266 | |||
| 267 | |||
| 268 | * Move to the alignment position: |
||
| 269 | |||
| 270 | |||
| 271 | (It may be necessary to adjust the position at the scope after this since |
||
| 272 | the goniometer can not move back to a position as accurately as we need |
||
| 273 | it). |
||
| 274 | |||
| 275 | |||
| 276 | |||
| 277 | |||
| 278 | * Leginon/Navigation/Stage Location> Choose the name for the stored |
||
| 279 | alignment position ("align" if you have been following the setup check |
||
| 280 | list). and send it "To Scope" |
||
| 281 | |||
| 282 | |||
| 283 | |||
| 284 | |||
| 285 | |||
| 286 | |||
| 287 | * scope> move the specimen to a position on the same square, either on a |
||
| 288 | recognizable object or burn a good hole through ice as preparation for lower mag |
||
| 289 | preset image shift correction |
||
| 290 | |||
| 291 | |||
| 292 | |||
| 293 | |||
| 294 | |||
| 295 | |||
| 296 | * Leginon/Presets Manager> send "fa" or other high mag preset with (0.0,0.0) |
||
| 297 | image shift "To Scope" |
||
| 298 | |||
| 299 | |||
| 300 | * Leginon/Navigation> "Acquire" an image. The recognizable object or the burn |
||
| 301 | hole should be at the center of the image, if not move the grid at the |
||
| 302 | microscope. |
||
| 303 | |||
| 304 | |||
| 305 | * Check, adjust, and save the lower mag preset image shift in hole and sq |
||
| 306 | |||
| 307 | |||
| 308 | |||
| 309 | |||
| 310 | * Leginon/Presets Manager> send the preset "To Scope". |
||
| 311 | |||
| 312 | |||
| 313 | * Leginon/Navigation> "Acquire" an image in the Camera |
||
| 314 | Configuration. |
||
| 315 | |||
| 316 | |||
| 317 | * Leginon/Navigation> "Navigate" by "Image Shift" to center on the |
||
| 318 | reference object or the burn mark made in the high mag presets. |
||
| 319 | |||
| 320 | |||
| 321 | * Leginon/Presets Manager> save the new image shift by retrieving them |
||
| 322 | "From Scope" |
||
| 323 | |||
| 324 | |||
| 325 | *"Navigate" means selecting the navigate tool on the top right of image |
||
| 326 | display and then left-clicking the location of the new center that is translated |
||
| 327 | by the given TEM Parameter. |
||
| 328 | |||
| 329 | |||
| 330 | |||
| 331 | |||
| 332 | |||
| 333 | |||
| 334 | * Leginon/Drift Manager/Toolbar>"Declare Drift"(declare drift icon) to force |
||
| 335 | target shift check at the next possible time point. This will correct further |
||
| 336 | targeting. |
||
| 337 | |||
| 338 | |||
| 339 | * "Continue" data collection on next hole or square target by clicking |
||
| 340 | Hole/Toolbar> "Continue" button or Square/> "Continue" button, |
||
| 341 | respectively. |
||
| 342 | |||
| 343 | |||
| 344 | |||
| 345 | |||
| 346 | |||
| 347 | |||
| 348 | Still not working: Your high mag "fc", "en", or "ef" has an image shift other than that |
||
| 349 | of "fa". The latter is commonly used in image shift alignment of presets. |
||
| 350 | |||
| 351 | |||
| 352 | Solution: |
||
| 353 | |||
| 354 | |||
| 355 | |||
| 356 | |||
| 357 | * Leginon/Presets Manager> select individual suspect to check its image shift x |
||
| 358 | and y, change them to the correct values but not send the preset to Scope. |
||
| 359 | |||
| 360 | |||
| 361 | |||
| 362 | |||
| 363 | |||
| 364 | |||
| 365 | |||
| 366 | |||
| 367 | |||
| 368 | h2. Targeting is consistently off in opposite direction from |
||
| 369 | where you pick |
||
| 370 | |||
| 371 | |||
| 372 | |||
| 373 | Commonly Why: Camera configuration is different in calibration and now. |
||
| 374 | |||
| 375 | |||
| 376 | |||
| 377 | |||
| 378 | * All your targeting would be wrong. "Abort" target processing in the Target |
||
| 379 | Watcher of "squares" and "holes". |
||
| 380 | |||
| 381 | |||
| 382 | * If you use the Tietz camera: scope/Leginon/Instrument/Camera> check your |
||
| 383 | camera configuration. mirror should be ['vertical']. Change it if not, and "Set" the |
||
| 384 | configuration. |
||
| 385 | |||
| 386 | |||
| 387 | * If you use the Gatan camera: scope/Digital Micrograph> check your camera |
||
| 388 | configuration. |
||
| 389 | |||
| 390 | |||
| 391 | * Leginon/Presets Manager> send "sq" or "grid" "To Scope" |
||
| 392 | |||
