Troubles with Targeting » History » Version 3

Eric Hou, 06/24/2010 02:28 PM

1 1 Amber Herold
h1. Troubles with Targeting
2
3
h2. Grid mosaic is too small to see the details
4
5 3 Eric Hou
h3. Why:
6 1 Amber Herold
* The image scale for creating mosaic is set too small.
7
8 3 Eric Hou
h3. Solution:
9
#  Leginon/Square Targeting/Mosaic Settings> Change "Scale Image" to a larger
10 1 Amber Herold
size
11 3 Eric Hou
 
12
#  Leginon/Square Targeting/Mosaic Settings> Click "Apply" and then "Create" the
13 1 Amber Herold
mosaic atlas.
14
15
h2. Grid mosaic is not patched together well
16
17 3 Eric Hou
h3. Commonly Why:
18 1 Amber Herold
* LM mode U-center defocus not set to 0.
19
20
* LM mode rotation center is off.
21
22
* Bad stage position matrix calibration
23
24 3 Eric Hou
h3. Solutions:
25 1 Amber Herold
26 3 Eric Hou
#  Correct the alignemnt error or redo <link linkend="mcal_st_pos">stage position matrix calibration</link>
27
&nbsp;
28
#  Acquire a new atlas
29
&nbsp;
30
**  Leginon/Grid Targeting/Setting> Enter a new label
31
&nbsp;
32
**  Leginon/Grid Targeting> Click "Calculate Atlas"
33
&nbsp;
34
**  Leginon/Grid Targeting> Click "Submit Targets" (play icon) to begin acquiring the atlas.
35
&nbsp;
36
**  Leginon/Square Targeting> Follow the atlas progress in the display.
37 1 Amber Herold
38 3 Eric Hou
h2. Grid mosaic display in Leginon is different from that of the Web Viewer
39 1 Amber Herold
40 3 Eric Hou
h3. Why:
41
* The goniometer does not reach the intended position due to ,most commonly, a pole touch error that temporaily disables the goniometer.
42 1 Amber Herold
43 3 Eric Hou
* The calibration used to move to the target in "Grid" node is not the same as the calibration parameter used to create the mosaic in "Square Targeting" node are not consistent.
44 1 Amber Herold
45 3 Eric Hou
h3. Solution:
46 1 Amber Herold
47 3 Eric Hou
*  Other than fix the goniometer, Targeting on these mosaic can still work if the error does not occur frequently enough to interfere drift management.
48 1 Amber Herold
49 3 Eric Hou
*  Check "Leginon/Grid/Settings/Use ?? to move to target" against "Leginon/Square Targeting/Mosaic Settings/CalibrationParameters". The latter should be changed if the mosaic is already acquire and what seen on the Web is better than what is in Leginon. The mosaic atlas can be recreated by clicking the "Create" button.
50 1 Amber Herold
51 3 Eric Hou
h3. More Explanation:
52
&nbsp;
53
*  The web tools and Leginon main program use different mechanism to paste together the atlas.
54
&nbsp;
55
The targets Leginon "Grid Targeting" node send out is an even spaced raster. "Grid" node uses that information and the move type set for the node to calculate the goniometer position it needs to aim for in order to get to where it really wants to go to and acquires the image. It is done this way so that it can handle any complex move type. For example, lets say that the model is p=x^2 so that to get to position p' from 0, the goniometer x needs to be sqrt(p'). "Grid Targeting" send the target as p', "Grid" figure out that x'=sqrt(p') and send the x value to the scope to move.
56
&nbsp;
57
When it needs to display the atlas in "Square targeting", Leginon uses the actual goniometer position of each tile image back calculates the targeted position according to the model. This has the advantage that if there is a problem with the goniometer and the stage does not move to where it needs to (such as after a pole touch error), the tile image will appear at its actual position. In the example above, the image will appear at p' from the image because from the goniometer position x'=sqrt(p'), p=x^2=(sqrt(p'))^2.
58
&nbsp;
59
The webtool, however, pastes the atlas together using the targets saved from "Grid Targeting". It does not care what you have set in "Square Targeting" node. Therefore, it will always look right, unless of course the stage did not move to the target when it acquired the image.
60 1 Amber Herold
61
h2. Commonly Why:
62
63
* You did not initiate refresh by clicking the "current position" tool
64
65
* More than one atlas is of the same label.
66
67 3 Eric Hou
h3. Solutions:
68 1 Amber Herold
69 3 Eric Hou
*  No user correction can be made with atlas of the same label. It requires modification of the database.
70 1 Amber Herold
71
*  Acquire a new atlas with a different label
72
73 3 Eric Hou
h2. Targeting is consistently off in the same direction and distance
74 1 Amber Herold
75 3 Eric Hou
Commonly Why: Image shifts in "hl", "sq", or "gr"presets are not aligned with the presets at higher mags. The low mag "sq" preset is especially prone to this problem.
76 1 Amber Herold
77 3 Eric Hou
***Important: When the targeted image is off, it normally means the image shift of the preset where the target comes from is off. That is, if hole images appear off, it is not the image shift of "hl" preset that needs correcting but that of "sq" preset.
78 1 Amber Herold
79
Solution:
80
81 3 Eric Hou
*  Pause before next hole or square by clicking Hole/Toolbar>Pause button or Square/Toolbar>Pause button, respectively. Which one to pause depends on what is the next target.
82 1 Amber Herold
83
*  Center the grid at a location that can be recognized by either:
84 3 Eric Hou
&nbsp;
85
##  Move to the alignment position:
86
&nbsp;
87
(It may be necessary to adjust the position at the scope after this since the goniometer can not move back to a position as accurately as we need it).
88
&nbsp;
89
***  Leginon/Navigation/Stage Location> Choose the name for the stored alignment position ("align" if you have been following the setup check list). and send it "To Scope"
90
&nbsp;
91
##  scope> move the specimen to a position on the same square, either on a recognizable object or burn a good hole through ice as preparation for lower mag preset image shift correction
92 1 Amber Herold
93 3 Eric Hou
*  Leginon/Presets Manager> send "fa" or other high mag preset with (0.0,0.0) image shift "To Scope"
94 1 Amber Herold
95 3 Eric Hou
*  Leginon/Navigation> "Acquire" an image. The recognizable object or the burn hole should be at the center of the image, if not move the grid at the microscope.
96 1 Amber Herold
97
*  Check, adjust, and save the lower mag preset image shift in hole and sq
98 3 Eric Hou
&nbsp;
99
##  Leginon/Presets Manager> send the preset "To Scope".
100
&nbsp;
101
##  Leginon/Navigation> "Acquire" an image in the Camera Configuration.
102
&nbsp;
103
##  Leginon/Navigation> "Navigate" by "Image Shift" to center on the reference object or the burn mark made in the high mag presets.
104
&nbsp;
105
##  Leginon/Presets Manager> save the new image shift by retrieving them "From Scope"
106
&nbsp;
107
"Navigate" means selecting the navigate tool on the top right of image display and then left-clicking the location of the new center that is translated by the given TEM Parameter.
108 1 Amber Herold
109 3 Eric Hou
*  Leginon/Drift Manager/Toolbar>"Declare Drift"(declare drift icon) to force target shift check at the next possible time point. This will correct further targeting.
110 1 Amber Herold
111 3 Eric Hou
*  "Continue" data collection on next hole or square target by clicking Hole/Toolbar> "Continue" button or Square/> "Continue" button, respectively.
112 1 Amber Herold
113 3 Eric Hou
Still not working: Your high mag "fc", "en", or "ef" has an image shift other than that of "fa". The latter is commonly used in image shift alignment of presets.