| 393 | |||
| 394 | * Leginon/Navigation> "Acquire" an image with "Use this configuration" Unchecked |
||
| 395 | in the Camera Configuration. |
||
| 396 | |||
| 397 | |||
| 398 | * Leginon/Navigation> "Navigate" by "Image Shift", "Beam Shift", or "Stage |
||
| 399 | Position" to center on a reference object or burn mark made in the high mag presets. |
||
| 400 | This is to check if it behaves correctly now. |
||
| 401 | |||
| 402 | |||
| 403 | * Select targets again, or, if atlas is wrong, republish the target list to make |
||
| 404 | the atlas. |
||
| 405 | |||
| 406 | |||
| 407 | |||
| 408 | |||
| 409 | |||
| 410 | |||
| 411 | |||
| 412 | |||
| 413 | |||
| 414 | h2. Hole targeting is inconsistently off or goes bad in the |
||
| 415 | next run after correction |
||
| 416 | |||
| 417 | |||
| 418 | |||
| 419 | Commonly Why: Not well understood but may be related to microscopy alignment and |
||
| 420 | settings such as normalization of lenses. up to 1.5 um error has been observed even with a |
||
| 421 | reasonable calibration. The following solutions only works some times. |
||
| 422 | |||
| 423 | |||
| 424 | |||
| 425 | |||
| 426 | * Leginon/Drift Manager>"Declare Drift" to force correction for possible |
||
| 427 | drift. |
||
| 428 | |||
| 429 | |||
| 430 | * Repeat image shift correction alignment <link linkend="toffinsamedir">as in the |
||
| 431 | case when targeting is consistently off</link>. |
||
| 432 | |||
| 433 | |||
| 434 | * Ignore the first targetted '"hole" in the "squares" image. The second and on are |
||
| 435 | more consistent when there is a hysteresis problem. |
||
| 436 | |||
| 437 | |||
| 438 | * Repeat <link linkend="gonm_m_only">"Mag Only" Modeled Stage Calibration</link> at |
||
| 439 | the "square" preset. |
||
| 440 | |||
| 441 | |||
| 442 | |||
| 443 | |||
| 444 | |||
| 445 | |||
| 446 | |||
| 447 | |||
| 448 | |||
| 449 | h2. Hole targeting is off more when the image shift target is |
||
| 450 | away from the center of the image |
||
| 451 | |||
| 452 | |||
| 453 | |||
| 454 | Commonly Why: Not well understood but may be related to microscopy image shift causes |
||
| 455 | different physical movement at different magnifications since Leginon uses the image shift |
||
| 456 | calibration at the lower mag where the target comes from to determine the required image |
||
| 457 | shift at the higher mag. |
||
| 458 | |||
| 459 | |||
| 460 | |||
| 461 | |||
| 462 | No current solution |
||
| 463 | |||
| 464 | |||
| 465 | |||
| 466 | |||
| 467 | |||
| 468 | h2. Automatic hole finding fails to pick up good ice thickness |
||
| 469 | after set up |
||
| 470 | |||
| 471 | |||
| 472 | |||
| 473 | Commonly Why: |
||
| 474 | * Electron beam intensity has changed |
||
| 475 | |||
| 476 | |||
| 477 | * Blob threshold is not inclusive enough to tolerate variations |
||
| 478 | |||
| 479 | |||
| 480 | * Ice contaimination |
||
| 481 | |||
| 482 | |||
| 483 | |||
| 484 | |||
| 485 | |||
| 486 | Solutions |
||
| 487 | * Monitor electron beam with a saved location with an empty hole in navigator and |
||
| 488 | check the recorded intensity periodically and adjust zero ice thickness |
||
| 489 | accordingly. |
||
| 490 | |||
| 491 | |||
| 492 | * Reduce Blob threshold in Alias nodes of the Hole Finder. |
||
| 493 | |||
| 494 | |||
| 495 | * Make sure the anticontamination cryo box is properly inserted on the scope. |
||
| 496 | |||
| 497 | |||
| 498 | * Remove the film camera from the scope if you don't need film. |
||
| 499 | |||
| 500 | |||
| 501 | |||
| 502 | |||
| 503 | |||
| 504 | |||
| 505 | |||
| 506 | |||
| 507 | h2. No broken grid square for dose measurement and ZLP alignment |
||
| 508 | (MSI-Tomography) |
||
| 509 | |||
| 510 | |||
| 511 | |||
| 512 | Solution |
||
| 513 | * Burn a hole clear of ice and pick it as the reference target in "Hole Targeting" |
||
| 514 | or "Tomography Targeting" |
||
| 515 | 2 | Amber Herold | |
| 516 | |||
| 517 | |||
| 518 | |||
| 519 | |||
| 520 | |||
| 521 | |||
| 522 | ______ |
||
| 523 | |||
| 524 | [[Troubles with Imaging - Troubleshooting|< Troubles with Imaging]] | [[Troubles with Focusing and Drift Check|Troubles with Focusing and Drift Check >]] |
||