114 1 Amber Herold
115
Solution:
116
117 3 Eric Hou
*  Leginon/Presets Manager> select individual suspect to check its image shift x and y, change them to the correct values but not send the preset to Scope.
118 1 Amber Herold
119 3 Eric Hou
h2. Targeting is consistently off in opposite direction from where you pick
120 1 Amber Herold
121
Commonly Why: Camera configuration is different in calibration and now.
122
123 3 Eric Hou
#  All your targeting would be wrong. "Abort" target processing in the Target Watcher of "squares" and "holes".
124
&nbsp;
125
#  If you use the Tietz camera: scope/Leginon/Instrument/Camera> check your camera configuration. mirror should be ['vertical']. Change it if not, and "Set" the configuration.
126
&nbsp;
127
#  If you use the Gatan camera: scope/Digital Micrograph> check your camera configuration.
128
&nbsp;
129
#  Leginon/Presets Manager> send "sq" or "grid" "To Scope"
130
&nbsp;
131
#  Leginon/Navigation> "Acquire" an image with "Use this configuration" Unchecked in the Camera Configuration.
132
&nbsp;
133
#  Leginon/Navigation> "Navigate" by "Image Shift", "Beam Shift", or "Stage Position" to center on a reference object or burn mark made in the high mag presets. This is to check if it behaves correctly now.
134
&nbsp;
135
#  Select targets again, or, if atlas is wrong, republish the target list to make the atlas.
136 1 Amber Herold
137 3 Eric Hou
h2. Hole targeting is inconsistently off or goes bad in the next run after correction
138 1 Amber Herold
139 3 Eric Hou
Commonly Why: Not well understood but may be related to microscopy alignment and settings such as normalization of lenses. up to 1.5 um error has been observed even with a reasonable calibration. The following solutions only works some times.
140 1 Amber Herold
141 3 Eric Hou
#  Leginon/Drift Manager>"Declare Drift" to force correction for possible drift.
142
&nbsp;
143
#  Repeat image shift correction alignment <link linkend="toffinsamedir">as in the case when targeting is consistently off</link>.
144
&nbsp;
145
#  Ignore the first targetted '"hole" in the "squares" image. The second and on are more consistent when there is a hysteresis problem.
146
&nbsp;
147
#  Repeat <link linkend="gonm_m_only">"Mag Only" Modeled Stage Calibration</link> at the "square" preset.
148 1 Amber Herold
149 3 Eric Hou
h2. Hole targeting is off more when the image shift target is away from the center of the image
150 1 Amber Herold
151 3 Eric Hou
Commonly Why: Not well understood but may be related to microscopy image shift causes different physical movement at different magnifications since Leginon uses the image shift calibration at the lower mag where the target comes from to determine the required image shift at the higher mag.
152 1 Amber Herold
153
No current solution
154
155 3 Eric Hou
h2. Automatic hole finding fails to pick up good ice thickness after set up
156 1 Amber Herold
157
Commonly Why:
158 3 Eric Hou
159 1 Amber Herold
* Electron beam intensity has changed
160
161
* Blob threshold is not inclusive enough to tolerate variations
162
163
* Ice contaimination
164
165
Solutions
166
167 3 Eric Hou
* Monitor electron beam with a saved location with an empty hole in navigator and check the recorded intensity periodically and adjust zero ice thickness accordingly.
168 1 Amber Herold
169
* Reduce Blob threshold in Alias nodes of the Hole Finder.
170
171
* Make sure the anticontamination cryo box is properly inserted on the scope.
172
173
* Remove the film camera from the scope if you don't need film.
174 2 Amber Herold
175 3 Eric Hou
h2. No broken grid square for dose measurement and ZLP alignment (MSI-Tomography)
176 2 Amber Herold
177
Solution
178
179 3 Eric Hou
* Burn a hole clear of ice and pick it as the reference target in "Hole Targeting" or "Tomography Targeting"
180 2 Amber Herold
181
______